da Vinci SP Surgical System, SP1098, 00886874114605 Product Usage: used with an endoscope, surgi...
FDA Device Recall #Z-0927-2020 — Class II — July 26, 2019
Recall Summary
| Recall Number | Z-0927-2020 |
| Classification | Class II — Moderate risk |
| Date Initiated | July 26, 2019 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Intuitive Surgical, Inc. |
| Location | Sunnyvale, CA |
| Product Type | Devices |
| Quantity | 4 Surgical Systems |
Product Description
da Vinci SP Surgical System, SP1098, 00886874114605 Product Usage: used with an endoscope, surgical instruments, and accessories to perform minimally invasive surgical procedures.
Reason for Recall
The firm become aware of a potentially defective capacitors on circuit boards installed on certain da Vinci Xi, X and SP system Vision Side Carts (VSC) and/or Patient Side Carts (PSC). As a result, you may experience either vision loss or non-recoverable errors 307 or 319, preventing further use of the system.
Distribution Pattern
Worldwide distribution - US Nationwide distribution in the states of Alabama, Alaska, Arizona, California, Colorado, Connecticut, Florida, Hawaii, Idaho, Illinois, Iowa, Kansas, Kentucky, Louisiana, Massachusetts, Michigan, Minnesota, Missouri, Montana, New Jersey, New Mexico, New York, North Carolina, Ohio, Oklahoma, Oregon, Pennsylvania, South Carolina, Texas, Virginia, Washington, West Virginia, Wisconsin and countries of Belgium, Brazil, France, Germany, India, Ireland, Italy, Japan, Norway, Panama, Poland, South Korea, Spain, Taiwan, and United Kingdom
Lot / Code Information
System Serial # PSC Serial # VSC Serial # SP0047 Not Affected 675054 SP0049 Not Affected 674261 SP0050 Not Affected 675051 SP0054 Not Affected 682482
Other Recalls from Intuitive Surgical, Inc.
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1355-2026 | Class II | da Vinci 5 Surgeon Console Viewer Display, ASSY... | Dec 30, 2025 |
| Z-1096-2026 | Class II | da Vinci SP, SP Access Port Kit, Large Incisio... | Dec 15, 2025 |
| Z-1134-2026 | Class II | Brand Name: Ion Endoluminal System Product Nam... | Dec 4, 2025 |
| Z-1428-2026 | Class II | Da Vinci 5 ASSY, DV5 CONSOLE, IS5000 Model/... | Nov 17, 2025 |
| Z-1232-2026 | Class II | Brand Name: Da Vinci Product Name: da Vinci X,... | Nov 7, 2025 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.