| Date | Product | Reason | Class | Firm |
|---|---|---|---|---|
| Nov 19, 2015 | Factor VIII Chromogenic Assay, REF/Catalog No. B4238-40, Siemens Material Num... | Factor VIII Chromogenic Assay Lot 44637 is showing an increased occurrence of irregular curves wh... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Nov 3, 2015 | Dimension Vista System B2 Microglobulin Flex reagent cartridge and urine sta... | Siemens Healthcare Diagnostics has confirmed an increase in the rate of Abnormal Assay errors a... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Oct 7, 2015 | ADVIA Chemistry XPT Systems, Software Version 1.0.3, SMN 11127538 Clinical... | Multiple Software issues. Auto Start-UP Fail, Calibration Interval Resets when a Reagent Blank is... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Sep 30, 2015 | Dimension Vista System (MG) ; Flex reagent cartridge Catalog number K3057, SM... | Erroneous low results on a small number of reagent wells. There is the potential for under-recove... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Sep 24, 2015 | ADVIA Chemistry Hemoglobin A1c_3, A1c_3M Reagent kits. Used on the ADVIA 12... | Reagent in these lots may demonstrate an increased occurrence of high %HbA1c bias, a positive bia... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Aug 28, 2015 | ADVIA Chemistry XPT Chemistry Systems, Siemens Material Number (SMN) 1072303... | ISE module may produce discrepant, unflagged results for Sodium, Potassium or Chloride when the r... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Aug 26, 2015 | syngo Lab Data Manager System - Product Usage: syngo Lab Data Manager is a cl... | Software Issues. | Class II | Siemens Healthcare Diagnostics, Inc. |
| Jul 27, 2015 | ADVIA Centaur¿ Folate Ascorbic Acid/Ascorbic Acid Diluent; SMN 10316877. Use... | Low red blood cell control recoveries and low results for red blood cell patient samples when usi... | Class II | Siemens Healthcare Diagnostics, Inc |
| Jul 23, 2015 | Dimension Vista B¿2-Microglobulin Flex Reagent Cartridge (B2MIC); Product... | customers are receiving a high number of Abnormal Assay Errors or are not able to calibrate B2MIC... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Jul 15, 2015 | Dimension¿ CREA (Creatinine) and Dimension Vista¿ CREA (Creatinine); The CRE... | Siemens has confirmed that the Dimension CREA and Dimension Vista CREA may produce falsely depres... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Jun 30, 2015 | ADVIA Chemistry Acetaminophen reagents; for in vitro diagnostic use in the qu... | A change in the concentration of N-acetylcysteine (NAC) that may cause interference for the ADVIA... | Class II | Siemens Healthcare Diagnostics, Inc. |
| May 21, 2015 | Siemens Healthcare Diagnostics Dimension Vista Systems, used with the Dimensi... | Two software issues may occur in Vista software version 3.6.1. Issue #1: Samples may stop process... | Class II | Siemens Healthcare Diagnostics, Inc. |
| May 19, 2015 | ADVIA Centaur Tni-Ultra Assay; SMN 10317708 - 100 tests and SMN 10317709 -500... | System-to-system bias between the Tni-Ultra assay on the ADVIA Centaur CP and ADVIA Centaur/ADVI... | Class II | Siemens Healthcare Diagnostics, Inc |
| May 19, 2015 | CentraLink" Data Management System; software is a network solution provider a... | Siemens Healthcare Diagnostics has determined that the sample query function that includes Instr... | Class II | Siemens Healthcare Diagnostics, Inc. |
| May 11, 2015 | Factor deficient plasma for the Siemens BCS¿ / BCS¿ XP System. In vitro di... | Discrepancies for Siemens BCS / BCS XP with dilution studies; erroneously high Factor IX results ... | Class II | Siemens Healthcare Diagnostics, Inc. |
| May 6, 2015 | The Rapidlab 1260 system is a blood gas analyzer used for laboratory testing ... | D50 and D51 Diagnostic error codes are not functional. | Class II | Siemens Healthcare Diagnostics, Inc |
| May 6, 2015 | IMMULITE 2000/IMMULITE 2000 XPi Italian Cypress 3g Allergy, Catalog # T23L4 (... | IMMULITE 2000/2000XPi Italian Cypress (T23L4), Lots 206 and 208, fail to meet specifications to s... | Class II | Siemens Healthcare Diagnostics, Inc. |
| May 6, 2015 | The Rapidlab 1265 system is a blood gas analyzer used for laboratory testing ... | D50 and D51 diagnostic error codes are not functional. | Class II | Siemens Healthcare Diagnostics, Inc |
| Apr 29, 2015 | EasyLink Data Management System; a clinical data management system that assis... | System Limitations and Software Issues related to the following features: Custom reference range... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Apr 20, 2015 | Dimension Vista Systems LOCI Thyroid Stimulating Hormone; the TSH method is a... | Dimension Vista LOCI Thyroid Stimulating Hormone (TSH) Lot 14237AA may produce falsely depressed ... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Apr 9, 2015 | IMMULITE/IMMULITE 1000 Systems TBG, catalog LKTB1 Quantitative measurement... | IMMULITE/IMMULITE 1000 TBG catalog LKTB lot 257 is failing the adjustment slope instrument parame... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Apr 2, 2015 | Dimension EXL with LM, Dimension EXL 200, Dimension RxL, Dimension RxL HM, Di... | Certain Dimension Integrated Chemistry Systems Instrument Power Configurations may have been inst... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Mar 25, 2015 | Dade Actin Activated Cephaloplastin Reagent; Liquid rabbit brain cephalin wit... | Recovery for assayed controls are out of assigned ranges. A drift >3 seconds in the normal APTT ... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Mar 23, 2015 | ADVIA Chemistry Lipase Reagent , Catalog Number B01-4840-01 REF: 01984894. | Siemens internal investigation confirmed that current contamination avoidance settings do not pre... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Feb 24, 2015 | QuikLYTE Dilution Check: used to monitor and adjust for the dilution ratio on... | A positive bias beyond 5% resulting in a dilution check failure or an unnecessary dilution check ... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Feb 11, 2015 | ADVIA Chemistry XPT System, Siemens Material Number 10723034, Software Versio... | Siemens Healthcare Diagnostics has confirmed an issue with ADVIA Chemistry XPT software version 1... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Jan 27, 2015 | ADVIA Centaur Systems Progesterone Kit, (5 -pack); In Vitro Diagnostic; Catal... | ADVIA Centaur Systems Progesterone Kit Lots Ending in 268 have a high bias on results greater th... | Class II | Siemens Healthcare Diagnostics, Inc |
| Jan 27, 2015 | ADVIA Centaur Systems Progesterone Kit, (5-pack Ref); In Vitro Diagnostic C... | ADVIA Centaur Systems Progesterone Kit Lots Ending in 268 have a high bias on results greater th... | Class II | Siemens Healthcare Diagnostics, Inc |
| Jan 27, 2015 | ADVIA Centaur Systems Progesterone Kit, (1 -pack) In Vitro Diagnostic; Catal... | ADVIA Centaur Systems Progesterone Kit Lots Ending in 268 have a high bias on results greater th... | Class II | Siemens Healthcare Diagnostics, Inc |
| Jan 19, 2015 | ADVIA Chemistry Systems Drug Calibrator I; For in vitro diagnostic use in the... | Incorrect calibrator lot values were assigned for ADVIA Chemistry Systems Drug Calibrator I, lot ... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Jan 15, 2015 | N Latex IgM; In-vitro diagnostic reagent for the quantitative determination o... | Complaints have been confirmed on failed proficiency testing due to serum samples recovering too ... | Class III | Siemens Healthcare Diagnostics, Inc. |
| Dec 16, 2014 | ADVIA Chemistry XPT interfaced to the CentraLink Data Management System V14x... | Enabling sending of preliminary/initial results on the ADVIA Chemistry XPT creates a risk that a ... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Nov 24, 2014 | Dimension Vista Myoglobin Calibrator (MYO CAL), Lot 4FD085 Product usage:... | Dimension Vista Myoglobin Calibrator (MYO CAL) may produce a positive shift in MYO QC and Patient... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Nov 5, 2014 | Dimension Vista 500 Intelligent Lab System or Dimension Vista 1500 Intelligen... | Customer complaints of discrepant flagged and/or un-flagged patient and QC results. Reagent probe... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Oct 27, 2014 | Turbo Troponin I For in vitro diagnostic use for the quantitative measurem... | Class II | Siemens Healthcare Diagnostics, Inc. | |
| Oct 20, 2014 | IMMULITE Calcitonin Control Module (IMMULITE, IMMULITE 1000, IMMULITE 2000, I... | A typographical error regarding the expiration date was identified in the control module instruct... | Class III | Siemens Healthcare Diagnostics, Inc. |
| Oct 17, 2014 | Rapid Neg BP Combo Panel Type 3; Catalog number B 1017-117. MicroScan Rapi... | An increase in false positive susceptible results on Rapid Neg BP Combo Panel Type 3 and Rapid Ne... | Class I | Siemens Healthcare Diagnostics, Inc. |
| Oct 17, 2014 | Rapid Neg Urine Combo Panel Type 1; Catalog number B 1017-167; Siemens Mater... | An increase in false positive susceptible results on Rapid Neg BP Combo Panel Type 3 and Rapid Ne... | Class I | Siemens Healthcare Diagnostics, Inc. |
| Oct 9, 2014 | ADVIA Centaur XP Immunoassay System (including refurbished) Professional u... | A slight crack of the tubing at the connecting fitting on the reservoir connector assembly may oc... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Oct 9, 2014 | ADVIA Centaur (including refurbished) Professional use in a laboratory for... | A slight crack of the tubing at the connecting fitting on the reservoir connector assembly may oc... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Oct 6, 2014 | ADVIA Centaur BNP Assay for in vitro diagnostic use in the quantitative deter... | Siemens Healthcare Diagnostics has confirmed that approximately 2000 ReadyPacks from the ADVIA Ce... | Class III | Siemens Healthcare Diagnostics, Inc |
| Sep 30, 2014 | Siemens Dimension HB1C Flex reagent cartridge; Dimension Hemoglobin A1c (HB1C... | Certain lots of the reagent cartridges show confirmed positive bias averaging 0.4% Hemoglobin A1c... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Sep 26, 2014 | Siemens Calibrator E; SMN 10309079 (2 pack), SMN 10321075 (6 pack), SMN 10335... | Positive bias for Calibrator E lot kits ending in 38 and 39 with the ADVIA Centaur Systems Cortis... | Class II | Siemens Healthcare Diagnostics, Inc |
| Sep 24, 2014 | siemens Dimension¿ Phenytoin Flex¿ Reagent Cartridge The PTN method used o... | Internal testing has confirmed a low absorbance range between the levels 1 and 5 calibrators with... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Sep 19, 2014 | External power supply adaptors that accompany Clinitek Status Power Supply Ad... | The external power supply for the Clintek Status analyzer, which is provided separately, is damag... | Class II | Siemens Healthcare Diagnostics, Inc |
| Sep 19, 2014 | External power supply adaptors that accompany Clinitek Status Connect System ... | The external power supply for the Clintek Status analyzer, which is provided separately, is damag... | Class II | Siemens Healthcare Diagnostics, Inc |
| Sep 19, 2014 | External power supply adaptors that accompany Clinitek Status+ The Clinite... | The external power supply for the Clintek Status analyzer, which is provided separately, is damag... | Class II | Siemens Healthcare Diagnostics, Inc |
| Sep 15, 2014 | ADVIA Centaur¿¿ Systems HBc Total (HBcT) Kit. For In Vitro Diagnostic Use. C... | Increased imprecision and positive bias is observed with patient samples and negative controls an... | Class II | Siemens Healthcare Diagnostics, Inc |
| Sep 15, 2014 | Siemens ADVIA Centaur¿ Systems HBc Total (HBcT), Reagent. For In Vitro Diagno... | Increased imprecision and positive bias is observed with patient samples and negative controls an... | Class II | Siemens Healthcare Diagnostics, Inc |
| Sep 12, 2014 | ADVIA Centaur¿ and ADVIA Centaur¿ XP and ADVIA Centaur¿ CP Cleaning Solution ... | Stability of the ADVIA Centaur Cleaning Solution is not assured | Class III | Siemens Healthcare Diagnostics, Inc |