Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; OPEN HEART SUPP...
FDA Device Recall #Z-0444-2026 — Class I — October 10, 2025
Recall Summary
| Recall Number | Z-0444-2026 |
| Classification | Class I — Serious risk |
| Date Initiated | October 10, 2025 |
| Status | Ongoing |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Medline Industries, LP |
| Location | Northfield, IL |
| Product Type | Devices |
| Quantity | 96,406 total units |
Product Description
Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; OPEN HEART SUPPLEMENT DYNJ909116D; VASCULAR DYNJ909124F
Reason for Recall
Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.
Distribution Pattern
International distribution to Mexico and Panama. US nationwide distribution to the following: AK AL AR AZ CA CO CT FL GA IA IL IN KS KY MA ME MI MN MO MS MT NC NJ NY OH OR PA PR RI SC TN TX WA WI;
Lot / Code Information
DYNJ909116D UDI-DI 40198459055356 (case), 10198459055355 (ea) Lots 25FBM604 25GBP604; DYNJ909124F UDI-DI 40198459130619 (case), 10198459130618 (ea) Lots 25FBB831 25GBI833
Other Recalls from Medline Industries, LP
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-2235-2026 | Class I | Medline kits containing Huons Co. Bupivacaine H... | Apr 10, 2026 |
| Z-2237-2026 | Class I | Medline kits containing Huons Co. Bupivacaine H... | Apr 10, 2026 |
| Z-2236-2026 | Class I | Medline kits containing Huons Co. Bupivacaine H... | Apr 10, 2026 |
| Z-2234-2026 | Class I | Medline kits containing Huons Co. Bupivacaine H... | Apr 10, 2026 |
| Z-2232-2026 | Class I | Medline kits containing Huons Co. Bupivacaine H... | Apr 10, 2026 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.