Browse Device Recalls
2,601 FDA medical device safety recalls.
FDA Medical Device Recall Enforcement Database
Browse 2,601 FDA medical device recall enforcement actions. Each entry includes the device description, reason for recall, classification (Class I through III based on severity), recalling firm, and distribution details. Use the filters below to search by year, classification, state, or keyword.
Medical Device Recalls
Browse 2,601 FDA device recalls in 2023.
| Date | Product | Reason | Class | Firm |
|---|---|---|---|---|
| Sep 11, 2023 | Disposable Spetzler-Malis Bipolar forceps, 20CM SLM DISP BAYONET 1.0MM, Catal... | An error was identified on the expiration date of the product label. The shelf life for the produ... | Class II | Stryker Corporation |
| Sep 11, 2023 | Disposable Spetzler-Malis Bipolar forceps, 23CM SM DISP BAYONET 1.0MM TIP, Ca... | An error was identified on the expiration date of the product label. The shelf life for the produ... | Class II | Stryker Corporation |
| Sep 11, 2023 | Disposable Spetzler-Malis Bipolar forceps, 23CM IM DISP BAYONET 0.5MM TIP, Ca... | An error was identified on the expiration date of the product label. The shelf life for the produ... | Class II | Stryker Corporation |
| Sep 11, 2023 | Disposable Spetzler-Malis Bipolar forceps, 20CM IM DISP BAYONET 1.5MM TIP, Ca... | An error was identified on the expiration date of the product label. The shelf life for the produ... | Class II | Stryker Corporation |
| Sep 11, 2023 | Disposable Spetzler-Malis Bipolar forceps, 23CM SLM DISP BAYONET 1.5MM, Catal... | An error was identified on the expiration date of the product label. The shelf life for the produ... | Class II | Stryker Corporation |
| Sep 11, 2023 | Sapphire Infusion Pumps, Models: Multi-Therapy (REF 15031-000-0028), Epidural... | Infusion Pumps with affected software revision may fail to detect air in line, which may lead to ... | Class I | EITAN MEDICAL LTD |
| Sep 11, 2023 | Disposable Spetzler-Malis Bipolar forceps, 20CM IM DISP BAYONET 0.5MM TIP, Ca... | An error was identified on the expiration date of the product label. The shelf life for the produ... | Class II | Stryker Corporation |
| Sep 11, 2023 | Disposable Spetzler-Malis Bipolar forceps, 18CM SM DISP BAYONET 1.5MM TIP, Ca... | An error was identified on the expiration date of the product label. The shelf life for the produ... | Class II | Stryker Corporation |
| Sep 11, 2023 | GS0008 2Y - Burn Pack - STL, Item Number 880082025 | Medical convenience kit contains Zimmer Dermatome Blades which were recalled by Zimmer due to 38 ... | Class II | ROi CPS LLC |
| Sep 11, 2023 | Disposable Spetzler-Malis Bipolar forceps, 18CM SM DISP BAYONET 0.5MM TIP, Ca... | An error was identified on the expiration date of the product label. The shelf life for the produ... | Class II | Stryker Corporation |
| Sep 11, 2023 | Disposable Spetzler-Malis Bipolar forceps, 23CM SM DISP BAYONET 1.5MM TIP, Ca... | An error was identified on the expiration date of the product label. The shelf life for the produ... | Class II | Stryker Corporation |
| Sep 8, 2023 | Centricity Universal Viewer Workflow Manager, Image processing radiological s... | When using Universal Viewer or Centricity Universal Viewer with Workflow Manager (WFM), if a thir... | Class II | GE Healthcare |
| Sep 8, 2023 | Omega Medical Imaging Elevating Monitor Suspension Systems, Part Number 1000-... | An actuator separated from the pivot mechanism on a lower monitor boom system column. The lower b... | Class II | Omega Medical Imaging, Inc. |
| Sep 8, 2023 | F2-01 Frame, physiological patient monitor (with arrhythmia detection or alarms) | There is a potential interruption of data communication between E-modules inserted in the F2-01 F... | Class II | GE Medical Systems China Co., Ltd. |
| Sep 8, 2023 | Universal Viewer Workflow Manager, Image processing radiological system | When using Universal Viewer or Centricity Universal Viewer with Workflow Manager (WFM), if a thir... | Class II | GE Healthcare |
| Sep 7, 2023 | Alcon Custom-Pak, Surgical Procedure Pack REF 14057-09, 45K W/DUOVISC CATARAC... | Custom surgical procedure packs contain latex gloves but are labeled as "latex free". | Class II | Alcon Research, LLC |
| Sep 7, 2023 | Real Intelligence CORI RI.KNEE v2.0 TKA (Total Knee Arthroplasty) software ve... | CORI software was missing a planning stage that appears in cases of significant knee deformation ... | Class II | Blue Belt Technologies, Inc |
| Sep 7, 2023 | Fogarty Thru-Lumen Embolectomy Catheter, REF: 12TLW805F35 | Inadvertent release of embolectomy catheters that may experience slower than normal balloon defla... | Class II | Edwards Lifesciences, LLC |
| Sep 6, 2023 | 2008T HD SYS. CDX BLUESTAR | Potential PCBA leaching from tubing of hemodialysis machines. | Class I | Fresenius Medical Care Holdings, Inc. |
| Sep 6, 2023 | Meera Mobile Operating Table-to support and position the patient immediately ... | Under certain conditions, an issue might prevent the device from performing as intended. A specif... | Class II | Getinge Usa Sales Inc |
| Sep 6, 2023 | BD Pyxis CII Safe ES, REF: 1116-00; and BD Pyxis : CII Safe ES Desktop PC, Wi... | When global edit is used to update multiple formulary properties simultaneously, the following pr... | Class II | CareFusion 303, Inc. |
| Sep 6, 2023 | RX Daytona Plus (without ISE)- IVD Clinical Chemistry Analyser (1) RX4041, (... | RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not mee... | Class II | Randox Laboratories Ltd. |
| Sep 6, 2023 | AU/DxC AU Chemistry Transferrin REF: OSR6152 Immune complexes formed in solu... | The firm has identified that the Transferrin reagent does not meet the Instructions for Use (IFU)... | Class II | Beckman Coulter Biomedical Division |
| Sep 6, 2023 | 2008T Hemodialysis System without CDX | Potential PCBA leaching from tubing of hemodialysis machines. | Class I | Fresenius Medical Care Holdings, Inc. |
| Sep 6, 2023 | 2008T HD SYS. CDX W/BIBAG BLUESTAR | Potential PCBA leaching from tubing of hemodialysis machines. | Class I | Fresenius Medical Care Holdings, Inc. |
| Sep 6, 2023 | 2008T GEN 2 Bibag without CDX | Potential PCBA leaching from tubing of hemodialysis machines. | Class I | Fresenius Medical Care Holdings, Inc. |
| Sep 6, 2023 | 2008T Hemodialysis System w/Bibag | Potential PCBA leaching from tubing of hemodialysis machines. | Class I | Fresenius Medical Care Holdings, Inc. |
| Sep 6, 2023 | RX Daytona Plus (with ISE)-IVD Clinical Chemistry Analyser Catalog Number: R... | RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not mee... | Class II | Randox Laboratories Ltd. |
| Sep 6, 2023 | 2008T HD SYS. W/O CDX W/BIBAG BLUESTAR | Potential PCBA leaching from tubing of hemodialysis machines. | Class I | Fresenius Medical Care Holdings, Inc. |
| Sep 6, 2023 | RX Imola-IVD Clinical Chemistry Analyzer Catalog Number: RX4900. RX4900R (re... | RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not mee... | Class II | Randox Laboratories Ltd. |
| Sep 6, 2023 | Portex Pro-Vent Arterial Blood Sampling Kit with Dry Lithium Heparin for Gase... | The complaint involved the absence of a component called "ASM Filter-Pro ABG SYR CAP" from the pa... | Class II | Smiths Medical ASD Inc. |
| Sep 6, 2023 | RX Monaco-IVD Clinical Chemistry Analyser Catalog Number: RX5000 | RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not mee... | Class II | Randox Laboratories Ltd. |
| Sep 6, 2023 | 2008T Hemodialysis SYS, with CDX | Potential PCBA leaching from tubing of hemodialysis machines. | Class I | Fresenius Medical Care Holdings, Inc. |
| Sep 6, 2023 | 2008T HD SYS. W/O CDX BLUESTAR | Potential PCBA leaching from tubing of hemodialysis machines. | Class I | Fresenius Medical Care Holdings, Inc. |
| Sep 6, 2023 | RX Modena-IVD Clinical Chemistry Analyser Catalog Number: RX9000 | RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not mee... | Class II | Randox Laboratories Ltd. |
| Sep 6, 2023 | RSP Glenoid Baseplate, 30mm P2 Porous Coated REF 508-32-204. DJO Surgical sho... | Coating hex did not engage with the Straight Mod Hexdriver or with the Hex Head Screw Driver | Class II | Encore Medical, LP |
| Sep 6, 2023 | Venue Go Standard Carts Ref: (H45181VC and H45103VCW), used with Venue Go R2,... | Some Venue Go Standard Carts can develop an internal failure of the vertical/horizontal tilt adju... | Class II | GE MEDICAL SYSTEMS, ISRAEL LTD. |
| Sep 5, 2023 | Zimmer Dermatome Blades- 10-pack boxes of blades. Intended for use with the d... | Skin grafts thin and non-uniform when using the affected blades. The issue would be identified at... | Class II | Zimmer Surgical Inc |
| Sep 5, 2023 | Atrium Oasis Dry Suction Water Seal Chest Drain, Part Number: 3600-100, DRAIN... | Product was re-processed and re-sterilized by a third party using packaging configurations and a ... | Class II | Atrium Medical Corporation |
| Sep 5, 2023 | Atrium Oasis Dry Suction Water Seal Chest Drain, Part Number: 3650-100, DRAIN... | Product was re-processed and re-sterilized by a third party using packaging configurations and a ... | Class II | Atrium Medical Corporation |
| Sep 5, 2023 | MAZOR X robotic guidance system, Model TPL0059, with software versions 4.2.2 ... | Software update | Class II | Mazor Robotics Ltd |
| Sep 1, 2023 | Segmental Stem, Canal Filling, Bowed 152MM, 1 EACH. Model Numbers: FB-11152-0... | Potential for breach of Tyvek seals. | Class II | Onkos Surgical, Inc. |
| Sep 1, 2023 | Male-Female Midsection with the following sizes and model numbers. a. Male... | Potential for breach of Tyvek seals. | Class II | Onkos Surgical, Inc. |
| Sep 1, 2023 | Segmental Stem, Cemented, Bowed, 152MM, 1 EACH. Model Numbers: CB-11152-03M, ... | Potential for breach of Tyvek seals. | Class II | Onkos Surgical, Inc. |
| Sep 1, 2023 | General Pack, REF BBGP31C | During an internal investigation, ACS identified that the components, I.V. Cath (part number 4252... | Class II | American Contract Systems, Inc. |
| Sep 1, 2023 | Pediatric Urology Wolson Pack, REF BPPU55A | During an internal investigation, ACS identified that the components, I.V. Cath (part number 4252... | Class II | American Contract Systems, Inc. |
| Sep 1, 2023 | Breast Plastics Pack, REF BBPL20D | During an internal investigation, ACS identified that the components, I.V. Cath (part number 4252... | Class II | American Contract Systems, Inc. |
| Sep 1, 2023 | Sub Q Recorder Pack, REF BPRC87A | During an internal investigation, ACS identified that the components, I.V. Cath (part number 4252... | Class II | American Contract Systems, Inc. |
| Sep 1, 2023 | (1) Over-ear headphones used with the MAGNETOM Amira MAGNETOM Free.Star MAGNE... | In some cases, when lubricating grease is present within the headband of the over-ear headphones... | Class II | Siemens Medical Solutions USA, Inc |
| Sep 1, 2023 | Robotics Pack, REF SSRO22D | During an internal investigation, ACS identified that the components, I.V. Cath (part number 4252... | Class II | American Contract Systems, Inc. |