VITROS 5600 Integrated System, Version 3.3.1 & below For use in the in vitro quantitative, se...

FDA Device Recall #Z-0650-2019 — Class II — October 19, 2018

Recall Summary

Recall Number Z-0650-2019
Classification Class II — Moderate risk
Date Initiated October 19, 2018
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Ortho-Clinical Diagnostics
Location Rochester, NY
Product Type Devices
Quantity 2413

Product Description

VITROS 5600 Integrated System, Version 3.3.1 & below For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Product Slides, VITROS Chemistry Products MicroTip Reagents and VITROS Immunodiagnostic Products Reagents.

Reason for Recall

Software anomaly may incorrectly extend the on-analyzer stability time for a loaded VITROS Reagent.

Distribution Pattern

US Nationwide Distribution

Lot / Code Information

1) Product Code 6802413 UDI 10758750002740; Serial Numbers J56000024 J56003358 2) Product Code 6802915 (Refurbished) UDI 10758750007110 Serial Numbers: 56000427 56000750 56001012 56001535 56001152 56001641 56000327 56000585 56001437 56000786 56000255 56001235 56000792 56001062 56000308 56001184 56000315 56000458 56001570 56000825 56000661 56000261 56000559 56000818 56000747 56001139 56000317 56000513 56000755 56000883 56001334 56001445 56001141 56000756 56000824 56000718 56001021 56000339 56000741 56000727 56001131 56001558 56001530 56001089 56001264 56000646 56000655 56001027 56001128 56000796 56000890 56000928 56001359 56001851 56000533 56001111 56001689

Other Recalls from Ortho-Clinical Diagnostics

Recall # Classification Product Date
Z-1372-2024 Class II VITROS Performance Verifier I Feb 27, 2024
Z-1373-2024 Class II VITROS Performance Verifier II Feb 27, 2024
Z-2358-2023 Class II VITROS Immunodiagnostic Products Troponin I ES ... Jun 5, 2023
Z-2357-2023 Class II VITROS Immunodiagnostic Products Troponin I ES ... Jun 5, 2023
Z-2129-2023 Class II VITROS Chemistry Products Calibrator Kit 20-in ... May 8, 2023

Frequently Asked Questions

A software recall means the device's embedded software or firmware has a defect that could affect its performance or safety. Many software recalls are corrected through firmware updates that can be applied without physically replacing the device. For implantable devices, the update may be delivered wirelessly during a routine clinic visit. For external devices, the manufacturer may provide updated software files or replacement units. Contact your healthcare provider to determine whether your specific device and software version are affected and what action is recommended.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.