Medical convenience kits Item Number/Description RCGY52D GYN LAPAROSCOPY SPH 0246965 UIGN82...
FDA Device Recall #Z-0179-2026 — Class II — September 4, 2025
Recall Summary
| Recall Number | Z-0179-2026 |
| Classification | Class II — Moderate risk |
| Date Initiated | September 4, 2025 |
| Status | Ongoing |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | American Contract Systems Inc. |
| Location | Grimes, IA |
| Product Type | Devices |
| Quantity | 32,433 total |
Product Description
Medical convenience kits Item Number/Description RCGY52D GYN LAPAROSCOPY SPH 0246965 UIGN82M GYN (PS 907645) UIHY16AE ASC HYSTEROSCOPY PACK (PS 020634)
Reason for Recall
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Distribution Pattern
US Nationwide distribution in the states of SD, IA, MN, WA, IL.
Lot / Code Information
FHGY09S UDI-DI 191072234360 lot 8532011 bag serial number 68874879; LMGY39U UDI-DI 191072209337 Lot 8490211 bag serial numbers 47860009 47861175; OWGY11Q UDI-DI 191072236555 lot 8389711 bag serial number 17622382; RCGY52D UDI-DI 191072232670 lot 8142411 bag serial number 17487909; UIGN82M UDI-DI 191072106674 lot 8084811 bag serial numbers 17391081 17391091 17391111 17391171 lot 8445711 bag serial numbers 17683242 lot 8530511 bag serial numbers 17712404; UIHY16AE UDI-DI 00191072234117 Lot 8171811 N/A * If bag serial number is listed as "n/a", then entire lot is affected
Other Recalls from American Contract Systems Inc.
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-0180-2026 | Class II | Medical convenience kits Item Number/Descri... | Sep 4, 2025 |
| Z-0185-2026 | Class II | Medical convenience kits Item Number/Descri... | Sep 4, 2025 |
| Z-0174-2026 | Class II | Medical convenience kits Item Number/Descri... | Sep 4, 2025 |
| Z-0173-2026 | Class II | Medical convenience kits Item number/ Produ... | Sep 4, 2025 |
| Z-0187-2026 | Class II | Medical convenience kits IHDC30R D AND C PACK | Sep 4, 2025 |
Frequently Asked Questions
Sterility recalls for medical devices vary in severity. If you have already had a procedure using a potentially non-sterile device, contact your healthcare provider immediately — you may need monitoring for signs of infection. Symptoms to watch for include fever, redness or swelling at the surgical site, unusual pain, or discharge. For devices that have not yet been used, they should be quarantined and returned to the manufacturer per the recall notice. Non-sterile implants can cause serious infections; early detection and treatment are critical.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.