Terumo¿ Advanced Perfusion System 1 (APS1, System 1) 100/120V AC Product Usage: The Terumo¿ ...

FDA Device Recall #Z-1973-2013 — Class II — July 26, 2013

Recall Summary

Recall Number Z-1973-2013
Classification Class II — Moderate risk
Date Initiated July 26, 2013
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Terumo Cardiovascular Systems Corporation
Location Ann Arbor, MI
Product Type Devices
Quantity 1657 total

Product Description

Terumo¿ Advanced Perfusion System 1 (APS1, System 1) 100/120V AC Product Usage: The Terumo¿ Advanced Perfusion System 1 is indicated for use in extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures only when used by a qualified perfusionist who is experienced in the operation of Sarns TM or similar equipment.

Reason for Recall

Terumo Cardiovascular Systems (TCVS) has become aware that an Advanced Perfusion System 1 (APS1) Centrifugal Pump 'Service Pump' message may occur due to user induced motor movement. In this instance, the APS1 will identify the movement (as it is designed) and display a Service Pump message error. The user might interpret this as a pump which requires service rather than a message displayed due to pump movement while not activated.

Distribution Pattern

Worlwide Distribution - US Nationwide: AL,AR,AZ, CA, CO, CT, DC, DE, FL,GA, HI, IA, ID, IL, IN, KY, LA, MA , MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV and the countries of: AUSTRALIA,UNITED ARAB EMIRATES (UAE),Hong Kong, onesia,Singapore,Taiwan,Thailand, Philippines,CHILE, Malaysia,USA,South Korea,BELGIUM,Japan,Mexico,CANADA

Lot / Code Information

Catalog Number or Part Number: 801763 Description: 100/120V AC, Advanced Perfusion System 1

Other Recalls from Terumo Cardiovascular Systems Corpora...

Recall # Classification Product Date
Z-2018-2025 Class II CDI OneView System BPM Probe: Cat. No. CDI75... May 28, 2025
Z-1492-2025 Class II Terumo CDI OneView Monitoring System Hematocrit... Mar 3, 2025
Z-1401-2024 Class II Cardiovascular Procedure Kits that include High... Feb 20, 2024
Z-1346-2023 Class II CAPIOX NX19 Hollow Fiber Oxygenator, Catalog Nu... Mar 1, 2023
Z-1347-2023 Class II CAPIOX FX25 Hollow Fiber Oxygenator, Catalog Nu... Mar 1, 2023

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.