Cardiohelp Emergency Drive, Part No. 70104.8002 Product Usage: The Cardiohelp Emergency Drive...

FDA Device Recall #Z-0986-2019 — Class I — February 8, 2019

Recall Summary

Recall Number Z-0986-2019
Classification Class I — Serious risk
Date Initiated February 8, 2019
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Maquet Cardiovascular Us Sales, Llc
Location Wayne, NJ
Product Type Devices
Quantity 10

Product Description

Cardiohelp Emergency Drive, Part No. 70104.8002 Product Usage: The Cardiohelp Emergency Drive is used in emergencies to manually drive the disposable if the Cardiohelp-i fails.

Reason for Recall

Upon servicing the gearwheel in the Emergency Drive the gearwheel may have been re-assembled incorrectly, leading to the inadequate operation of the Emergency Drive.

Distribution Pattern

US Nationwide Distribution in the states of California, Florida, Illinois, Kentucky, Michigan, and New York.

Lot / Code Information

Affected serial numbers: 90420359, 90421204, 90421188, 90421279, 90421228, 90421761, 90420904, 90420231, 90420455, 90421304

Other Recalls from Maquet Cardiovascular Us Sales, Llc

Recall # Classification Product Date
Z-0632-2021 Class II HLS Set Advanced Model Name: BEQ-HLS 5050 USA; ... Nov 20, 2020
Z-0633-2021 Class II HLS Set Advanced, Model: BEQ-HLS 7050 USA; HLS ... Nov 20, 2020
Z-0443-2021 Class II Blood Monitoring Unit (BMU 40), model no. 70104... Sep 30, 2020
Z-0133-2021 Class II Yuno II Mobile Operating Table - Product Usage:... Sep 24, 2020
Z-0132-2021 Class II Yuno Mobile Operating Table - Product Usage: In... Sep 24, 2020

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.