Artis icono biplane with software VE20C-diagnostic imaging angiography system Model: 11327600

FDA Device Recall #Z-0516-2022 — Class II — October 12, 2021

Recall Summary

Recall Number Z-0516-2022
Classification Class II — Moderate risk
Date Initiated October 12, 2021
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Siemens Medical Solutions USA, Inc
Location Malvern, PA
Product Type Devices
Quantity 72 units

Product Description

Artis icono biplane with software VE20C-diagnostic imaging angiography system Model: 11327600

Reason for Recall

Four potential software issues with Artis pheno and Artis icono systems with software VE20C: 1. Updated calibration data not saved with measurement after scene+/-; 2. No x-ray possible, system shutdown/restart might be required during intervention; 3. Corrupted Image during Roadmap; 4. Unintended shutdown of Imaging System with UPS (Uninterruptable Power Supply) option

Distribution Pattern

US Nationwide distribution.

Lot / Code Information

Serial Numbers: 180062 180318 180054 180019 180065 180057 180041 180317 180034 180072 180349 180060 180337 180070 180048 180083 180369 180370 180345 180315 180343 180080 180301 180305 180067 180330 180040 180360 180333 180029 180352 180336 180316 180068 180363 180084 180334 180018 180350 180354 180046 180047 180335 180085 180035 180339 180357 180358 180361 180362 180081 180331 180021 180342 180321 180309 180356 180368 180341 180323 180329 180043 180049 180077 180086 180053 180055 180015 180014 180066 180082 180037 UDI: 04056869063317 Expanded Recall 12/20/21 S/N: 180056

Other Recalls from Siemens Medical Solutions USA, Inc

Recall # Classification Product Date
Z-1175-2026 Class II Interventional Fluoroscopic X-Ray System: LUMIN... Dec 29, 2025
Z-1174-2026 Class II Interventional Fluoroscopic X-Ray System: LUMIN... Dec 29, 2025
Z-1241-2026 Class II NAEOTOM Alpha.Prime Software applications: s... Dec 19, 2025
Z-1240-2026 Class II NAEOTOM Alpha Software applications: syngo.C... Dec 19, 2025
Z-1239-2026 Class II SOMATOM X.ceed Software applications: syngo.... Dec 19, 2025

Frequently Asked Questions

A software recall means the device's embedded software or firmware has a defect that could affect its performance or safety. Many software recalls are corrected through firmware updates that can be applied without physically replacing the device. For implantable devices, the update may be delivered wirelessly during a routine clinic visit. For external devices, the manufacturer may provide updated software files or replacement units. Contact your healthcare provider to determine whether your specific device and software version are affected and what action is recommended.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.