Merlin PCS 3650 programmer Model 3330 software

FDA Device Recall #Z-1495-2022 — Class II — June 17, 2022

Recall Summary

Recall Number Z-1495-2022
Classification Class II — Moderate risk
Date Initiated June 17, 2022
Status Ongoing
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm St. Jude Medical, Cardiac Rhythm Management Division
Location Sylmar, CA
Product Type Devices
Quantity 29,996 programs

Product Description

Merlin PCS 3650 programmer Model 3330 software

Reason for Recall

There is a potential for programmer software (Merlin PCS) and remote monitoring software application (Merlin.net) to display overestimated predicted battery longevity for certain pacemakers.

Distribution Pattern

US: AK AL AR AZ CA CO CT DC DE FL GA HI IA ID IL IN KS KY LA MA MD ME MI MN MO MS MT NC ND NE NH NJ NM NV NY OH OK OR PA PR RI SC SD TN TX UT VA VI VT WA WI WV WY , Guam OUS: ALBANIA ALGERIA ANGOLA Argentina ARMENIA Australia AUSTRIA AZERBAIJAN BAHRAIN Bangladesh BELGIUM BENIN BOLIVIA BOSNIA-HERZ. BRAZIL BULGARIA Burnei Cambodia CAMEROON Canada Chile China Colombia COTE D'IVOIRE CUBA CYPRUS CZECH REPUBLIC DENMARK DOMINICAN REP. ECUADOR EGYPT EL SALVADOR ESTONIA ETHIOPIA FAROE ISLANDS FINLAND FRANCE GEORGIA GERMANY GREECE GREENLAND GUATEMALA Hong Kong HUNGARY ICELAND India Indonesia IRELAND ISRAEL ITALY JORDAN KAZAKHSTAN KOSOVO KUWAIT KYRGYZSTAN LATVIA LEBANON LIBYA LIECHTENSTEIN LITHUANIA LUXEMBOURG MACEDONIA Malaysia MALTA MAURITIUS Mexico MONGOLIA MONTENEGRO MOROCCO Myanmar Nepal NETHERLANDS New Zealand NORWAY OMAN Pakistan PALESTINE PANAMA PARAGUAY PERU Philippines POLAND PORTUGAL QATAR ROMANIA RUSSIAN FED. SAUDI ARABIA SENEGAL Singapore SLOVAKIA SLOVENIA SOUTH AFRICA South Korea SPAIN Srilanka SUDAN SWEDEN SWITZERLAND Taiwan Thailand TRINIDAD TUNISIA TURKEY UKRAINE United Arab Emirates UNITED KINGDOM UZBEKISTAN VENEZUELA Vietnam VIRGIN ISLAND YEMEN ZAMBIA

Lot / Code Information

Software Model 3330 v26.0.1 rev 1 or earlier; UDI/DI 05414734509725

Other Recalls from St. Jude Medical, Cardiac Rhythm Mana...

Recall # Classification Product Date
Z-1496-2022 Class II Merlin 2 PCS MER37000 programmer Model MER3400 ... Jun 17, 2022
Z-1497-2022 Class II Merlin.net model MN5000 Software Jun 17, 2022
Z-0990-2022 Class II Gallant HF Implantable Cardioverter Defibrillat... Mar 10, 2022
Z-0988-2022 Class II Gallant DR Implantable Cardioverter Defibrillat... Mar 10, 2022
Z-0965-2022 Class II Merlin 2 PCS MER3700 programmer Model MER3400 s... Mar 10, 2022

Frequently Asked Questions

A software recall means the device's embedded software or firmware has a defect that could affect its performance or safety. Many software recalls are corrected through firmware updates that can be applied without physically replacing the device. For implantable devices, the update may be delivered wirelessly during a routine clinic visit. For external devices, the manufacturer may provide updated software files or replacement units. Contact your healthcare provider to determine whether your specific device and software version are affected and what action is recommended.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.