| Date | Product | Reason | Class | Firm |
|---|---|---|---|---|
| Jul 17, 2019 | Kit BD Max ExK DNA 1 EU LUO; Catalog # 442818 | The foil bags containing extraction tubes for BD MAX" Reagent products may not been sealed properly. | Class II | Becton Dickinson & Co. |
| Jul 17, 2019 | Kit BD Max ExK TNA 2; Catalog # 442825 | The foil bags containing extraction tubes for BD MAX" Reagent products may not been sealed properly. | Class II | Becton Dickinson & Co. |
| Jul 17, 2019 | Kit BD Max Enteric Viral Panel EU Catalog # 443985 | The foil bags containing extraction tubes for BD MAX" Reagent products may not been sealed properly. | Class II | Becton Dickinson & Co. |
| Jul 17, 2019 | Kit BD Max Check-Points CPO IVD EU; Catalog # 278102 | The foil bags containing extraction tubes for BD MAX" Reagent products may not been sealed properly. | Class II | Becton Dickinson & Co. |
| Jul 17, 2019 | Kit EXT Enteric Bacterial Panel; Catalog # 443812 | The foil bags containing extraction tubes for BD MAX" Reagent products may not been sealed properly. | Class II | Becton Dickinson & Co. |
| Jul 17, 2019 | Kit BD Max ExK DNA 1 USA; Catalog # 442817 | The foil bags containing extraction tubes for BD MAX" Reagent products may not been sealed properly. | Class II | Becton Dickinson & Co. |
| Jul 17, 2019 | Kit BD Max StaphSR; Catalog # 443418 | The foil bags containing extraction tubes for BD MAX" Reagent products may not been sealed properly. | Class II | Becton Dickinson & Co. |
| Jul 17, 2019 | Kit BD Max Enteric Viral Panel RUO; Catalog # 443715 | The foil bags containing extraction tubes for BD MAX" Reagent products may not been sealed properly. | Class II | Becton Dickinson & Co. |
| Apr 3, 2019 | BD Veritor Plus Analyzer- In-Vitro Diagnostic intended to use with the BD Ve... | Background calibration has the potential to cause false positive or false negative test results. ... | Class II | Becton Dickinson & Co. |
| Apr 3, 2019 | BD Veritor System Reader -In Vitro Diagnostic for use with BD Veritor System ... | Background calibration has the potential to cause false positive or false negative test results. ... | Class II | Becton Dickinson & Co. |
| Feb 21, 2019 | BD Synapsys Laboratory Solutions Catalog Number:444150 Product Usage: B... | BD Synapsys version 2.1 allowed the potential for a test order to be associated to an incorrect c... | Class II | Becton Dickinson & Co. |
| Jan 14, 2019 | BD MAX Enteric Bacterial Panel -For In Vitro Diagnostic Use with the BD MAX S... | False negative results with the BD MAX Extended Enteric Bacterial Panel when testing specifically... | Class II | Becton Dickinson & Co. |
| Dec 20, 2018 | Phoenix AST-s Indicator, Catalog Number 246009 | Panels inoculated using certain lots of the indicator solution are demonstrating an increased occ... | Class II | Becton Dickinson & Co. |
| Dec 20, 2018 | Phoenix AP AST Indicator Bag, Catalog Number 246006 | Panels inoculated using certain lots of the indicator solution are demonstrating an increased occ... | Class II | Becton Dickinson & Co. |
| Dec 20, 2018 | Phoenix AST Indicator, Catalog Number 246004 | Panels inoculated using certain lots of the indicator solution are demonstrating an increased occ... | Class II | Becton Dickinson & Co. |
| Oct 22, 2018 | BD BBL Trypticase Soy Agar with 5% Sheep Blood, Catalog No. 221261, (non-sele... | a portion of a single lot was not manufactured in accordance with the internal formulation result... | Class II | Becton Dickinson & Co. |
| Sep 24, 2018 | BD Max Instrument Reader Spare Part, Catalog Number 8090369/443364, 44336409 ... | A trend for false positive Candida glabrata results has been confirmed. The discrepant results i... | Class II | Becton Dickinson & Co. |
| Sep 24, 2018 | BD Max Instrument, Catalog Number 441916 441927 (refurbished instrument) - Pr... | A trend for false positive Candida glabrata results has been confirmed. The discrepant results i... | Class II | Becton Dickinson & Co. |
| Sep 21, 2018 | BD FocalPoint Slide Profiler and BD FocalPoint GS Review Station; Catalog Nos... | The device may not function at the lower end (<15 deg. C) of the labeled operating temperatures. ... | Class III | Becton Dickinson & Co. |
| Jun 21, 2018 | BD BBL GC-Lect Agar; Cat. No. 297715 The device is a selective medium pro... | A portion of this lot was manufactured using Gentamicin instead of Vancomycin. Gentamicin is an i... | Class II | Becton Dickinson & Co. |
| Mar 2, 2018 | BD MAX DNA MMK Lab Use, catalog no. 442828 | The package insert incorrectly lists the volume of the primer and probe diluent to use for the re... | Class III | Becton Dickinson & Co. |
| Mar 2, 2018 | BD MAX DNA MMK (SPC) For Laboratory Use, catalog no. 442829 | The package insert incorrectly lists the volume of the primer and probe diluent to use for the re... | Class III | Becton Dickinson & Co. |
| Oct 3, 2017 | Difco" Agar Noble, 500g bottle Noble Agar is a solidifying agent that is e... | HighPuragar was mislabeled and inadvertently sold as Noble Agar. | Class II | Becton Dickinson & Co. |
| Jul 18, 2017 | BD MAX" CT/GC/TV. Catalog Number(s): 443824,442970,443590 (RUO).for colle... | BD recently notified customers in July 2017 regarding an increased rate of unresolved (UNR) and i... | Class II | Becton Dickinson & Co. |
| Jul 18, 2017 | BD MAX" UVE Specimen Collection Kit. Catalog Number(s): 443376 for collecti... | BD recently notified customers in July 2017 regarding an increased rate of unresolved (UNR) and i... | Class II | Becton Dickinson & Co. |
| Jul 18, 2017 | BD MAXTM Vaginal Panel. Catalog Number(s): 443712 443710 443711 (RUO)for col... | BD recently notified customers in July 2017 regarding an increased rate of unresolved (UNR) and i... | Class II | Becton Dickinson & Co. |
| Jun 1, 2017 | BD PhoenixTM AP, Catalog Number 448010 The BD Phoenix" AP instrument is desi... | Potential unexpected movement of robot arm | Class III | Becton Dickinson & Co. |
| May 22, 2017 | TrypticaseTM Soy Agar with 5% Sheep Blood (TSA IITM), Catalog Number 221261 ... | Listeria monocytogenes contamination of non-sterile plated media | Class II | Becton Dickinson & Co. |
| May 11, 2017 | BD Phoenix PMIC-107, Catalog Number 448417, UDI #s (01)30382904484171, (10)70... | BD has recently confirmed quality control failures, related to a manufacturing issue, for the ind... | Class II | Becton Dickinson & Co. |
| Jun 1, 2016 | PhoenixSpec Calibrator & Kit is used to calibrate the PhoenixSpec & Phoeni... | BD has determined that the BBL PhoenixSpec Calibrator kits referenced in this recall from the fol... | Class III | Becton Dickinson & Co. |
| Jun 1, 2016 | PhoenixSpec AP Calibrator is used to calibrate the PhoenixSpec & Phoenix ... | BD has determined that the PhoenixSpec AP Calibrator referenced in this recall from the following... | Class III | Becton Dickinson & Co. |
| Mar 28, 2016 | Trypticase TM Soy Agar with 5% Sheep Blood (TSA IITM) // MacConkey II Agar C... | Trypticase TM Soy Agar with 5% Sheep Blood (TSA IITM) // MacConkey II Agar is contaminated with L... | Class II | Becton Dickinson & Co. |
| Feb 12, 2016 | BD SurePathTM Preservative Collection Vials, catalog numbers 490505, 490527, ... | BD has confirmed that a portion of BD SurePath" collection vials associated with specific product... | Class II | Becton Dickinson & Co. |
| Aug 27, 2015 | BD AffirM VPIII Microbial Identification Test | BD has confirmed that a portion of tests associated with the affected lots( 446252 and 446257 ) o... | Class II | Becton Dickinson & Co. |
| Jun 29, 2015 | Vancomycin Screen Agar | The affected expired lots of product may have exhibited breakthrough growth of vancomycin suscep... | Class II | Becton Dickinson & Co. |
| Jun 29, 2015 | Enterococcus Screen Agar QUAD Plate | The affected expired lots of product may have exhibited breakthrough growth of vancomycin suscep... | Class II | Becton Dickinson & Co. |
| Mar 27, 2015 | BBL Trypticase Soy Agar with 5% Sheep Blood (TSA II) and Chocolate II Agar (G... | BD has confirmed that one lot of TSA II and Chocolate Agar I-Plates may fail to recover Haemophil... | Class III | Becton Dickinson & Co. |
| Nov 21, 2014 | BBL Port A Cul tubes, catalog number 221606. Patient sample collection tubes... | The products may exhibit indications of excessive oxidation. Excessive oxidation may affect the ... | Class II | Becton Dickinson & Co. |
| Nov 21, 2014 | BBL Port A Cul tubes with Swabs Sterile Pack, catalog number 221607. For pati... | The products may exhibit indications of excessive oxidation. Excessive oxidation may affect the ... | Class II | Becton Dickinson & Co. |
| Nov 7, 2014 | BD Viper LT System, catalog number 442839 when used in conjunction with the B... | Potential for false positive Chlamydia trachomatis (CT) results and Neisseria gonorrhoeae (GC) re... | Class II | Becton Dickinson & Co. |
| Aug 25, 2014 | EpiCenter Plus Microbiology Data Management System software version 6.20A, wh... | The information in the EpiCenter interface translation tables was not updated to match the Bruker... | Class II | Becton Dickinson & Co. |
| Jul 28, 2014 | BD BBL Vancomycin Screen Agar Catalog number 222204 Vancomycin Screen Agar... | These lots of Agar may have exhibited breakthrough growth of vancomycin susceptible Enterococcus ... | Class II | Becton Dickinson & Co. |
| Jul 23, 2014 | The BD MAX MRSA Assay, catalogue #442953. An automated qualitative in vitr... | The performance of the assay no longer conforms to this statement in the Analytical Specificity s... | Class II | Becton Dickinson & Co. |
| Jun 6, 2014 | CDC Anaerobe 5% Sheep Blood Agar CDC Anaerobe 5% Sheep Blood Agar is an en... | A low level of surface and subsurface contamination of Listeria monocytogenes on 9 lots of non-st... | Class II | Becton Dickinson & Co. |
| Jun 6, 2014 | Columbia Agar with 5% Sheep Blood Columbia Agar with 5% Sheep Blood is a g... | A low level of surface and subsurface contamination of Listeria monocytogenes on 9 lots of non-st... | Class II | Becton Dickinson & Co. |
| Jun 6, 2014 | TrypticaseTM Soy Agar with 5% Sheep Blood (TSA IITM) // MacConkey II Agar ... | A low level of surface and subsurface contamination of Listeria monocytogenes on 9 lots of non-st... | Class II | Becton Dickinson & Co. |
| May 21, 2014 | BBL CrystaLEAD Enteric/Nonfermentor (E/NF) Identification (ID) System, catal... | A portion of the BBL Crystal Enteric/Nonfermentor (E/NF) Identification kits from lot numbers 322... | Class III | Becton Dickinson & Co. |
| Dec 9, 2013 | BD BACTEC FX- Top Unit instrument, catalog number 441385, and Bottom Unit ins... | Improperly functioning component of a diagnostic medical device may cause false negative results ... | Class II | Becton Dickinson & Co. |
| Dec 9, 2013 | BD BACTEC FX 40 instrument, catalog number 442296, manufactured by BD Diagnos... | Improperly functioning component of a diagnostic medical device may cause false negative results ... | Class II | Becton Dickinson & Co. |
| Oct 11, 2013 | BD GeneOhm Cdiff Assay, Part #441401, containing Lysis Kit sample buffer tube... | An incorrect reagent was packed with an in vitro diagnostic kit, which may cause an increase in u... | Class II | Becton Dickinson & Co. |