| Date | Product | Reason | Class | Firm |
|---|---|---|---|---|
| Nov 25, 2016 | Brilliance iCT Product Usage: Computed Tomography X-ray systems intended... | Failure to correctly document the installation of four M12 Bolts into the system rotor. | Class II | Philips Medical Systems (Cleveland) Inc |
| Nov 18, 2016 | FV-RF (Clinix RF Plus) Product Usage: A Stationary X-ray System, a device... | There's a potential for liquid penetration into the back side of the front control panel of the X... | Class II | Philips Medical Systems (Cleveland) Inc |
| Nov 14, 2016 | IQon Spectral CT is a Computed Tomography X-Ray System intended to produce cr... | A software defect is causing issues with the IQon Spectral CT and Philips Spectral CT Application... | Class II | Philips Medical Systems (Cleveland) Inc |
| Nov 14, 2016 | Philips Spectral CT Applications supports viewing and analysis of images at e... | A software defect is causing issues with the IQon Spectral CT and Philips Spectral CT Application... | Class II | Philips Medical Systems (Cleveland) Inc |
| Nov 12, 2016 | Brilliance 64 728232 Computed Tomography X-ray systems intended to produce c... | Software error due to the filament on timer | Class II | Philips Medical Systems (Cleveland) Inc |
| Nov 12, 2016 | Ingenuity Core 128 728323 | Software error due to the filament on timer | Class II | Philips Medical Systems (Cleveland) Inc |
| Nov 12, 2016 | Ingenuity CT 728326 Computed Tomography X-ray systems intended to produce cr... | Software error due to the filament on timer | Class II | Philips Medical Systems (Cleveland) Inc |
| Nov 12, 2016 | Ingenuity Core 728321 Computed Tomography X-ray systems intended to produce ... | Software error due to the filament on timer | Class II | Philips Medical Systems (Cleveland) Inc |
| Nov 12, 2016 | Brilliance 64 728231 Computed Tomography X-ray systems intended to produce c... | Software error due to the filament on timer | Class II | Philips Medical Systems (Cleveland) Inc |
| Oct 13, 2016 | Ingenuity CT Computed Tomography X-ray system, Model 728326 | The product label does not include the correct current rating. | Class III | Philips Medical Systems (Cleveland) Inc |
| Oct 13, 2016 | Ingenuity Core 128 Computed Tomography X-ray system, Model 728323 | The product label does not include the correct current rating. | Class III | Philips Medical Systems (Cleveland) Inc |
| Oct 13, 2016 | IQon Spectral CT Computed Tomography X-ray system, Model 728332 | The product label does not include the correct current rating. | Class III | Philips Medical Systems (Cleveland) Inc |
| Oct 13, 2016 | Brilliance ICT SP Computed Tomography X-ray system, Model 728311 | The product label does not include the correct current rating. | Class III | Philips Medical Systems (Cleveland) Inc |
| Oct 13, 2016 | Ingenuity Core Computed Tomography X-ray system, Model 728321 | The product label does not include the correct current rating. | Class III | Philips Medical Systems (Cleveland) Inc |
| Oct 13, 2016 | Brilliance iCT Computed Tomography X-ray system, Model 728306 | The product label does not include the correct current rating. | Class III | Philips Medical Systems (Cleveland) Inc |
| Jul 20, 2016 | Ingenuity Core128, Model number 728323 Product Usage: Computed Tomography... | After upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical... | Class II | Philips Medical Systems (Cleveland) Inc |
| Jul 20, 2016 | Ingenuity Core, Model number 728321 Product Usage: Computed Tomography X-... | After upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical... | Class II | Philips Medical Systems (Cleveland) Inc |
| Jul 20, 2016 | Brilliance 16, Model number 728246 Product Usage: Computed Tomography X-r... | After upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical... | Class II | Philips Medical Systems (Cleveland) Inc |
| Jul 20, 2016 | Brilliance BigBore Radiology CT Model number 728244 Product Usage: Comput... | After upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical... | Class II | Philips Medical Systems (Cleveland) Inc |
| Jul 20, 2016 | Brilliance CT Big Bore Oncology, Model number 728243 Product Usage: Compu... | After upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical... | Class II | Philips Medical Systems (Cleveland) Inc |
| Jul 20, 2016 | Brilliance 64 CT Model number 728231 Product Usage: Computed Tomography X... | After upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical... | Class II | Philips Medical Systems (Cleveland) Inc |
| Jul 20, 2016 | Brilliance CT 16 Power, Model number 728240 Product Usage: Computed Tomog... | After upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical... | Class II | Philips Medical Systems (Cleveland) Inc |
| Jun 21, 2016 | Ingenuity Flex, model # 728317; Intended to produce cross- sectional images ... | Supplier design change contributed to a thinner plastic which was used as part of the cover on th... | Class II | Philips Medical Systems (Cleveland) Inc |
| Jun 21, 2016 | Ingenuity Core, model # 728326; Intended to produce cross- sectional images ... | Supplier design change contributed to a thinner plastic which was used as part of the cover on th... | Class II | Philips Medical Systems (Cleveland) Inc |
| Jun 21, 2016 | Ingenuity Core 128, model #728323; Intended to produce cross- sectional imag... | Supplier design change contributed to a thinner plastic which was used as part of the cover on th... | Class II | Philips Medical Systems (Cleveland) Inc |
| Jun 21, 2016 | Brilliance CT Big Bore Oncology, model #728243; Intended to produce cross- se... | Supplier design change contributed to a thinner plastic which was used as part of the cover on th... | Class II | Philips Medical Systems (Cleveland) Inc |
| Jun 21, 2016 | Brilliance iCT model #728306; Intended to produce cross- sectional images of... | Supplier design change contributed to a thinner plastic which was used as part of the cover on th... | Class II | Philips Medical Systems (Cleveland) Inc |
| Jun 14, 2016 | Big Bore Oncology tomography X-ray system scanner, model number 728243. The... | Socket Head Cap Screws (SHCS) used in the Z-Axis Adjustment Plate were found to be made of stainl... | Class II | Philips Medical Systems (Cleveland) Inc |
| Jun 14, 2016 | Big Bore Radiology tomography X-ray system scanner, model number 728244. Th... | Socket Head Cap Screws (SHCS) used in the Z-Axis Adjustment Plate were found to be made of stainl... | Class II | Philips Medical Systems (Cleveland) Inc |
| May 25, 2016 | BrightView X, model #882478; gamma camera designed for single or dual detecto... | The firm became aware of an issue where during a non-Auto Body Contouring (ABC), non-circular, re... | Class II | Philips Medical Systems (Cleveland) Inc |
| May 25, 2016 | BrightView XCT Upgrade Model 882454 | The firm became aware of an issue where during a non-Auto Body Contouring (ABC), non-circular, re... | Class II | Philips Medical Systems (Cleveland) Inc |
| May 25, 2016 | BrightView XCT Model 882482, gamma camera for Single Photon Emission Computed... | The firm became aware of an issue where during a non-Auto Body Contouring (ABC), non-circular, re... | Class II | Philips Medical Systems (Cleveland) Inc |
| Apr 29, 2016 | Ingenuity Core128 Model No. 728323; To produce cross-sectional images of the... | Software issues in v4.1.3/4.1.4/4.1.5 in the Philips Ingenuity CT products that could affect the ... | Class II | Philips Medical Systems (Cleveland) Inc |
| Apr 29, 2016 | Brilliance iCT, Model No. 728306; To produce cross-sectional images of the b... | Software issues in versions v4.1 .3/4.1.5 in the Philips Brilliance iCT) iCT SP products that, co... | Class II | Philips Medical Systems (Cleveland) Inc |
| Apr 29, 2016 | Ingenuity CT Model No. 728326; To produce cross-sectional images of the body. | Software issues in v4.1.3/4.1.4/4.1.5 in the Philips Ingenuity CT products that could affect the ... | Class II | Philips Medical Systems (Cleveland) Inc |
| Apr 29, 2016 | Ingenuity Core Model No. 728321; To produce cross-sectional images of the body. | Software issues in software versions v4.1.3/4.1.4/4.1.5 in the Philips Brilliance 64 and Ingenuit... | Class II | Philips Medical Systems (Cleveland) Inc |
| Apr 29, 2016 | Brilliance iCT SP Model No. 728311; To produce cross-sectional images of the ... | Software issues found in v4.1 .3/4.1.5 in the Philips Brilliance iCT/ iCT SP products that could... | Class II | Philips Medical Systems (Cleveland) Inc |
| Apr 29, 2016 | Brilliance 64 CT Model 728231; To produce cross-sectional images of the body. | Software issues found in software versions v4.1.3/4.1.4/4.1.5 in the Philips Brilliance 64 produ... | Class II | Philips Medical Systems (Cleveland) Inc |
| Apr 20, 2016 | MX 16-slice whole body computed tomography X-ray system. Imaging diagnostic t... | The following issues have been found in MX 16-slice systems with software version 1.1.4.21426: ... | Class II | Philips Medical Systems (Cleveland) Inc |
| Apr 20, 2016 | MX 16-slice SKD whole body computed tomography X-ray system. Imaging diagnos... | The following issues have been found in MX 16-slice systems with software version 1.1.4.21426: ... | Class II | Philips Medical Systems (Cleveland) Inc |
| Apr 8, 2016 | Ingenuity Computed Tomography X-ray Systems Product Usage: Intended to pr... | The firm became aware of a problem where the system may not map Varian drive after CT user logout... | Class II | Philips Medical Systems (Cleveland) Inc |
| Apr 8, 2016 | Brilliance 64 Computed Tomography X-ray Systems Product Usage: Intended t... | The firm became aware of a problem where the system may not map Varian drive after CT user logout... | Class II | Philips Medical Systems (Cleveland) Inc |
| Apr 8, 2016 | Ingenuity Core Computed Tomography X-ray Systems, Product Usage: Inten... | The firm became aware of a problem where the system may not map Varian drive after CT user logout... | Class II | Philips Medical Systems (Cleveland) Inc |
| Apr 8, 2016 | Ingenuity Core 128 Computed Tomography X-ray Systems Product Usage: Inten... | The firm became aware of a problem where the system may not map Varian drive after CT user logout... | Class II | Philips Medical Systems (Cleveland) Inc |
| Apr 1, 2016 | Ingenuity Core-Computed Tomography X-ray system Product Usage: The Ingenu... | Philips Healthcare received reports from the field that certain Ingenuity Core systems running so... | Class II | Philips Medical Systems (Cleveland) Inc |
| Apr 1, 2016 | Ingenuity Core 128-Computed Tomography X-ray system Product Usage: The In... | Philips Healthcare received reports from the field that certain Ingenuity Core128 systems running... | Class II | Philips Medical Systems (Cleveland) Inc |
| Apr 1, 2016 | Brilliance 64, Computed Tomography X-ray system Product Usage: The Brilli... | Philips Healthcare received reports from the field that certain Brilliance 64 systems running sof... | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 24, 2016 | Ingenuity Core Computed Tomography X-ray systems is intended to produce cross... | Software defect causing intermittently slow response of Host. | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 24, 2016 | Brilliance CT 16 Power Computed Tomography X-ray systems is intended to produ... | Software defect causing intermittently slow response of Host. | Class II | Philips Medical Systems (Cleveland) Inc |
| Mar 24, 2016 | Brilliance 64 Computed Tomography X-ray systems is intended to produce cross-... | Software defect causing intermittently slow response of Host. | Class II | Philips Medical Systems (Cleveland) Inc |