Dimension Vista Myoglobin Calibrator (MYO CAL), Lot 4FD085 Product usage: in vitro diagnostic...
FDA Recall #Z-0946-2015 — Class II — November 24, 2014
Product Description
Dimension Vista Myoglobin Calibrator (MYO CAL), Lot 4FD085 Product usage: in vitro diagnostic product for the calibration of Myoglobin (MYO) method on the Dimension Vista System.
Reason for Recall
Dimension Vista Myoglobin Calibrator (MYO CAL) may produce a positive shift in MYO QC and Patient test results that exceeds acceptance criteria for this product. A positive shift observed up to 12% at MYO concentrations within and above the reference range of the assay. Depending on quality control limits, this drift may not be detected.
Recalling Firm
Siemens Healthcare Diagnostics, Inc. — Newark, DE
Classification
Class II — May cause temporary or medically reversible adverse health consequences.
Product Type
Devices
Product Quantity
1230
Distribution
Worldwide Distribution: US (nationwide) and country of: Canada.
Code Information
Lot 4FD085, exp. 12-01-2014
Status
Terminated
Voluntary / Mandated
Voluntary: Firm initiated