Medtronic Sutureless Pump Connector Revision Kit, model 8578. Contents: catheter interface with...

FDA Device Recall #Z-1573-2013 — Class I — June 3, 2013

Recall Summary

Recall Number Z-1573-2013
Classification Class I — Serious risk
Date Initiated June 3, 2013
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Medtronic Neuromodulation
Location Minneapolis, MN
Product Type Devices
Quantity 115,722 total

Product Description

Medtronic Sutureless Pump Connector Revision Kit, model 8578. Contents: catheter interface with attached sutureless pump connector, catheter, connector pin, and strain-relief sleeve to be used with Medtronic SynchroMed implantable drug infusion pumps. Contents of inner package are STERILE. The Medtronic Model 8578 Sutureless Pump Connector Revision Kit is used when a pump connector for an Indura 1P Model 8709 or Model 8709SC catheter is required. The catheter is part of an infusion system that stores and delivers parenteral drugs to the intrathecal space. The implanted infusion system components consist of a Medtronic pump and an Indura 1P Model 8709 or Model 8709SC catheter. The catheter connects to the pump with the Model 8578 sutureless pump connector at the catheter port.

Reason for Recall

The Sutureless Connector (SC) Intrathecal Catheter connector has been redesigned to reduce the potential for occlusion at the catheter to pump interface. Medtronic is removing the unused products from the market that were manufactured with the previous design, and recommend the previous design no longer be used due to greater potential for misalignment and subsequent occlusion.

Distribution Pattern

Worldwide distribution: US (nationwide) and countries of: Aruba, Australia, Austria, Belgium, Canada, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Finland, France, Georgia, Germany, Greece, Guadeloupe, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Jordan, Korea, Kuwait, Lebanon, Lithuania, Luxembourg, Malta, Martinique, Morocco, Netherlands, Netherlands Antilles, New Zealand, Norway, Panama, Poland, Portugal, Puerto Rico, Qatar, Romania, Russian Federation, San Marino, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Thailand, Trinidad and Tobago, Turkey, United Arab Emirates, United Kingdom, Uruguay, and Venezuela.

Lot / Code Information

Product having a Use By Date prior to 25 Aug 2014

Other Recalls from Medtronic Neuromodulation

Recall # Classification Product Date
Z-1043-2026 Class II A71200 Vanta" Clinician Programmer Application ... Dec 3, 2025
Z-0996-2026 Class II Restore Clinician Programmer Application (CP Ap... Nov 20, 2025
Z-1704-2025 Class II Enhanced Verify Evaluation Handset (CFN HH90130FA) Apr 3, 2025
Z-1543-2025 Class II Stimulation RC Clinician Programmer Application... Mar 4, 2025
Z-1506-2025 Class II A820 myPTM Software Application associated with... Jan 9, 2025

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.