Oxygen Masks, Elongated (under chin), Adult Med Conc., No Tubing, Product Code 104-E;
FDA Device Recall #Z-2228-2017 — Class II — May 10, 2017
Recall Summary
| Recall Number | Z-2228-2017 |
| Classification | Class II — Moderate risk |
| Date Initiated | May 10, 2017 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | ConvaTec, Inc |
| Location | Greensboro, NC |
| Product Type | Devices |
| Quantity | 816,709 |
Product Description
Oxygen Masks, Elongated (under chin), Adult Med Conc., No Tubing, Product Code 104-E;
Reason for Recall
Reports of tubing detaching from oxygen masks either prior to use (in packaging) or during use were made by different end users and for different batches. The failures were caused by an improper solvent bond between the oxygen tubing and the oxygen mask connector. While the defect affects less than 2% of recalled product, there is a potential for harm when oxygen supply to the user is disrupted.
Distribution Pattern
CA, CT, FL, GA, IL, MO, NJ, NY, NC, OH, PA, RI, VA , Puerto Rico, Australia, Belgium, Canada, Columbia, Czech, Denmark, Finland, France, Germany, Greece, Israel, Italy, Japan, Jordan, Malaysia, Malta, Netherlands, Norway, Rwanda, Saudi Arabia, Sweden, Switzerland, United Kingdom, and Vietnam
Lot / Code Information
Lot Number 102153, 102339, 102941, 103176, 103453, 103718, 103919, 104471, 105034, 105282, 105427, 105567, 105996, 106960, 108658, 112024, 112530
Other Recalls from ConvaTec, Inc
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1584-2026 | Class II | Convatec, EsteemBody Drainable Pouch REF:42365... | Feb 16, 2026 |
| Z-1583-2026 | Class II | Convatec, EsteemBody Drainable Pouch REF:42365... | Feb 16, 2026 |
| Z-1582-2026 | Class II | Convatec, EsteemBody Drainable Pouch REF:42364... | Feb 16, 2026 |
| Z-0160-2026 | Class II | DuoDERM" Extra Thin dressings are highly flexib... | Aug 29, 2025 |
| Z-0903-2025 | Class II | Esteem Synergy Stomahesive Skin Barrier, part o... | Nov 29, 2024 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.