AD-TECH Spencer Probe Depth Electrode, REF SD08R-AP58X-000 8-Contact 1.12mm (A style) Depth Elect...

FDA Recall #Z-0221-2025 — Class II — September 19, 2024

Recall #Z-0221-2025 Date: September 19, 2024 Classification: Class II Status: Ongoing

Product Description

AD-TECH Spencer Probe Depth Electrode, REF SD08R-AP58X-000 8-Contact 1.12mm (A style) Depth Electrode

Reason for Recall

Ad-Tech is recalling these A-Style SD Depth Electrodes because a label discrepancy was identified on A style contact SD electrodes where the MR Conditional symbol was incorrectly placed instead of the MR Unsafe symbol. The A style contacts were not cleared for MR Conditional use. Our records indicate you have received one or more of the affected lots listed above.

Recalling Firm

Ad-Tech Medical Instrument Corporation — Oak Creek, WI

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

14 units

Distribution

US Nationwide distribution in the states of CA, OH.

Code Information

UDI/DI 00841823107688, Lot Number 168788 855856857, 172999 873874875R, 174915 879880881

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated