IOLMaster 700

FDA Recall #Z-1133-2021 — Class II — December 7, 2020

Recall #Z-1133-2021 Date: December 7, 2020 Classification: Class II Status: Terminated

Product Description

IOLMaster 700

Reason for Recall

When using software 1.90.2.09 or 1.90.8.06 and using modality worklist functionality for patient data transfer, the selection of the patient in the patient list may not match the patient information displayed on the right side of the screen

Recalling Firm

Carl Zeiss Meditec AG — Jena, N/A

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

613 devices

Distribution

U.S.: AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, LA, MA, MD, ME, MI, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OK, PA, Puerto Rico, SC, TX, UT, VA, VT, WA, WI, and WV. O.U.S.: Austria, Australia, Canada, Switzerland, Czech Republic, Germany, Denmark, Estonia, Spain, France, United Kingdom, Guatemala, Hong Kong, Indonesia, Ireland, India, Italy, Japan, Kuwait, Malaysia, Netherlands, New Zealand, Philippines, Poland, Portugal, Russia, Sweden, Singapore, Slovakia, Thailand, Taiwan, and South Africa.

Code Information

Model: IOLMaster 700 Catalog Number device: 000000-1932-169

Status

Terminated

Voluntary / Mandated

Voluntary: Firm initiated