Boston Scientific, EMBLEM S-ICD Programmer Model 3200 Programmer.
FDA Device Recall #Z-1178-2017 — Class II — January 12, 2017
Recall Summary
| Recall Number | Z-1178-2017 |
| Classification | Class II — Moderate risk |
| Date Initiated | January 12, 2017 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Boston Scientific Corporation |
| Location | Saint Paul, MN |
| Product Type | Devices |
| Quantity | 4,500 |
Product Description
Boston Scientific, EMBLEM S-ICD Programmer Model 3200 Programmer.
Reason for Recall
There is a potential for radio frequency (RF) interference to alter wireless communication from a programmer, which in rare instances may cause an S-ICD to perform an unintended command. This behavior can only occur during an active, in-clinic interrogation/programming session with the Model 3200 S-ICD programmer. There is no risk of this behavior occurring when the LATITUDE Patient Management System communicates with an S-ICD in an ambulatory setting.
Distribution Pattern
Worldwide Distribution - USA ( nationwide) including PR and to the states of : AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY., and to the countries of :Canada, Austria, Belgium, Czech Republic, Denmark, Finland, France, French Polynesia, Guadeloupe, Monaco, New Caledonia, Reunion, Germany, Greece, Hungary, Ireland, Israel, Italy, Latvia, Lithuania, Luxembourg, Netherlands, Norway, Poland, Portugal, Slovakia, Spain, Sweden, Switzerland, United Kingdom, Australia, China, Egypt, Hong Kong, India, Iran, Japan, Jordan, Kuwait, Lebanon, Malaysia, New Zealand, Qatar, Saudi Arabia, Singapore, South Africa, Thailand, Tunisia, Turkey, United Arab Emirates, Argentina, Bermuda and Brazil.
Lot / Code Information
All serial numbers.
Other Recalls from Boston Scientific Corporation
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1551-2026 | Class II | EndoVive Safety PEG Kit; Outer Box Material Nu... | Feb 10, 2026 |
| Z-1160-2026 | Class I | AXIOS Stent and Electrocautery-Enhanced Deliver... | Dec 19, 2025 |
| Z-1162-2026 | Class I | *HOT AXIOS Stent and Electrocautery-Enhanced De... | Dec 19, 2025 |
| Z-1159-2026 | Class I | AXIOS Stent and Electrocautery-Enhanced Deliver... | Dec 19, 2025 |
| Z-1163-2026 | Class I | *HOT AXIOS Stent and Electrocautery-Enhanced De... | Dec 19, 2025 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.