Beacon Tip Royal Flush Plus High-Flow Catheter Catheter, Intravascular, Diagnostic The catheter...

FDA Device Recall #Z-2611-2016 — Class II — April 15, 2016

Recall Summary

Recall Number Z-2611-2016
Classification Class II — Moderate risk
Date Initiated April 15, 2016
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Cook Inc.
Location Bloomington, IN
Product Type Devices
Quantity 532,395

Product Description

Beacon Tip Royal Flush Plus High-Flow Catheter Catheter, Intravascular, Diagnostic The catheters are intended for use in angiographic procedures by physicians trained and experienced in angiographic techniques. Standard techniques for placement of vascular access sheaths, angiographic catheters and wire guides should be employed.

Reason for Recall

Increase in reports of polymer degradation of the catheter tip, resulting in tip fracture and/or separation. degradation could include loss of device function, separation of a device segment leading to medical intervention, or complications resulting from a separated segment. Such complications include device fragments in the vascular system, genitourinary system, or other soft tissues. Fragments within the vascular system could result in embolization to the heart or lungs, or occluding blood flow to end organs.

Distribution Pattern

Worldwide Distribution - US (Nationwide) and Internationally to Algeria, Andorra, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Barbados, Belgium, Bermuda, Brazil, Brunei Darussalam, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Djibouti, Dominican Republic, Ecuador, Egypt, El Salvador, Finland, France, French Polynesia, Georgia, Germany, Greece, Guadeloupe, Guatemala, Guernsey, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jersey, Jordan, Kazakhstan, Kenya, Korea, Kosovo, Kuwait, Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Macau, Malaysia, Malta, Martinique, Mauritius, Mexico, Monaco, Mongolia, Morocco, Myanmar, Netherlands, Netherlands Antilles, New Caledonia, New Zealand, Nicaragua, Norway, Oman, Pakistan, "Palestinian Territory, Occupied", Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Trinidad And Tobago, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Vietnam, U.S. Virgin Islands, Yugoslavia.

Lot / Code Information

All Lots. Catalog prefixes HNR5.0 Catalog/product number; Global/order number HNR5.0-35-100-P-10S-0 G09697 HNR5.0-35-100-P-10S-CFP G11264 HNR5.0-35-100-P-10S-PIG G09059 HNR5.0-35-100-P-8S-VCF G31403 HNR5.0-35-100-P-NS-0 G10313 HNR5.0-35-100-ST-10S-0 G09565 HNR5.0-35-100-ST-10S-PIG G09525 HNR5.0-35-110-P-10S-0 G11241 HNR5.0-35-110-P-10S-PIG G11265 HNR5.0-35-65-P-8S-VCF G12129 HNR5.0-35-70-P-10S-0 G09411 HNR5.0-35-70-P-10S-CFP G11262 HNR5.0-35-70-P-10S-PIG G08832 HNR5.0-35-70-P-10S-PIG-RICKETTS-081983 G11280 HNR5.0-35-70-P-4S-PED G11281 HNR5.0-35-70-P-NS-0 G11959 HNR5.0-35-70-ST-10S-0 G09564 HNR5.0-35-70-ST-10S-PIG G09050 HNR5.0-35-80-P-8S-NEFF-CANADA-082098 G11860 HNR5.0-35-90-P-10S-0 G09412 HNR5.0-35-90-P-10S-CFP G11266 HNR5.0-35-90-P-10S-PIG G09408 HNR5.0-35-90-P-10S-PIG-RICKETTS-081983 G11282 HNR5.0-35-90-P-6S-PED G11729 HNR5.0-35-90-P-8S-VCF G12799 HNR5.0-35-90-ST-10S-0 G11641 HNR5.0-35-90-ST-10S-PIG G09197 HNR5.0-38-100-P-10S-0 G11267 HNR5.0-38-100-P-10S-CFP G11268 HNR5.0-38-100-P-10S-PIG G09285 HNR5.0-38-100-P-NS-0 G27874 HNR5.0-38-110-P-10S-0 G09461 HNR5.0-38-110-P-10S-PIG G11269 HNR5.0-38-70-P-10S-0 G11270 HNR5.0-38-70-P-10S-CFP G11271 HNR5.0-38-70-P-10S-PIG G11272 HNR5.0-38-70-P-NS-0 G11289 HNR5.0-38-80-P-10S-OPC G27877 HNR5.0-38-90-P-10S-0 G11273 HNR5.0-38-90-P-10S-CFP G11274 HNR5.0-38-90-P-10S-PIG G11275

Other Recalls from Cook Inc.

Recall # Classification Product Date
Z-0388-2022 Class II Flexor Check-Flo Introducer Ansel Modification ... Nov 12, 2021
Z-0387-2022 Class II Flexor Check-Flo Introducer Ansel Modificatio... Nov 12, 2021
Z-0414-2022 Class II Skinny Needle with Chiba Tip. RPN 090010, 09001... Oct 13, 2021
Z-0222-2022 Class II Transseptal Needle, Trocar. RPN TSNC-18-71.0, T... Oct 8, 2021
Z-0223-2022 Class II Transseptal Needle with Catheter. RPN TSN-17-75... Oct 8, 2021

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.