Captus 3000 Thyroid Uptake System, Model # 5430-0076 and 5430-0077

FDA Device Recall #Z-0550-2019 — Class II — November 5, 2018

Recall Summary

Recall Number Z-0550-2019
Classification Class II — Moderate risk
Date Initiated November 5, 2018
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Capintec Inc
Location Florham Park, NJ
Product Type Devices
Quantity 385

Product Description

Captus 3000 Thyroid Uptake System, Model # 5430-0076 and 5430-0077

Reason for Recall

There is a potential for the spring arm failure as a result of a broken tension rod (a component inside the arm). A break in the tension rod can casue the collimator to fall downward to its lowest point of travel, approximately 25 inches from the ground.

Distribution Pattern

Worldwide distribution - US nationwide distributed in the states of AR, AZ, CA, CO, CT, DC, FL, IA, IN, LA, MD, MI, MN, NC, NJ, OH, OK, PA, SC, SD, TX, VA, WI, WV, and countries of Argentina, Brazil, Canada, China, Cyprus, Greece, Hong Kong, India, Japan, South Korea, Vietnam.

Lot / Code Information

Serial Numbers 901100 through 901309. Serial Numbers 901310 through 901965 as of 1/15/19.

Other Recalls from Capintec Inc

Recall # Classification Product Date
Z-1564-2020 Class II Captus 4000e Thyroid Uptake System, Model Numbe... Mar 3, 2020
Z-0551-2019 Class II Captus 700 Thyroid Uptake System, Model # 5430-... Nov 5, 2018
Z-2312-2012 Class II Capintec Captus 3000 Thyroid Uptake System ... Jul 24, 2012

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.