Medical convenience kits Item Number/Description HSAN63N ANGIO PACK IHAN02W ANGIOGRAPHY PAC...

FDA Device Recall #Z-0180-2026 — Class II — September 4, 2025

Recall Summary

Recall Number Z-0180-2026
Classification Class II — Moderate risk
Date Initiated September 4, 2025
Status Ongoing
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm American Contract Systems Inc.
Location Grimes, IA
Product Type Devices
Quantity 32,433 total

Product Description

Medical convenience kits Item Number/Description HSAN63N ANGIO PACK IHAN02W ANGIOGRAPHY PACK MHCA98AK CORONARY ANGIO PACK- 205991 RCAN45C ANGIO CUSTOM RCH UDCO10F CVL CORONARY PACK 219224

Reason for Recall

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Distribution Pattern

US Nationwide distribution in the states of SD, IA, MN, WA, IL.

Lot / Code Information

ANCA80AP UDI-DI 191072232168 Lot 8077111 Bag serial number 68119740; HSAN63N UDI-DI 191072232267 Lot 8097211 Bag serial number N/A; IHAN02W UDI-DI 191072201928 Lot 7961811 Bag serial number 68086881 Lot 8037411 Bag serial number N/A; MHCA98AK UDI-DI 191072216014 Lot 8006711 Bag serial number 47415134 Lot 8133011 Bag serial number N/A; RCAN45C UDI-DI 191072229632 Lot 8537111 Bag serial number 68872542; UDCO10F UDI-DI 191072216236 Lot 7987811 Bag serial number N/A * If bag serial number is listed as "N/A", then whole lot is affected

Other Recalls from American Contract Systems Inc.

Recall # Classification Product Date
Z-0185-2026 Class II Medical convenience kits Item Number/Descri... Sep 4, 2025
Z-0174-2026 Class II Medical convenience kits Item Number/Descri... Sep 4, 2025
Z-0173-2026 Class II Medical convenience kits Item number/ Produ... Sep 4, 2025
Z-0187-2026 Class II Medical convenience kits IHDC30R D AND C PACK Sep 4, 2025
Z-0170-2026 Class II Basic Biopsy Tray, Item Number NMBP44L Sep 4, 2025

Frequently Asked Questions

Sterility recalls for medical devices vary in severity. If you have already had a procedure using a potentially non-sterile device, contact your healthcare provider immediately — you may need monitoring for signs of infection. Symptoms to watch for include fever, redness or swelling at the surgical site, unusual pain, or discharge. For devices that have not yet been used, they should be quarantined and returned to the manufacturer per the recall notice. Non-sterile implants can cause serious infections; early detection and treatment are critical.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.