HAND DRAPING SJH-Procedure Kit Catalog Number: SJHD67F
FDA Recall #Z-2300-2024 — Class II — May 15, 2024
Product Description
HAND DRAPING SJH-Procedure Kit Catalog Number: SJHD67F
Reason for Recall
Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.
Recalling Firm
American Contract Systems Inc — Tiffin, OH
Classification
Class II — May cause temporary or medically reversible adverse health consequences.
Product Type
Devices
Product Quantity
352 units
Distribution
US Nationwide distribution.
Code Information
UDI-DI: 00191072192349 LOT#"s: 662231 981241 917241 882241
Status
Ongoing
Voluntary / Mandated
Voluntary: Firm initiated
What Should You Do?
Stop using this medical device immediately and consult your healthcare provider. Contact the recalling firm for instructions on return or replacement. If you experienced an adverse event, inform your doctor.