Esteem Synergy Stomahesive Skin Barrier, part of Esteem synergy, which is a two-piece ostomy syst...
FDA Device Recall #Z-0903-2025 — Class II — November 29, 2024
Recall Summary
| Recall Number | Z-0903-2025 |
| Classification | Class II — Moderate risk |
| Date Initiated | November 29, 2024 |
| Status | Ongoing |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | ConvaTec, Inc |
| Location | Greensboro, NC |
| Product Type | Devices |
| Quantity | 240 |
Product Description
Esteem Synergy Stomahesive Skin Barrier, part of Esteem synergy, which is a two-piece ostomy system consisting of an adhesive skin barrier and an adhesive pouch: ESTEEM SYN FLX WHT16/35MM PRECUT(1X10), REF: 405473; ESTEEM SYN FLX WHT35/35MM PRECUT(1X10), REF: 405479
Reason for Recall
Ostomy skin barrier package contains a pre-cut wafer of 28mm instead of 16mm and/or 35mm pre-cut wafers, as stated package labeling. If pre-cut wafers are too big the device cannot be used and if the pre-cut wafer is too small and the device user tries to apply, this could cause abrasion, stomal cut, and bleeding
Distribution Pattern
US distribution to states of: OH, NY, IL
Lot / Code Information
REF/UDI-DI/Lot: 405473/00768455185399/3J03813, 405479/00768455185450/3J03814
Other Recalls from ConvaTec, Inc
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1584-2026 | Class II | Convatec, EsteemBody Drainable Pouch REF:42365... | Feb 16, 2026 |
| Z-1583-2026 | Class II | Convatec, EsteemBody Drainable Pouch REF:42365... | Feb 16, 2026 |
| Z-1582-2026 | Class II | Convatec, EsteemBody Drainable Pouch REF:42364... | Feb 16, 2026 |
| Z-0160-2026 | Class II | DuoDERM" Extra Thin dressings are highly flexib... | Aug 29, 2025 |
| Z-1495-2024 | Class II | ConvaTec CarboFlex, 15cm x 20cm, 6 in. x 8 in.,... | Feb 27, 2024 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.