571 FDA medical device safety recalls.
FDA Medical Device Recall Enforcement Database
Browse 571 FDA medical device recall enforcement actions. Each entry includes the device description, reason for recall, classification (Class I through III based on severity), recalling firm, and distribution details. Use the filters below to search by year, classification, state, or keyword.
Medical Device Recalls
Browse 571 FDA device recalls.
| Date | Product | Reason | Class | Firm |
|---|---|---|---|---|
| Oct 15, 2021 | Dimension Vista 1500 In-vitro diagnostic device intended to duplicate manual ... | Incorrect Default Hemolysis, Icterus or Lipemia (HIL) Index for Five Dimension Vista Assays | Class II | Siemens Healthcare Diagnostics, Inc. |
| Oct 15, 2021 | Dimension Vista 500- In-vitro diagnostic device intended to duplicate manual ... | Incorrect Default Hemolysis, Icterus or Lipemia (HIL) Index for Five Dimension Vista Assays | Class II | Siemens Healthcare Diagnostics, Inc. |
| Oct 6, 2021 | Aptio Automation Interface Module to ADVIA Centaur XP/XPT. Aptio by Siemens C... | The Aptio Automation Firmware for the ADVIA Centaur XP/XPT Interface Module may lead, in some cas... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Sep 28, 2021 | Atellica IM 1300 Analyzer, SMN 11066001 | Ancillary reagent packs used on Atellica IM 1300 and Atellica IM 1600 analyzers may overflow when... | Class II | Siemens Healthcare Diagnostics, Inc |
| Sep 28, 2021 | Atellica IM 1600 Analyzer, SMN 11066000 | Ancillary reagent packs used on Atellica IM 1300 and Atellica IM 1600 analyzers may overflow when... | Class II | Siemens Healthcare Diagnostics, Inc |
| Sep 22, 2021 | Atellica IM SARS-CoV-2 Antigen Assay (100 Test Kit), SMN # 11207861 | Data provided in the Emergency Use Authorization (EUA) submission was insufficient to determine t... | Class II | Siemens Healthcare Diagnostics, Inc |
| Sep 22, 2021 | ADVIA Centaur SARS-CoV-2 Antigen Assay (100 Test Kit), SMN # 11207866 | Data provided in the Emergency Use Authorization (EUA) submission was insufficient to determine t... | Class II | Siemens Healthcare Diagnostics, Inc |
| Sep 20, 2021 | Atellica CH Urinary/Cerebrospinal Fluid Protein (UCFP), SMN 11097543 | The firm has confirmed the potential for Atellica CH Urinary/Cerebrospinal Fluid Protein reagent ... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Sep 20, 2021 | Advia Chemistry Urinary/Cerebrospinal Fluid Protein (UCFP), SMN 11319151 | The firm has confirmed the potential for ADVIA Chemistry Urinary/Cerebrospinal Fluid Protein reag... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Jul 6, 2021 | Atellica IM BR 27.29 (BR) Assay 50 Test Kit- in vitro diagnostic use in the ... | Product does not meet the Instructions for Use (IFU) claimed Pack Calibration Interval of 10 days... | Class II | Siemens Healthcare Diagnostics, Inc |
| Jul 6, 2021 | Atellica IM BR 27.29 (BR) Assay 250 Test Kit (Reference)- in vitro diagnostic... | Product does not meet the Instructions for Use (IFU) claimed Pack Calibration Interval of 10 days... | Class II | Siemens Healthcare Diagnostics, Inc |
| Jul 6, 2021 | Atellica IM BR 27.29 (BR) Assay 250 Test Kit- in vitro diagnostic use in the ... | Product does not meet the Instructions for Use (IFU) claimed Pack Calibration Interval of 10 days... | Class II | Siemens Healthcare Diagnostics, Inc |
| May 18, 2021 | The ADVIA Chemistry XPT Chemistry System, Siemens Material Number (SMN) 10723... | ADVIA Chemistry System and Software Versions (V1.4 SMN 11314625 and V1.4 China SMN 11316885) may ... | Class II | Siemens Healthcare Diagnostics, Inc. |
| May 13, 2021 | Atellica CH 930 Analyzer - multi-component system for in vitro diagnostic te... | Valves used on the Atellica CH 930, Atellica IM 1300 and Atellica IM 1600 analyzers may develop a... | Class II | Siemens Healthcare Diagnostics, Inc. |
| May 13, 2021 | Atellica IM 1600 Analyzer- multi-component system for in vitro diagnostic te... | Valves used on the Atellica CH 930, Atellica IM 1300 and Atellica IM 1600 analyzers may develop a... | Class II | Siemens Healthcare Diagnostics, Inc. |
| May 13, 2021 | Atellica IM 1300 Analyzer - multi-component system for in vitro diagnostic t... | Valves used on the Atellica CH 930, Atellica IM 1300 and Atellica IM 1600 analyzers may develop a... | Class II | Siemens Healthcare Diagnostics, Inc. |
| May 4, 2021 | Atellica CH 930 Analyzer with Photometer Lamp (SMN 11075676) Siemens Materia... | Photometer Lamp May Reach Saturation Without Flagging Results, may generate erroneously elevated... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Apr 21, 2021 | Centaur Syphilis Assay-SYP Kit, 200 Test - Product Usage: For in vitro diagno... | Potential for the ADVIA Centaur Syphilis assay to carryover into other commercial assays. This ca... | Class II | Siemens Healthcare Diagnostics, Inc |
| Mar 5, 2021 | ADVIA 2120 Dual Aspirate Autosampler (SMN 10313419) - US, automated hematolog... | Potential Sample Identification (SID) Mismatch with14-Character Barcodes | Class II | Siemens Healthcare Diagnostics, Inc. |
| Mar 5, 2021 | ADVIA 2120i Refurb Single Aspirate (SMN 11314044) - US, automated hematology... | Potential Sample Identification (SID) Mismatch with14-Character Barcodes | Class II | Siemens Healthcare Diagnostics, Inc. |
| Mar 5, 2021 | ADVIA 2120 REFURB SAA AUTOSAMPLER (SMN 10374453) - US, automated hematology a... | Potential Sample Identification (SID) Mismatch with14-Character Barcodes | Class II | Siemens Healthcare Diagnostics, Inc. |
| Mar 5, 2021 | ADVIA 2120i Dual Aspirate Autosampler (SMN 10285573) - US, automated hematol... | Potential Sample Identification (SID) Mismatch with14-Character Barcodes | Class II | Siemens Healthcare Diagnostics, Inc. |
| Mar 5, 2021 | ADVIA 2120i with Single Aspirate Autosampler (SMN 10488923) - OUS, automated ... | Potential Sample Identification (SID) Mismatch with14-Character Barcodes | Class II | Siemens Healthcare Diagnostics, Inc. |
| Mar 5, 2021 | ADVIA 2120 REFURB DAA AUTOSAMPLER (SMN 10374454) - US, automated hematology ... | Potential Sample Identification (SID) Mismatch with14-Character Barcodes | Class II | Siemens Healthcare Diagnostics, Inc. |
| Mar 5, 2021 | ADVIA 2120 Dual Aspirate Autosampler (SMN 10313419) - OUS, automated hematolo... | Potential Sample Identification (SID) Mismatch with14-Character Barcodes | Class II | Siemens Healthcare Diagnostics, Inc. |
| Mar 5, 2021 | ADVIA 2120i (RoHS) with Single Aspirate Autosampler (SMN 11219529) - OUS, aut... | Potential Sample Identification (SID) Mismatch with14-Character Barcodes | Class II | Siemens Healthcare Diagnostics, Inc. |
| Mar 5, 2021 | ADVIA 2120i Dual Aspirate Autosampler (SMN 10285573) - OUS, automated hematol... | Potential Sample Identification (SID) Mismatch with14-Character Barcodes | Class II | Siemens Healthcare Diagnostics, Inc. |
| Mar 5, 2021 | ADVIA 2120 REFURB DAA AUTOSAMPLER (SMN 10374454) - OUS, automated hematology ... | Potential Sample Identification (SID) Mismatch with14-Character Barcodes | Class II | Siemens Healthcare Diagnostics, Inc. |
| Mar 5, 2021 | ADVIA 2120i Refurb Dual Aspirate (SMN 11314045) - OUS, automated hematology a... | Potential Sample Identification (SID) Mismatch with14-Character Barcodes | Class II | Siemens Healthcare Diagnostics, Inc. |
| Mar 5, 2021 | ADVIA 2120i Refurb Dual Aspirate (SMN 11314045) - US, automated hematology a... | Potential Sample Identification (SID) Mismatch with14-Character Barcodes | Class II | Siemens Healthcare Diagnostics, Inc. |
| Mar 5, 2021 | ADVIA 2120i (RoHS) Dual Aspirate Autosampler (SMN 11219530) - US, automated... | Potential Sample Identification (SID) Mismatch with14-Character Barcodes | Class II | Siemens Healthcare Diagnostics, Inc. |
| Mar 5, 2021 | ADVIA 2120 REFURB SAA AUTOSAMPLER (SMN 10374453) - OUS, automated hematology ... | Potential Sample Identification (SID) Mismatch with14-Character Barcodes | Class II | Siemens Healthcare Diagnostics, Inc. |
| Mar 5, 2021 | ADVIA 2120 with Single Aspirate Autosampler (SMN 10316162) - US, automated he... | Potential Sample Identification (SID) Mismatch with14-Character Barcodes | Class II | Siemens Healthcare Diagnostics, Inc. |
| Mar 5, 2021 | ADVIA 2120i Refurb Single Aspirate (SMN 11314044) - OUS, automated hematology... | Potential Sample Identification (SID) Mismatch with14-Character Barcodes | Class II | Siemens Healthcare Diagnostics, Inc. |
| Mar 5, 2021 | ADVIA 2120 with Single Aspirate Autosampler (SMN 10316162) - OUS, automated h... | Potential Sample Identification (SID) Mismatch with14-Character Barcodes | Class II | Siemens Healthcare Diagnostics, Inc. |
| Mar 5, 2021 | ADVIA 2120i with Single Aspirate Autosampler (SMN 10488923) - US, automated h... | Potential Sample Identification (SID) Mismatch with14-Character Barcodes | Class II | Siemens Healthcare Diagnostics, Inc. |
| Mar 5, 2021 | ADVIA 2120i (RoHS) with Single Aspirate Autosampler (SMN 11219529) - US, aut... | Potential Sample Identification (SID) Mismatch with14-Character Barcodes | Class II | Siemens Healthcare Diagnostics, Inc. |
| Mar 5, 2021 | ADVIA 2120i (RoHS) Dual Aspirate Autosampler (SMN 11219530) - OUS, automated ... | Potential Sample Identification (SID) Mismatch with14-Character Barcodes | Class II | Siemens Healthcare Diagnostics, Inc. |
| Mar 2, 2021 | BCS XP System: Multipurpose system for in vitro coagulation studies. Product ... | Siemens Healthcare Diagnostics Products GmbH has confirmed that the BCS XP System may be affected... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Feb 8, 2021 | Enzyme linked immunosorbent assay, for herpes simplex virus, hsv-2 | Potential false positive HSV results. Siemens Healthcare Diagnostics, Inc. initiated a correctio... | Class II | Siemens Healthcare Diagnostics, Inc |
| Feb 3, 2021 | Dimension Vista HDLC - colorimetric method, lipoproteins - Product Usage: use... | Siemens Healthcare Diagnostics Inc. has confirmed the potential for erroneous results due to disc... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Dec 2, 2020 | Atellica IM 1600 Analyzer - In vitro diagnostic testing of clinical specimens... | Multiple issues in Software v1.23.2 and lower-may lead to the reporting of discrepant patient res... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Dec 2, 2020 | Atellica CH 930 Analyzer - In vitro diagnostic testing of clinical specimens.... | Multiple issues in Software v1.23.2 and lower-may lead to the reporting of discrepant patient res... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Dec 2, 2020 | Atellica IM 1300 Analyzer - In vitro diagnostic testing of clinical specimens... | Multiple issues in Software v1.23.2 and lower-may lead to the reporting of discrepant patient res... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Nov 23, 2020 | ADVIA Gamma-Glutamyl Transferase (GGT) Reagent - Product Usage: For in vitro ... | Gamma-Glutamyl Transferase (GGT) Reagent Low End Repeatability-Imprecision. Reagent may demonst... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Nov 23, 2020 | Atellica Gamma-Glutamyl Transferase (GGT) Reagent - Product Usage: For in vit... | Gamma-Glutamyl Transferase (GGT) Reagent Low End Repeatability-Imprecision. Reagent may demonst... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Oct 26, 2020 | Atellica IM 1600 Analyzer - Product Usage: automated, immunoassay analyzers d... | Test Definition scanning may reset custom settings to defaults. When a 2D Master Curve and TDef b... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Oct 26, 2020 | Atellica IM 1300 Analyzer - Product Usage: automated, immunoassay analyzers d... | Test Definition scanning may reset custom settings to defaults causing falsely low or high result... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Sep 22, 2020 | Atellica CH Systems Fructosamine (FRUC)-For in vitro diagnostic use in the qu... | Fructosamine QC and patient samples may exhibit a 60 umol/L positive bias across the measuring ra... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Sep 22, 2020 | ADVIA Chemistry Systems-Fructosamine (FRUC)-For in vitro diagnostic use in th... | Fructosamine QC and patient samples may exhibit a 60 umol/L positive bias across the measuring ra... | Class II | Siemens Healthcare Diagnostics, Inc. |