C7000 - CUSA¿ Clarity Ultrasonic Tissue Ablation System

FDA Device Recall #Z-0646-2020 — Class II — October 21, 2019

Recall Summary

Recall Number Z-0646-2020
Classification Class II — Moderate risk
Date Initiated October 21, 2019
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Integra LifeSciences Corp.
Location Plainsboro, NJ
Product Type Devices
Quantity N/A

Product Description

C7000 - CUSA¿ Clarity Ultrasonic Tissue Ablation System

Reason for Recall

The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operators manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.

Distribution Pattern

US: AL AR AZ CA CO CT FL GA IA ID IL IN KS KY LA MA MD MI MN MO MS NC NE NJ NY OH OK OR PA SC TN UT VA WA WI WV International: SG MX CA JP AU CO TH HK

Lot / Code Information

Korean OPS Manual P/N 60905212

Other Recalls from Integra LifeSciences Corp.

Recall # Classification Product Date
Z-1295-2025 Class II AURORA Surgiscope System, Sterile, single use d... Feb 6, 2025
Z-1168-2025 Class II Brand Name: MediHoney¿ Gel Product Name: MediH... Jan 10, 2025
Z-0977-2025 Class II CODMAN CERTAS Plus Programmable Valve Inline Sm... Dec 16, 2024
Z-0978-2025 Class II CODMAN CERTAS Plus Programmable Valve Inline Sm... Dec 16, 2024
Z-0979-2025 Class II CODMAN CERTAS Plus Programmable Valve Right Ang... Dec 16, 2024

Frequently Asked Questions

Sterility recalls for medical devices vary in severity. If you have already had a procedure using a potentially non-sterile device, contact your healthcare provider immediately — you may need monitoring for signs of infection. Symptoms to watch for include fever, redness or swelling at the surgical site, unusual pain, or discharge. For devices that have not yet been used, they should be quarantined and returned to the manufacturer per the recall notice. Non-sterile implants can cause serious infections; early detection and treatment are critical.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.