TSX-101A Aquilion 32/64 ONE Computed Tomography X-Ray System; TXS-101/R Aquilion RXL Computed Tom...
FDA Device Recall #Z-1113-2014 — Class II — October 1, 2013
Recall Summary
| Recall Number | Z-1113-2014 |
| Classification | Class II — Moderate risk |
| Date Initiated | October 1, 2013 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Toshiba American Medical Systems Inc |
| Location | Tustin, CA |
| Product Type | Devices |
| Quantity | 338 units |
Product Description
TSX-101A Aquilion 32/64 ONE Computed Tomography X-Ray System; TXS-101/R Aquilion RXL Computed Tomography X-Ray System; TSX-301A Aquilion One Computed Tomography X-Ray System; TSX-301B Aquilion Premium Computed Tomography X-Ray System; TSX-302A Aquilion Prime Computed Tomography X-Ray System. This device is indicated to acquire and display cross sectional volumes of the whole body, to include the head. The Aquilion RXL has the capability to provide volume sets. These volume sets can be used to perform specialized studies, using indicated software/hardware, by a trained and qualified physician.
Reason for Recall
Toshiba intitiated this recall because their investigation revealed that due to a problem of the X-ray output control software, the X-ray output conditions may not be set correctly. As a result, X-ray exposure under the scan conditions used for contrast study may not start.
Distribution Pattern
Nationwide Distribution.
Lot / Code Information
SERIAL NUMBER JDA1062176, JDA1062177, JDA1072183, JDA1072184, JDA1072186, JDA1072187, JDA1082188, JDA1082189, JDA1082190, JDA10X2191, JDA10X2192, JDA10X2193, JDA10X2194, JDA10X2195, JDA10X2196, JDA10X2197, JDA10X2198, JDA10Y2199, JDA10Y2200, JDA10Y2201, JDA10Y2202, JDA10Y2203, JDA10Y2204, JDA10Y2205, JDA10Z2206, JDA10Z2207, JDA10Z2209, JDA10Z2210, JDA10Z2211, JDA10Z2212, JDA10Z2213, JDA10Z2214, JDA10Z2215, JDA1112216, JDA1112217, JDA1112218, JDA1112219, JDA1112221, JDA1112222, JDA1112224, JDA1112225, JDA1112226, JDA1112227, JDA1122231, JDA1122234, JDA1142236, JDA1162237, JDA1162238, JDA1162239, JDA1182240, JDA1182241, JDA1182242, JDA1182243, JDA1182244, JDA11X2245, JDA11X2246, JDA11X2247, JDA11Z2249, JDA11Z2250, JDA11Z2252, JDA11Z2253, JDA11Z2254, JDA11Z2255, JDA1222257, JDA1222258, JDA1222259, JDA1222260, JDA1222261, JDA1242264, JDA1252265, JDA1252266, JDA1252267, JDA12Z2271, JDA1312272, JGA1062178, JGA1062181, JGA1062182, JGA1112220, JGA1112228, JGA11Z2256, JGA1222262, JGA12X2269, JLA1062179, JLA1062180, JLA1072185, JLA10Z2208, JLA1122232, JLA1122233, JLA1132235, JLA11Y2248, JLA11Z2251, JLA1232263, JLA1262268, JLA12Y2270, KDA1062080, KDA1072081, KDA1072082, KDA1072083, KDA1072085, KDA1072086, KDA1072088, KDA1082090, KDA1082091, KDA1082092, KDA1092093, KDA10X2094, KDA10X2095, KDA10X2096, KDA10X2097, KDA10X2098, KDA10X2099, KDA10Y2100, KDA10Y2101, KDA10Y2102, KDA10Z2104, KDA1112107, KDA1122110, KDA1122111, KDA1132113, KDA1142114, KDA1152115, KDA1152116, KDA1152117, KDA1152118, KDA1162119, KDA1162120, KDA1162121, KDA1162122, KDA1162123, KDA1162124, KDA1172125, KDA1182126, KDA1182127, KDA1182128, KDA1182129, KDA1192130, KDA1192131, KDA11X2132, KDA11X2133, KDA11Y2134, KDA11Z2136, KDA11Z2137, KDA11Z2138, KDA11Z2139, KDA11Z2140, KDA1222141, KDA1222142, KDA1222143, KDA1232144, KDA1242145, KDA1242146, KDA1252147, KDA1272148, KDA1282149, KDA1292151, KDA12X2152, KDA12Y2153, KDA12Z2154, KDA1322155, KGA1082089, KGA10Z2103, KGA10Z2105, KGA1112106, KGA1112108, KGA1122109, KGA1122112, KGA11Z2135, KLA1072084, KLA1072087, KLA1292150, RDA1292019, RDA12X2023, RDA12X2024, RDB12Y2029, RDB12Z2032, RDB1312037, RDB1312038, RGA12X2025, RGB12Y2028, RGB1312034, 2DA0792002, 2DA0792004, 2DA07X2005, 2DA07X2008, 2DA07Y2010, 2DA0842023, 2DA0842026, 2DA0862034, 2DA0862036, 2DA0872038, 2DA0882048, 2DA0882049, 2DA0882051, 2DA0882052, 2DA0882053, 2DA08X2061, 2DA08X2067, 2DA08X2070, 2DA08X2071, 2DA08X2072, 2DA08Y2077, 2DA08Y2080, 2DA08Z2092, 2DA0952115, 2DA0962116, 2DA0972123, 2DA0982131, 2DA0982133, 2DA09X2139, 2DA09X2143, 2DA09Y2148, 2DA09Y2151, 2DA09Z2158, 2DA09Z2161, 2DA09Z2162, 2DA1012165, 2DA1012167, 2DA1012168, 2DA1012169, 2DA1012170, 2DA1012174, 2DA1062208, 2DA1072212, 2DB1082219, 2DB1082220, 2DB1082222, 2DB10X2241, 2DB10Y2253, 2DB10Y2257, 2DB10Z2265, 2DB1112275, 2DB1112283, 2DB1122284, 2DB1122291, 2DB1122294, 2DC1142308, 2DC1142311, 2DC1152316, 2DC1162324, 2DC1172328, 2DC1172332, 2DC1172333, 2DC11X2352, 2DC11Y2365, 2DC11Y2367, 2DC11Z2372, 2DC11Z2374, 2DC1212377, 2DC1212383, 2DC1232392, 2DC1232393, 2DC1232396, 2DC1232397, 2DC1232398, 2DC1242407, 2DC1242409, 2DC1252410, 2DC1262412, 2DC1272417, 2DC1272421, 2DC1282423, 2DC12X2427, 2DC12Y2430, 2DC12Y2431, 2DC12Z2433, 2DC1312435, 2DC1312436, 2DC1322437, 1DA0962006, 1DA0972007, 1DA0972008, 1DA0992011, 1DA09Y2014, 1DA1062015, 1DB1072016, 1DB1082017, 1DB1082018, 1DB1082019, 1DB10Y2020, 1DB10Y2021, 1DB10Y2022, 1DB1112025, 1DB1122027, 1DC1142030, 1DC1152031, 1DC1162032, 1DC1162033, 1DC1172035, 1DC11Y2037, 1DC11Y2038, 1DC11Z2040, 1DC1222042, 1DC1232043, 1DC1232044, 1DC1252045, 1DC1252046, 1DC1262048, 1DC1272050, 1DC1282051, 1DC1292052, 1DC12X2053, 1DC12Y2054, 1DC12Y2055, 1DC12Z2056, 1DC1312057, 1DC1312058, 1DC1342059, 1GC1262047, 1GC1272049, 1DA1222096, 1DA1222097, 1DA1232099, 1DA1232100, 1DA1232103, 1DA1252124, 1DA1252131, 2DA1252001, 2DA1262003, 2DA1272004, 2DA1272005, 2DA1272006, 2DA1282007, 2DA1292008, 2DA12X2009, 2DA12X2010, 2DA12X2012, 2DA12X2013, 2DA12Y2014, 2DA12Y2015, 2DA12Y2016, 2DA12Z2017, 2DA12Z2018, 2DA1312019, 2DA1312020, 2DA1312021, 2DA1312022, 2GA1262002, 2GA12X2011, 2DA1072211, 2DA08X2063, 2DC11Y2368.
Other Recalls from Toshiba American Medical Systems Inc
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-2315-2018 | Class II | Diagnostic Ultrasound System, Aplio I-series. ... | Jan 4, 2018 |
| Z-0843-2018 | Class II | INFX-8000C Fluoroscopic X-Ray Systems | Dec 29, 2017 |
| Z-0842-2018 | Class II | INFX-8000F Fluoroscopic X-Ray Systems | Dec 29, 2017 |
| Z-0845-2018 | Class II | INFX-8000H Fluoroscopic X-Ray Systems | Dec 29, 2017 |
| Z-0844-2018 | Class II | INFX-8000V Fluoroscopic X-Ray Systems; INFINIX ... | Dec 29, 2017 |
Frequently Asked Questions
A software recall means the device's embedded software or firmware has a defect that could affect its performance or safety. Many software recalls are corrected through firmware updates that can be applied without physically replacing the device. For implantable devices, the update may be delivered wirelessly during a routine clinic visit. For external devices, the manufacturer may provide updated software files or replacement units. Contact your healthcare provider to determine whether your specific device and software version are affected and what action is recommended.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.