Terumo Advanced Perfusion System 1 Ultrasonic Air Sensor 3/8x3/32, Catalog Number 5773 - Product ...

FDA Device Recall #Z-2959-2020 — Class II — August 25, 2020

Recall Summary

Recall Number Z-2959-2020
Classification Class II — Moderate risk
Date Initiated August 25, 2020
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Terumo Cardiovascular Systems Corporation
Location Ann Arbor, MI
Product Type Devices
Quantity 100

Product Description

Terumo Advanced Perfusion System 1 Ultrasonic Air Sensor 3/8x3/32, Catalog Number 5773 - Product Usage: intended for use with Terumo and Sarns brand air detection and heart-lung systems to detect gross air bubbles in the line during extracorporeal procedures.

Reason for Recall

Internal testing at the supplier, revealed that a crimped wire was placed into an air bubble detector (ABD) sensor connector backwards, preventing the crimped wire from locking into the mating wire connector. This intermittent connection may result in false air bubble alarms.

Distribution Pattern

Worldwide distribution - US Nationwide distribution.

Lot / Code Information

UDI: 00886799000038; Serial Numbers: 20288 20289 20290 20291 20292 20293 20294 20295 20296 20299 20300 20301 20302 20306 20307 20308 20309 20310 20312 20317 20318 20320 20321 20322 20323 20324 20326 20328 20329 20331 20332 20334 20338 20339 20341 20342 20343 20344 20345 20346 20347 20348 20349 20350 20351 20352 20358 20365 20368 20373 20374 20375 20381 20382 20383 20384 20385 20387 20388 20392 20393 20395 20396 20397 20398 20399 20401 20402 20403 20404 20405 20408 20410 20411 20412 20555 20556 20557 20558 20559 20560 20561 20562 20563 20564 20565 20566 20567 20568 20569 20570 20571 20572 20573 20574 20575 20576 20577 20578 20579

Other Recalls from Terumo Cardiovascular Systems Corpora...

Recall # Classification Product Date
Z-2018-2025 Class II CDI OneView System BPM Probe: Cat. No. CDI75... May 28, 2025
Z-1492-2025 Class II Terumo CDI OneView Monitoring System Hematocrit... Mar 3, 2025
Z-1401-2024 Class II Cardiovascular Procedure Kits that include High... Feb 20, 2024
Z-1346-2023 Class II CAPIOX NX19 Hollow Fiber Oxygenator, Catalog Nu... Mar 1, 2023
Z-1347-2023 Class II CAPIOX FX25 Hollow Fiber Oxygenator, Catalog Nu... Mar 1, 2023

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.