DJO EMPOWR Tibial Knee Punch Handle Part No. 801-05-040 and 801-05-018 contained in Instrument Tr...
FDA Device Recall #Z-0060-2021 — Class II — August 3, 2020
Recall Summary
| Recall Number | Z-0060-2021 |
| Classification | Class II — Moderate risk |
| Date Initiated | August 3, 2020 |
| Status | Ongoing |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Encore Medical, LP |
| Location | Austin, TX |
| Product Type | Devices |
| Quantity | 190 units |
Product Description
DJO EMPOWR Tibial Knee Punch Handle Part No. 801-05-040 and 801-05-018 contained in Instrument Tray for EMPOWR Porous Knee - Product Usage: The Tibial Punch Handle is used in conjunction with one of three appropriately sized Tibial Punch Guides to broach the tibial canal in preparation for the tibial baseplate keel.
Reason for Recall
There is a potential for the retaining pin of the tibial punch handle to become loose and fall out while impacting. If undetected, there is a potential of leaving debris in the joint space.
Distribution Pattern
U.S. Nationwide distribution including in the states of VA, TN, MN, TX, IL, IN, KY, OH, SC, RI, MD, PA, NY, CA, KS, OK, AR, LA, FL, AL, GA, AZ, NV, WA, MS, ME, NV and Puerto Rico.
Lot / Code Information
Model/Catalog no. 801-05-040; :Lot/serial No.s 281905L08, 281905L15, 283808L01, 305197L02, 318871L10 Model/Catalog no. 801-05-018, Lot/serial No.s: 215488L08, 236760L01, 228231L01, 241946L02, 259710L01, 314969L01
Other Recalls from Encore Medical, LP
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1459-2026 | Class II | Brand Name: EMPOWR 3D KNEE Product Name: EMPOW... | Jan 15, 2026 |
| Z-1458-2026 | Class II | Brand Name: Reverse Shoulder Prosthesis (RSP) ... | Jan 15, 2026 |
| Z-1460-2026 | Class II | Brand Name: Reverse Shoulder Prosthesis (RSP) ... | Jan 15, 2026 |
| Z-1234-2026 | Class II | EMPOWR Acetabular System: Liner, 10 degree Hood... | Jan 5, 2026 |
| Z-1233-2026 | Class II | EMPOWR 3D Knee Tibial Insert 5R 14MM, VE, REF: ... | Jan 5, 2026 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.