DateProductReasonClassFirm
Aug 28, 2014 Dimension Vista¿ Mass CKMB Isoenzyme Calibrator (MMB CAL) an in vitro diagnos... Siemens Healthcare Diagnostics has determined that Dimension Vista® Mass CKMB Isoenzyme Calib... Class II Siemens Healthcare Diagnostics, Inc.
Aug 19, 2014 Siemens CentraLink Data Management System, software version 14. The Cent... Siemens Healthcare Diagnostics has identified a task scheduler issue with the CentraLink" Da... Class II Siemens Healthcare Diagnostics, Inc.
Aug 18, 2014 Siemens Dimension Vista¿ CHEM 1 CAL, lot 3GM081. The CHEM 1 CAL is an in v... Siemens Healthcare Diagnostics has received complaints for low recovery of Magnesium (MG) Quality... Class II Siemens Healthcare Diagnostics, Inc.
Aug 14, 2014 Dimension Vista¿ HbA1c In March, Siemens Healthcare Diagnostics issued an Urgent Medical Device Correction for decreased... Class II Siemens Healthcare Diagnostics, Inc.
Jul 16, 2014 Siemens Dimension EXL Integrated Chemistry System in vitro diagnostic devices... The Reagent Lid hinge may lose its effectiveness and slowly shift downward during maintenance pro... Class II Siemens Healthcare Diagnostics, Inc.
Jul 15, 2014 Siemens Dimension Chemistry Wash (RD701) is an accessory of the flex cartridg... Siemens Healthcare Diagnostics has received complaints for the Dimension Chemistry Wash (RD701) o... Class II Siemens Healthcare Diagnostics, Inc.
Jun 23, 2014 SIEMENS ADVIA Centaur CP Immunoassay System BNP kit; SMN 10309045 - 500 test ... Complaints of failed calibration. Class II Siemens Healthcare Diagnostics, Inc
Jun 23, 2014 SIEMENS ADVIA Centaur CP Immunoassay System BNP kit; SMN 10309044 - 100 test ... Complaints of failed calibration. Class II Siemens Healthcare Diagnostics, Inc
Jun 2, 2014 Siemens Healthcare ADVIA Acid/Base Reagent 1 and 2 Kit containing ADVIA Centa... Swollen acid bottles due to a trace metal contamination in the Acid Reagent bottles Class II Siemens Healthcare Diagnostics, Inc
Jun 2, 2014 MicroScan WalkAway-40 plus Instrument and MicroScan Walkaway-96 plus instrume... Springs contained in the access door hinge assembly on the Walk Away Plus instruments are becomin... Class II Siemens Healthcare Diagnostics, Inc.
May 30, 2014 ADVIA Centaur¿ Systems TnI-Ultra assay; Catalog Number 02789602 (100 tests), ... The solid phase reagent in some of the ADVIA Centaur¿ TnI -Ultra ReadyPacks¿, lot 084, is darker ... Class II Siemens Healthcare Diagnostics, Inc
May 27, 2014 Dimension¿ Integrated Chemistry Systems intended to measure a variety of anal... There is a misassembly issue with the photometer installed on Dimension¿ Integrated Chemistry Sy... Class III Siemens Healthcare Diagnostics, Inc.
May 20, 2014 syngo Lab Data Manager, SMN 10800057, 10803189, and 10804573. Certain configuration parameters may cause the reference ranges to not hold results as expected a... Class II Siemens Healthcare Diagnostics, Inc.
Apr 14, 2014 Siemens Healthcare Diagnostics Vista Air Compressor. Vista AirCompressor ... A small number of Vista Air Compressors were miswired, which if installed and powered up, could c... Class II Siemens Healthcare Diagnostics, Inc.
Apr 10, 2014 ADVIA Centaur Systems VitD 100 test, Ready Pack, Catalog Number: 10491994. ... Samples containing fluorescein may show interference with the ADVIA Centaur¿ Systems TSH3 Ultra, ... Class II Siemens Healthcare Diagnostics, Inc
Apr 10, 2014 ADVIA Centaur Systems TSH3 Ultra 2500 tests Catalog Number: 04862625; Siemen... Samples containing fluorescein may show interference with the ADVIA Centaur¿ Systems TSH3 Ultra, ... Class II Siemens Healthcare Diagnostics, Inc
Apr 10, 2014 ADVIA Centaur Systems TSH3 Ultra Ready Pack, 100 test Catalog Number: 06491... Samples containing fluorescein may show interference with the ADVIA Centaur¿ Systems TSH3 Ultra, ... Class II Siemens Healthcare Diagnostics, Inc
Apr 10, 2014 ADVIA Centaur Systems BRAHMS PCT 100 test Sold only OUS; Catalog Number: 103... Samples containing fluorescein may show interference with the ADVIA Centaur¿ Systems TSH3 Ultra, ... Class II Siemens Healthcare Diagnostics, Inc
Apr 10, 2014 ADVIA Centaur Systems TSH3 Ultra , Ready Pack, 500 test Catalog Number: 0649... Samples containing fluorescein may show interference with the ADVIA Centaur¿ Systems TSH3 Ultra, ... Class II Siemens Healthcare Diagnostics, Inc
Apr 10, 2014 ADVIA Centaur Systems VitD 500 test, Ready Pack, Catalog Number: 10631021.... Samples containing fluorescein may show interference with the ADVIA Centaur¿ Systems TSH3 Ultra, ... Class II Siemens Healthcare Diagnostics, Inc
Apr 10, 2014 ADVIA Centaur Systems VitD Total 100 test, Ready Pack, For OUS; SMN 1069920... Samples containing fluorescein may show interference with the ADVIA Centaur¿ Systems TSH3 Ultra, ... Class II Siemens Healthcare Diagnostics, Inc
Mar 25, 2014 Siemens ADVIA Centaur Folate Cat No. 06891541 SMN 10325366 (500 tests) Cat... Serum controls may go out of range low with the ADVIA Centaur ¿Systems Folate assay and a shift i... Class II Siemens Healthcare Diagnostics, Inc
Mar 19, 2014 Dimension Vista¿ CSA and CSAE The Cyclosporine method is an in vitro diagn... A complaint was received of Dimension Vista¿ CSA and CSAE under-recovery of cyclosporine when pat... Class II Siemens Healthcare Diagnostics, Inc.
Jan 6, 2014 Dimension Vista Enzyme 1 Calibrator (ENZ 1 CAL) lots 3FD034 and 3HD012. An... Complaints were received regarding Quality Control (QC) shifts with Pseudocholinesterase (PCHE) a... Class II Siemens Healthcare Diagnostics, Inc.
Dec 18, 2013 ADVIA Centaur¿ Systems Phenytoin Calibrator N ( PHTN) 6-pack SMN 10314977 ... ADVIA Centaur¿ Systems Phenytoin Calibrator N, lots CN58 and higher, change in recovery observe... Class II Siemens Healthcare Diagnostics, Inc
Dec 18, 2013 ADVIA Centaur¿ Systems Phenytoin Calibrator N ( PHTN) 2-pack SMN 10311391 ... ADVIA Centaur¿ Systems Phenytoin Calibrator N, lots CN58 and higher, change in recovery observe... Class II Siemens Healthcare Diagnostics, Inc
Dec 17, 2013 Siemens ADVIA 1200 and ADVIA 1650 Chemistry Systems. Automated, clinical c... An overflow flag //// is displayed in the result column for a test when the algorithm for the tes... Class II Siemens Healthcare Diagnostics, Inc.
Dec 5, 2013 Dimension Vista Cyclosporine Flex reagent cartridge an in vitro diagnostic... Complaints were received regarding variability in recovery of QC and patient results with certain... Class II Siemens Healthcare Diagnostics, Inc.
Nov 6, 2013 Siemens ADVIA Chemistry Systems models 1200, 1650, 1800 and 2400 Product Us... The ADVIA Chemistry software systems using certain software versions allows the system to automat... Class II Siemens Healthcare Diagnostics, Inc.
Oct 30, 2013 Siemens CentraLink" Data Management System V14.0.4, V14.0.5 or V14.0.8 The... Under extremely rare circumstances, a patient result that had been previously disabled may be rel... Class II Siemens Healthcare Diagnostics, Inc.
Oct 7, 2013 User Defined Method Flex Assignment/Siemens Dimension Vista 500 or Dimension ... When utilizing the Routine Inventory screen to enter a User Defined Method (EMPTY) Flex the syste... Class II Siemens Healthcare Diagnostics, Inc.
Sep 18, 2013 Siemens Dimension Vista Homocysteine Flex Reagent Cartridge The HCYS metho... Dimension Vista HCYS lot 13064MA shows erratic recovery of QC and patient samples. Class II Siemens Healthcare Diagnostics, Inc.
Sep 17, 2013 Siemens Dimension LOCI TNI Troponin I Calibrator RC 621. The LOCI TNI ... Siemens Dimension LOCI TNI Troponin I calibrator lot 3DD007 may demonstrate an average upward shi... Class II Siemens Healthcare Diagnostics, Inc.
Aug 30, 2013 Emit 2000 Phenobarbital Assay and Syva Emit 2000 Phenobarbital Assay Imprecision with certain lots Class II Siemens Healthcare Diagnostics, Inc.
Aug 21, 2013 Microscan rapID/S panels NC3.33 Sold in Japan only. Product Usage: Mi... Confirmed increase in false susceptible and false intermediate misreads for Imipenem and/or merop... Class I Siemens Healthcare Diagnostics, Inc.
Aug 21, 2013 Microscan plus Negative Combo 3 Sold outside of US only. Product Usage:... Confirmed increase in false susceptible and false intermediate misreads for Imipenem and/or merop... Class I Siemens Healthcare Diagnostics, Inc.
Aug 21, 2013 Microscan rapID/S panels NC3.12 Sold in Japan only. Product Usage: Mi... Confirmed increase in false susceptible and false intermediate misreads for Imipenem and/or merop... Class I Siemens Healthcare Diagnostics, Inc.
Aug 21, 2013 Microscan Synergies Plus Negative Urine Combo 1 Product Usage: MicroScan¿... Confirmed increase in false susceptible and false intermediate misreads for Imipenem and/or merop... Class I Siemens Healthcare Diagnostics, Inc.
Aug 21, 2013 Microscan Synergies Plus Negative Combo 2 Product Usage: MicroScan¿ Syne... Confirmed increase in false susceptible and false intermediate misreads for Imipenem and/or merop... Class I Siemens Healthcare Diagnostics, Inc.
Aug 21, 2013 Microscan plus Negative Urine Combo 4 Sold outside of US only. Product ... Confirmed increase in false susceptible and false intermediate misreads for Imipenem and/or merop... Class I Siemens Healthcare Diagnostics, Inc.
Aug 21, 2013 Microscan plus Negative Breakpoint Combo 4 Sold outside of US only. Pro... Confirmed increase in false susceptible and false intermediate misreads for Imipenem and/or merop... Class I Siemens Healthcare Diagnostics, Inc.
Aug 21, 2013 MicroscanrapID/S panels NC3.11 Sold in Japan only. Product Usage: Mic... Confirmed increase in false susceptible and false intermediate misreads for Imipenem and/or merop... Class I Siemens Healthcare Diagnostics, Inc.
Aug 21, 2013 Microscan Synergies Plus Negative Breakpoint Combo 7 Product Usage: Micro... Confirmed increase in false susceptible and false intermediate misreads for Imipenem and/or merop... Class I Siemens Healthcare Diagnostics, Inc.
Aug 21, 2013 Microscan Synergies Plus Negative Urine Combo 2 Product Usage: MicroScan¿... Confirmed increase in false susceptible and false intermediate misreads for Imipenem and/or merop... Class I Siemens Healthcare Diagnostics, Inc.
Aug 21, 2013 Microscan Synergies Plus Negative Urine Combo 5 Product Usage: MicroSca... Confirmed increase in false susceptible and false intermediate misreads for Imipenem and/or merop... Class I Siemens Healthcare Diagnostics, Inc.
Aug 16, 2013 ADVIA Centaur Intact PTH Calibrator (6 pack); Catalog Number: SMN 10492387 ... PTH Calibrator packaged with only high calibrator. Class III Siemens Healthcare Diagnostics, Inc
Aug 13, 2013 Siemens EasyLink Informatics System V4.0/V5 Siemens, Material Number (SMN) ... May not perform as intended under certain conditions, causing the release of results to the Labor... Class II Siemens Healthcare Diagnostics, Inc.
Jul 30, 2013 ADVIA Chemistry A1c_3 Calibrators, Catalog Number 10491408, IVD --- COMMON/US... Siemens confirmed that the ADVIA¿ Chemistry Systems HbA1c method exhibits a positive bias of up t... Class II Siemens Healthcare Diagnostics
Jul 25, 2013 ADVIA Centaur Folate (100 Tests), SMN 10310308, Cat No. 06367974 In vitro... Calibration failures and/or significant negative shifts in quality control (QC) and patient results. Class II Siemens Healthcare Diagnostics, Inc
Jul 25, 2013 ADVIA Centaur Folate (500 tests), REF SMN 10331250, Cat No. 00203473 In vi... Calibration failures and/or significant negative shifts in quality control (QC) and patient results. Class II Siemens Healthcare Diagnostics, Inc