Ovation Prime Abdominal Stent Graft System - 29mm Aortic Body stent grafts. The TriVascular Ov...

FDA Device Recall #Z-0125-2015 — Class I — September 20, 2014

Recall Summary

Recall Number Z-0125-2015
Classification Class I — Serious risk
Date Initiated September 20, 2014
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Trivascular, Inc
Location Santa Rosa, CA
Product Type Devices
Quantity 0 Devices

Product Description

Ovation Prime Abdominal Stent Graft System - 29mm Aortic Body stent grafts. The TriVascular Ovation Prime Abdominal Stent Graft System is a low-profile endovascular device delivered via catheter to treat descending abdominal aortic aneurysms (AAAs).

Reason for Recall

TriVascular has identified select lots of Ovation Prime 29mm Aortic Body stent grafts that may be associated with rapid emptying of the fill polymer syringe when it is connected to the Autoinjector during the fill polymer injection step of the primary implant procedure.

Distribution Pattern

International Distribution Only -- Canada, Turkey, Austria, Ireland, Belgium, Italy, Switzerland, Spain, France, and Cyprus.

Lot / Code Information

The devices subject to this action are labeled for distribution outside of the United States; therefore, no UDI is on the device label or package. Model Number: TV-AB2980-D Lot Numbers: FS031814-11 FS052214-22 FS052814-36 FS040414-21 FS050614-56 FS040714-53 FS040514-02 FS042614-05 FS050614-53 FS032614-46 FS050214-27 FS050314-12 FS052214-48 FS052214-53 FS031814-44 FS031814-49 FS032714-04 FS032714-20 FS040414-30 FS040514-08 FS050114-01 FS050114-08 FS050514-18 FS050514-27 FS050614-54 FS050614-55 FS051914-32 FS051914-34 Expiry dates ranging "May-17" through "Jun-17"

Other Recalls from Trivascular, Inc

Recall # Classification Product Date
Z-1297-2014 Class II Ovation Prime Abdominal Stent Graft System; Ma... Feb 27, 2014
Z-0288-2013 Class II TriVascular brand Fill Polymer Kit, Endovascula... Oct 19, 2012

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.