| Date | Product | Reason | Class | Firm |
|---|---|---|---|---|
| Sep 18, 2013 | Siemens Dimension Vista Homocysteine Flex Reagent Cartridge The HCYS metho... | Dimension Vista HCYS lot 13064MA shows erratic recovery of QC and patient samples. | Class II | Siemens Healthcare Diagnostics, Inc. |
| Sep 17, 2013 | Siemens Dimension LOCI TNI Troponin I Calibrator RC 621. The LOCI TNI ... | Siemens Dimension LOCI TNI Troponin I calibrator lot 3DD007 may demonstrate an average upward shi... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Aug 30, 2013 | Emit 2000 Phenobarbital Assay and Syva Emit 2000 Phenobarbital Assay | Imprecision with certain lots | Class II | Siemens Healthcare Diagnostics, Inc. |
| Aug 21, 2013 | Microscan rapID/S panels NC3.33 Sold in Japan only. Product Usage: Mi... | Confirmed increase in false susceptible and false intermediate misreads for Imipenem and/or merop... | Class I | Siemens Healthcare Diagnostics, Inc. |
| Aug 21, 2013 | Microscan plus Negative Combo 3 Sold outside of US only. Product Usage:... | Confirmed increase in false susceptible and false intermediate misreads for Imipenem and/or merop... | Class I | Siemens Healthcare Diagnostics, Inc. |
| Aug 21, 2013 | Microscan rapID/S panels NC3.12 Sold in Japan only. Product Usage: Mi... | Confirmed increase in false susceptible and false intermediate misreads for Imipenem and/or merop... | Class I | Siemens Healthcare Diagnostics, Inc. |
| Aug 21, 2013 | Microscan Synergies Plus Negative Urine Combo 1 Product Usage: MicroScan¿... | Confirmed increase in false susceptible and false intermediate misreads for Imipenem and/or merop... | Class I | Siemens Healthcare Diagnostics, Inc. |
| Aug 21, 2013 | Microscan Synergies Plus Negative Combo 2 Product Usage: MicroScan¿ Syne... | Confirmed increase in false susceptible and false intermediate misreads for Imipenem and/or merop... | Class I | Siemens Healthcare Diagnostics, Inc. |
| Aug 21, 2013 | Microscan plus Negative Urine Combo 4 Sold outside of US only. Product ... | Confirmed increase in false susceptible and false intermediate misreads for Imipenem and/or merop... | Class I | Siemens Healthcare Diagnostics, Inc. |
| Aug 21, 2013 | Microscan plus Negative Breakpoint Combo 4 Sold outside of US only. Pro... | Confirmed increase in false susceptible and false intermediate misreads for Imipenem and/or merop... | Class I | Siemens Healthcare Diagnostics, Inc. |
| Aug 21, 2013 | MicroscanrapID/S panels NC3.11 Sold in Japan only. Product Usage: Mic... | Confirmed increase in false susceptible and false intermediate misreads for Imipenem and/or merop... | Class I | Siemens Healthcare Diagnostics, Inc. |
| Aug 21, 2013 | Microscan Synergies Plus Negative Breakpoint Combo 7 Product Usage: Micro... | Confirmed increase in false susceptible and false intermediate misreads for Imipenem and/or merop... | Class I | Siemens Healthcare Diagnostics, Inc. |
| Aug 21, 2013 | Microscan Synergies Plus Negative Urine Combo 2 Product Usage: MicroScan¿... | Confirmed increase in false susceptible and false intermediate misreads for Imipenem and/or merop... | Class I | Siemens Healthcare Diagnostics, Inc. |
| Aug 21, 2013 | Microscan Synergies Plus Negative Urine Combo 5 Product Usage: MicroSca... | Confirmed increase in false susceptible and false intermediate misreads for Imipenem and/or merop... | Class I | Siemens Healthcare Diagnostics, Inc. |
| Aug 13, 2013 | Siemens EasyLink Informatics System V4.0/V5 Siemens, Material Number (SMN) ... | May not perform as intended under certain conditions, causing the release of results to the Labor... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Jul 30, 2013 | ADVIA Chemistry A1c_3 Calibrators, Catalog Number 10491408, IVD --- COMMON/US... | Siemens confirmed that the ADVIA¿ Chemistry Systems HbA1c method exhibits a positive bias of up t... | Class II | Siemens Healthcare Diagnostics |
| Jul 1, 2013 | Dimension Tacrolimus Flex reagent cartridge (DF107) and Dimension TACR CAL (D... | Siemens initiated a recall due to confirmed complaints of low patient sample recovery with Dimens... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Jun 19, 2013 | Dimension Vista 500 and Dimension Vista 1500 System Aliquot Well Double Disp... | Siemens is conducting a field correction for Dimension Vista 500 and Dimension Vista 1500 systems... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Jun 5, 2013 | Dimension(R) Mass Creatine Kinase MB Isozyme Calibrator (RC420) The Mass C... | Siemens Healthcare Diagnostics has received complaints that the lyophilized cake is not fully dis... | Class II | Siemens Healthcare Diagnostics, Inc. |
| May 20, 2013 | Dimension Vista(R) 500 and Dimension Vista(R) 1500 - Reloaded Vista Vials - l... | There is the potential for the vial volume of "in use" Vista calibrators, controls, or sample dil... | Class II | Siemens Healthcare Diagnostics, Inc. |
| May 10, 2013 | syngo(R) Lab Data Manager - Version VA11B and VA12A Systems Securely conso... | Under certain conditions the system may not perform as intended, causing the release of results t... | Class II | Siemens Healthcare Diagnostics, Inc. |
| May 6, 2013 | Dimension(R) Chemistry Wash (RD701), Siemens Material Number (SMN) 10445052 ... | Siemens Healthcare Diagnostics received complaints of Chemistry Wash causing QC and patient sampl... | Class II | Siemens Healthcare Diagnostics, Inc. |
| May 2, 2013 | Siemens brand MicroScan Dried Negative Breakpoint Combo 47 panels (B1017-417)... | The product has the potential for false negative or delayed reaction for Arginine (ARG) results. ... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Apr 25, 2013 | Dimension(R) TACR Flex(R) reagent cartridge (DF107) The TACR method is an ... | Siemens has confirmed that the TACR method may demonstrate reduced on-board stability which may r... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Apr 22, 2013 | HB1C Flex(R) Reagent Cartridge (DF105A, Siemens Material Number 10483822) for... | Positive bias on the lots listed compared to alternate methods and a higher frequency of "Above A... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Apr 22, 2013 | HbA1c Flex(R) reagent cartridge (K3105A, Siemens Material Number 10470481) fo... | Positive bias on the lots listed compared to alternate methods and a higher frequency of "Above A... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Mar 11, 2013 | Dade(R) Actin(R) FSL Activated PTT Reagent. For use in the determination o... | Siemens has observed that the listed lot numbers show an increase in Heparin sensitivity over the... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Mar 6, 2013 | Siemens Healthcare Diagnostics Inc. MIcroScan MICroSTREP plus Type 1 B1027-20... | Customer complaint investigation confirmed the issue of false skipped wells and false susceptible... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Feb 27, 2013 | Siemens Dimension(R) EXL(TM) LM (Siemens Material Number 10486890). Intend... | Through investigation of customer complaints, Siemens Healthcare Diagnostics has confirmed an inc... | Class III | Siemens Healthcare Diagnostics, Inc. |
| Feb 27, 2013 | Siemens Dimension(R) EXL(TM) 200 (Siemens Material Number 10636929). Inten... | Through investigation of customer complaints, Siemens Healthcare Diagnostics has confirmed an inc... | Class III | Siemens Healthcare Diagnostics, Inc. |
| Feb 27, 2013 | Siemens Dimension(R) EXL(TM) LM STM (Siemens Material Number 106393377). I... | Through investigation of customer complaints, Siemens Healthcare Diagnostics has confirmed an inc... | Class III | Siemens Healthcare Diagnostics, Inc. |
| Feb 22, 2013 | Siemens Dimension(R) Phosphorus (PHOS) Flex(R) reagent cartridge (DF61) Th... | Siemens Healthcare Diagnostics has initiated a voluntary field action for Dimension(R) PHOS Flex(... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Feb 21, 2013 | Siemens Dimension Vista(R) System Vista Drug 4 Calibrator. The device is used... | Siemens has confirmed complaints of low QC and patient recovery with the Dimension Vista(R) LOCI(... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Dec 20, 2012 | Dimension(R) IRON Calibrator (DC85), Siemens material number 10445010. The... | Siemens Healthcare Diagnostics has received complaints of IRON calibrations outside of acceptable... | Class III | Siemens Healthcare Diagnostics, Inc. |
| Dec 20, 2012 | Siemens Total Iron Binding Capacity (IBCT) Flex Reagent Cartridge. An in... | The analytical sensitivity claim of 6 ug/dL (1.1 umol/L) in the Instructions for Use (IFU) is no ... | Class III | Siemens Healthcare Diagnostics, Inc. |
| Dec 19, 2012 | Dimension Vista (R) Urinary/Cerebrospinal Fluid Protein (UCFP) Flex (R) Reage... | Siemens has confirmed an increase on absorbance errors when calibrating Dimension Vista Urinary/C... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Dec 10, 2012 | Emit(R) 2000 Phenobarbital Assay (OSR4D229) and Syva(R) Emit(R) Phenobarbital... | Firm has confirmed higher imprecision with the recalled lots on Beckman Coulter AU(R) Systems and... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Dec 6, 2012 | Siemens Emit(R) tox(TM) Serum Calibrators, catalog number 7B019UL The Emit... | Incorrect expiration date was assigned to the D1 lot of Emit(R) tox(TM) Serum Calibrators. The c... | Class III | Siemens Healthcare Diagnostics, Inc. |
| Nov 27, 2012 | Siemens Dimension(R) Tacrolimus Flex(R) reagent cartridge (DF107). The TACR m... | Siemens has confirmed customer complaints of low patient sample recovery with Dimension(R) TACR F... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Oct 11, 2012 | Siemens StreamLAB(R) Analytical Workcell; Catalog number 746002.901. Desig... | Firm has received reports from customers that the small pins at the bottom of the center door pan... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Oct 10, 2012 | Siemens Dimension Vista(R) Alkaline Phosphatase (ALP) Calibrator The ALP C... | Correlation slope bias and accuracy shift | Class III | Siemens Healthcare Diagnostics, Inc. |
| Oct 8, 2012 | Dimension Vista(R) IRON Flex(R) reagent cartridge (K3085) The IRON method ... | Siemens Healthcare Diagnostics has confirmed the potential for falsely elevated results with IRON... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Aug 9, 2012 | Dimension Vista(R) Systems Vista(R) Theophylline THEO Flex(R) Regent Cartridg... | Firm has received complaints from customers of "Abnormal Reaction" errors occurring on calibratio... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Aug 3, 2012 | Siemens Dimension Vista(R) Homocysteine Flex(R) Reagent Cartridge This is ... | Cartridges are leaking. The leaking is observed when opening the wrapper that holds the reagent ... | Class III | Siemens Healthcare Diagnostics, Inc. |
| Jul 30, 2012 | Emit(R) 2000 Phenobarbital Assay (OSR4D229) and Syva(R) Emit(R) Phenobarbital... | Firm has confirmed highter imprecision with the recalled lots on Beckman Coulter AU(R) Systems. ... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Jul 17, 2012 | Dimension Vista(R) Vancomycin Flex(R) Reagent Cartridge (VANC, Catalog # K408... | Siemens has received complaints of customers experiencing slopes >1.05 when calibrating Flex(R) r... | Class III | Siemens Healthcare Diagnostics, Inc. |
| Jul 17, 2012 | Dimension(R) Reagent Management System (RMS) - Power Cord. The Siemens Hea... | Siemens Healthcare Diagnostics has confirmed that partial insertion of the power cord on the Dime... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Jun 7, 2012 | EasyLink(TM) Informatics System, Part Number 1000034805. EasyLink(TM) Informa... | There is a potential for the EasyLink system to display and transmit to the Laboratory Informatio... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Jun 7, 2012 | Siemens syngo(R) Lab Data Manager. The syngo(R) Lab Data Manager is classi... | The recalling firm has discovered that when using a Sysmex(R) CA-7000 or CA-1500 coagulation anal... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Jun 7, 2012 | Siemens Stratus(R) CS Acute Care(TM) D-dimer Calibrator (DDM CalPak) lots 502... | Siemens has confirmed that the Stratus(R) CS Acute Care(TM) D-Dimer CalPak calibrator lots 502058... | Class III | Siemens Healthcare Diagnostics, Inc. |