Yuno Mobile Operating Table - Product Usage: Intended for support and position the patient immedi...
FDA Device Recall #Z-0132-2021 — Class II — September 24, 2020
Recall Summary
| Recall Number | Z-0132-2021 |
| Classification | Class II — Moderate risk |
| Date Initiated | September 24, 2020 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Maquet Cardiovascular Us Sales, Llc |
| Location | Wayne, NJ |
| Product Type | Devices |
| Quantity | 62 units US |
Product Description
Yuno Mobile Operating Table - Product Usage: Intended for support and position the patient immediately before, during and after surgical interventions as well as for examination and treatment. Part 143301F0 (US) Part 143301B0 (OUS)
Reason for Recall
Plastic socket of the IEC 60320-1 socket pulled out of the metal cage and electronic parts exposed; where the socket was pulled out and main cable is still connected to the main socket (power source), there is a risk that a user or third person may suffer an electric shock when coming in contact with the exposed electronic components
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of ARGENTINA, AUSTRALIA, BAHRAIN, BOLIVIA, BRAZIL, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, CROATIA, CZECH REPUBLIC, FINLAND, FRANCE GERMANY, HONG KONG, HUNGARY, INDIA, INDONESIA, IRAQ, ISRAEL, ITALY, JAPAN, KENYA, KOREA, KUWAIT, LATVIA, LEBANON, MACAU, MACEDONIA, MALAYSIA, MEXICO, NEW ZEALAND, OMAN, POLAND, QATAR, ROMANIA, RUSSIAN FEDERATION, RWANDA, SAUDI ARABIA, SERBIA, SINGAPORE, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TUNESIA, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM.
Lot / Code Information
Part 143301F0: SN 1 to SN 90 Part 143301B0: SN 1 to SN 267 (OUS)
Other Recalls from Maquet Cardiovascular Us Sales, Llc
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-0633-2021 | Class II | HLS Set Advanced, Model: BEQ-HLS 7050 USA; HLS ... | Nov 20, 2020 |
| Z-0632-2021 | Class II | HLS Set Advanced Model Name: BEQ-HLS 5050 USA; ... | Nov 20, 2020 |
| Z-0443-2021 | Class II | Blood Monitoring Unit (BMU 40), model no. 70104... | Sep 30, 2020 |
| Z-0133-2021 | Class II | Yuno II Mobile Operating Table - Product Usage:... | Sep 24, 2020 |
| Z-2957-2020 | Class III | Maquet/Getinge Rotaflow System Console US (115 ... | Aug 7, 2020 |
Frequently Asked Questions
Contact your physician immediately if you use an electrical device or implant affected by this recall. For external devices (monitors, pumps), stop using the device and contact the manufacturer for a replacement. For implanted devices (pacemakers, ICDs), your cardiologist will assess whether you need device replacement surgery — the decision depends on how dependent you are on the device and the actual failure rate observed in the field. Do not attempt to repair or modify a recalled electrical medical device yourself. Monitor for any new or unusual symptoms and report them to your doctor.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.