Centaur Syphilis Assay-SYP Kit, 200 Test - Product Usage: For in vitro diagnostic use in the qual...
FDA Device Recall #Z-1674-2021 — Class II — April 21, 2021
Recall Summary
| Recall Number | Z-1674-2021 |
| Classification | Class II — Moderate risk |
| Date Initiated | April 21, 2021 |
| Status | Ongoing |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Siemens Healthcare Diagnostics, Inc |
| Location | East Walpole, MA |
| Product Type | Devices |
| Quantity | 22550 units |
Product Description
Centaur Syphilis Assay-SYP Kit, 200 Test - Product Usage: For in vitro diagnostic use in the qualitative determination of antibodies to Treponema pallidum in human serum or plasma using the ADVIA Centaur XP/XPT System and ADVIA Centaur CP Systems, SMN:10492493
Reason for Recall
Potential for the ADVIA Centaur Syphilis assay to carryover into other commercial assays. This carryover effect only occurs when the impacted assays are used immediately following testing with the Syphilis assay on the ADVIA Centaur XP/XPT or ADVIA Centaur CP systems
Distribution Pattern
Worldwide distribution - US Nationwide distribution and the countries of Austria, Belgium, Burkina Faso, Croatia, Czech Republic, Egypt, Finland, France, Fren.Polynesia, Georgia, Germany, Germany, Greece, Guadeloupe, Italy, Kazakhstan, Latvia, Netherlands, Norway, Poland, Portugal, Qatar, Russian Fed., Serbia, Slovakia, Spain, Sweden, Switzerland, Turkey, U.A.E., Ukraine, United Kingdom.
Lot / Code Information
All Lots
Other Recalls from Siemens Healthcare Diagnostics, Inc
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-0125-2022 | Class II | Atellica IM 1600 Analyzer, SMN 11066000 | Sep 28, 2021 |
| Z-0124-2022 | Class II | Atellica IM 1300 Analyzer, SMN 11066001 | Sep 28, 2021 |
| Z-0196-2022 | Class II | ADVIA Centaur SARS-CoV-2 Antigen Assay (100 Tes... | Sep 22, 2021 |
| Z-0195-2022 | Class II | Atellica IM SARS-CoV-2 Antigen Assay (100 Test ... | Sep 22, 2021 |
| Z-2426-2021 | Class II | Atellica IM BR 27.29 (BR) Assay 250 Test Kit (R... | Jul 6, 2021 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.