Carl Zeiss Meditec AG VISUREF150 REF 2227-967 - Product Usage: is a diagnostic instrument used to...
FDA Device Recall #Z-0614-2021 — Class II — August 6, 2020
Recall Summary
| Recall Number | Z-0614-2021 |
| Classification | Class II — Moderate risk |
| Date Initiated | August 6, 2020 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Carl Zeiss Meditec AG |
| Location | Jena, N/A |
| Product Type | Devices |
| Quantity | 132 units |
Product Description
Carl Zeiss Meditec AG VISUREF150 REF 2227-967 - Product Usage: is a diagnostic instrument used to determine the objective refractive values and keratometric characteristics of the human eye. The results are used to assist the process of prescribing optical aids such as eyeglasses and contact lenses.
Reason for Recall
Due to the incorrect DC/DC converter being assembled onto the communication board. As a result the effectiveness of the communication board cannot be guaranteed in that the safety function of the board would be insufficient resulting in an increased leakage of current and potentially cause electrical shocks to the user.
Distribution Pattern
Worldwide distribution - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, KS, KY, LA, MD, MI, MN, NC, NE, NJ, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, TX, UT, VA, WA and WY. The countries of Argentina, Austria, Australia, Belgium, Bolivia, Canada, Colombia, Croatia, Cyprus, Denmark, France, Germany, Hong Kong, Hungary, India, Ireland, Italy, Japan, Latvia, Luxembourg, Malaysia, Maldives, Netherlands, Portugal, Singapore, South Africa, South Korea, Spain, Switzerland, Taiwan, Turkey, and United Kingdom.
Lot / Code Information
Model: VISUREF 150 Catalog Number: 000000-2227-967 Serial Numbers: Pending
Other Recalls from Carl Zeiss Meditec AG
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1007-2022 | Class II | ZEISS miLOOP Lens Fragmentation Device REF 3030... | Mar 16, 2022 |
| Z-1133-2021 | Class II | IOLMaster 700 | Dec 7, 2020 |
| Z-0613-2021 | Class II | Carl Zeiss Meditec Ag VISULENS 550 REF 2268-038... | Aug 6, 2020 |
| Z-0108-2019 | Class II | Carl Zeiss Meditec AG, Goeschwitzer Strasse 51-... | Jun 25, 2018 |
| Z-0001-2018 | Class II | Carl Zeiss INTRABEAM PRS 500, product #304534-0... | Aug 30, 2017 |
Frequently Asked Questions
Contact your physician immediately if you use an electrical device or implant affected by this recall. For external devices (monitors, pumps), stop using the device and contact the manufacturer for a replacement. For implanted devices (pacemakers, ICDs), your cardiologist will assess whether you need device replacement surgery — the decision depends on how dependent you are on the device and the actual failure rate observed in the field. Do not attempt to repair or modify a recalled electrical medical device yourself. Monitor for any new or unusual symptoms and report them to your doctor.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.