96 FDA medical device safety recalls.
FDA Medical Device Recall Enforcement Database
Browse 96 FDA medical device recall enforcement actions. Each entry includes the device description, reason for recall, classification (Class I through III based on severity), recalling firm, and distribution details. Use the filters below to search by year, classification, state, or keyword.
Medical Device Recalls
Browse 96 FDA device recalls.
| Date | Product | Reason | Class | Firm |
|---|---|---|---|---|
| Nov 13, 2025 | Alinity ci-series System Control Module (SCM), REF: 03R70-01, 03R70-20 | Abbott Laboratories is recalling their Alinity ci-series System Control Module, a chemistry/immun... | Class II | Abbott Laboratories |
| Sep 18, 2025 | Brand Name: Alinity hq Analyzer Product Name: Alinity hq Analyzer Model/Cat... | Software issue for hq analyzer results in system not visibly applying appropriate flagging to res... | Class II | Abbott Laboratories |
| Mar 18, 2025 | Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number:... | When expired reagents are scanned or manually entered, the system will change the expiration date... | Class II | Abbott Laboratories |
| Mar 18, 2025 | Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number:... | When expired reagents are scanned or manually entered, the system will change the expiration date... | Class II | Abbott Laboratories |
| Mar 18, 2025 | Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number:... | When expired reagents are scanned or manually entered, the system will change the expiration date... | Class II | Abbott Laboratories |
| Mar 18, 2025 | Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number:... | When expired reagents are scanned or manually entered, the system will change the expiration date... | Class II | Abbott Laboratories |
| Mar 18, 2025 | Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number:... | When expired reagents are scanned or manually entered, the system will change the expiration date... | Class II | Abbott Laboratories |
| Jun 14, 2024 | Alinity hq Analyzer, REF: 09P68-01 | 1)If Complete Blood Count with differential and reticulocyte(CBC+Diff+Retic) sample is run, then ... | Class II | Abbott Laboratories |
| Feb 21, 2024 | ARCHITECT STAT Myoglobin Reagent Kit, LN 2K43-25 and 2K43-20. | The reason for the recall is the failure of calibration and quality controls due to a manufacturi... | Class II | Abbott Laboratories |
| Oct 19, 2023 | ARCHITECT c4000 Processing Module REF: 02P24-01 02P24-02 02P24-40 01R24-... | The devices contain a dry natural rubber (latex) subcomponent, which are not labeled as containin... | Class II | Abbott Laboratories |
| Oct 19, 2023 | Abbott CELL-DYN Ruby, Model CD-Ruby, List number 08H67-01, and Abbott CELL-DY... | The devices contain a dry natural rubber (latex) subcomponent, which are not labeled as containin... | Class II | Abbott Laboratories |
| Oct 19, 2023 | Alinity c Processing Module REF 03R67-01 The Alinity c Processing Module... | The devices contain a dry natural rubber (latex) subcomponent, which are not labeled as containin... | Class II | Abbott Laboratories |
| Oct 19, 2023 | Alinity i Processing Module REF 03R65-01 The Alinity i Processing Module... | The devices contain a dry natural rubber (latex) subcomponent, which are not labeled as containin... | Class II | Abbott Laboratories |
| May 30, 2023 | Alinity ci-series System Control Module, REF: 03R70-01 | There are potential performance issues found in the Alinity ci-series System software versions 3.... | Class II | Abbott Laboratories |
| Nov 24, 2022 | Alinity Reaction Vessels, REF: 06P1401, for use with Alinity i Systems; ARCHI... | Reaction vessels may contain a potential contaminant that could affect their optical performance. | Class III | Abbott Laboratories |
| Jul 26, 2022 | ARCHITECT STAT Troponin-I Reagent Kit, REF 2K41-27 | One lot of reagent was manufactured with insufficient quantities of the heterophilic blocking age... | Class II | Abbott Laboratories |
| Jun 9, 2022 | REF: 08P4320, Alinity c Hemoglobin Reagent Kit, IVD, CE, R1 5 x 46.0 mL, R2... | There is a potential for falsely elevated Hemoglobin %A1c and Hemoglobin A1c results to be genera... | Class II | Abbott Laboratories |
| Jun 9, 2022 | REF: 4P52-20 Hemoglobin A1c for use with Architect, IVD, CE, R1 1 x 52 mL, R2... | There is a potential for falsely elevated Hemoglobin %A1c and Hemoglobin A1c results to be genera... | Class II | Abbott Laboratories |
| Jun 9, 2022 | REF: 08P4377, Alinity c Hemoglobin Reagent Kit, IVD, CE, R1 5 x 46.0 mL, R2... | There is a potential for falsely elevated Hemoglobin %A1c and Hemoglobin A1c results to be genera... | Class II | Abbott Laboratories |
| Mar 8, 2022 | ARCHITECT STAT Myoglobin Calibrators, RES 2K43-01 The ARCHITECT STAT Myogl... | This calibrator lot did not meet acceptance criteria during ongoing stability testing. When a cal... | Class II | Abbott Laboratories |
| Feb 3, 2022 | Alinity s System software version2.8.0, REF LN 06P16-01, IVD, OEI: (01)003807... | Software error associated with the immunoassay analyzer wash cycle which is using 1 mL of wash b... | Class II | Abbott Laboratories, Inc |
| Sep 29, 2021 | ARCHITECT c4000 REF 02P24-01/2P24, 1P86, 1R24, and 1R25; ARCHITECT c8000 REF ... | Twelve software-related issues affecting software version 9.41 and earlier were identified. The i... | Class II | Abbott Laboratories |
| Sep 29, 2021 | ARCHITECT i1000SR REF 01L86-01/1L86 and 1L87; ARCHITECT i2000SR REF 03M74-02/... | Twelve software-related issues affecting software version 9.41 and earlier were identified. The i... | Class II | Abbott Laboratories |
| Jul 26, 2021 | Alinity ci-series System Control Module clinical chemistry and immunoassay an... | Due to potential performance issues with software version 3.2.3 and earlier. Performance issues ... | Class II | Abbott Laboratories |
| Jul 12, 2021 | Alinity s System REF 06P16-01. An automated immunoassay analyzer for In Vit... | A design defect (hardware and software) allows liquid waste pressure to build up and potentially ... | Class II | Abbott Laboratories, Inc |
| Jun 10, 2021 | REF 8L92-22, Abbott Activated Alanine aminotransferase, for use with Architec... | There is a potential for >10% under-recovery on Activated Alanine Aminotransferase (A-ALT) sample... | Class II | Abbott Laboratories |
| Jun 10, 2021 | REF 8L92-21, Abbott Activated Alanine Aminotransferase, for use with Architec... | There is a potential for >10% under-recovery on Activated Alanine Aminotransferase (A-ALT) sample... | Class II | Abbott Laboratories |
| Jun 10, 2021 | REF: 08P1820, Abbott Alinity c Activated Alanine Aminotransferase reagent Ki... | There is a potential for 15% under-recovery on Activated Alanine Aminotransferase (A-ALT) samples... | Class II | Abbott Laboratories |
| Jun 10, 2021 | REF 8L92-42, Abbott Activated Alanine Aminotransferase, for use with Architec... | There is a potential for >10% under-recovery on Activated Alanine Aminotransferase (A-ALT) sample... | Class II | Abbott Laboratories |
| May 5, 2021 | v.1.2 Upgrade Kit or v.1.2 upgrade installed on WorkMate Claris System, Model... | Due to a software error, the user may lose functionality of the system or the screen may turn bla... | Class II | Abbott Laboratories Inc. (St Jude Medical) |
| May 5, 2021 | WorkMate Claris v.1.2 Upgrade Kits, Model Numbers H800007 and H800012 | Due to a software error, the user may lose functionality of the system or the screen may turn bla... | Class II | Abbott Laboratories Inc. (St Jude Medical) |
| May 5, 2021 | v.1.2 Upgrade Kit or v.1.2 upgrade installed on WorkMate Claris" System Displ... | Due to a software error, the user may lose functionality of the system or the screen may turn bla... | Class II | Abbott Laboratories Inc. (St Jude Medical) |
| Apr 22, 2021 | NeuroTherm NT2000iX Radiofrequency (RF) Generator, Model Number RFG-NT-2000 | Potential capacitor failure which typically manifests as in inability to provide output on one or... | Class II | Abbott Laboratories Inc. (St Jude Medical) |
| Apr 2, 2020 | Accelerator APS Aliquoter Module, List number: 07L40 - Product Usage: The ACC... | In the case of an AQM clot detection error a portion of the distilled water from the Aliquoter Mo... | Class II | Abbott Laboratories |
| Apr 1, 2020 | ARCHITECT C Peptide Calibrator - Product Usage: used as an aid in the diagnos... | Error in the renewal documentation process for license in Mexico. All in date ARCHITECT C Peptide... | Class III | Abbott Laboratories |
| Apr 1, 2020 | ARCHITECT C Peptide Reagent - Product Usage: used as an aid in the diagnosis ... | Error in the renewal documentation process for license in Mexico. All in date ARCHITECT C Peptide... | Class III | Abbott Laboratories |
| Apr 1, 2020 | ARCHITECT C Peptide Calibrator - Product Usage: used as an aid in the diagnos... | Error in the renewal documentation process for license in Mexico. All in date ARCHITECT C Peptide... | Class III | Abbott Laboratories |
| May 24, 2019 | ARCHITECT BNP Controls Ptoduct Usage: The ARCHITECT BNP Controls are for... | Presence of a time dependent, stability drift in patient and control results returned from ARCHIT... | Class III | Abbott Laboratories |
| May 24, 2019 | ARCHITECT BNP Calibrators Ptoduct Usage: The ARCHITECT BNP Calibrators ar... | Presence of a time dependent, stability drift in patient and control results returned from ARCHIT... | Class III | Abbott Laboratories |
| May 22, 2019 | "HEMOGLOBIN A1c CALIBRATORS *** REF 4P52-01 *** 307261/R03 *** S4P5X0 *** FOR... | Calibrator level 2 bottle may have been manufactured incorrectly for a portion of the lot. The ma... | Class II | Abbott Laboratories |
| May 9, 2019 | Emerald Diluent Tubing, List number 09H50-01, accompanying CELL-DYN Emerald i... | The diluent tubing used on the instrument appears cloudy or to have a powder/film on the inner an... | Class II | Abbott Laboratories |
| May 1, 2019 | ARCHITECT C System Mixer - Product Usage: The ARCHITECT Clinical Chemistry Sy... | Mixer blade may separate from mixer due to the screw and nut failure and detachment. As a result ... | Class II | Abbott Laboratories, Inc |
| Apr 9, 2019 | ARCHITECT Intact PTH Controls - Product Usage: The ARCHITECT Intact PTH Calib... | Devices were delivered without the required dry ice. | Class III | Abbott Laboratories |
| Apr 9, 2019 | ARCHITECT Sirolimus Calibrators - Product Usage: The ARCHITECT Sirolimus Cali... | Devices were delivered without the required dry ice. | Class III | Abbott Laboratories |
| Apr 9, 2019 | ARCHITECT HbA1C Calibrators - Product Usage: The ARCHITECT HbA1c Calibrators ... | Devices were delivered without the required dry ice. | Class III | Abbott Laboratories |
| Apr 9, 2019 | Technopath Multichem S Plus Level 1 - Product Usage: Multichem S Plus (Assaye... | Devices were delivered without the required dry ice. | Class III | Abbott Laboratories |
| Apr 9, 2019 | ARCHITECT STAT High Sensitive Troponin-I Controls - Product Usage: The ARCHIT... | Devices were delivered without the required dry ice. | Class III | Abbott Laboratories |
| Apr 9, 2019 | Technopath Multichem S Plus Level 2 - Product Usage: Multichem S Plus (Assaye... | Devices were delivered without the required dry ice. | Class III | Abbott Laboratories |
| Apr 9, 2019 | ALINITY i Anti-TPO Calibrators - Product Usage: The Alinity i Anti-TPO Calibr... | Devices were delivered without the required dry ice. | Class III | Abbott Laboratories |
| Apr 9, 2019 | ARCHITECT STAT High Sensitive Troponin-I Calibrators - Product Usage: The AR... | Devices were delivered without the required dry ice. | Class III | Abbott Laboratories |