Situate Laparotomy Sponge, RF and X-ray Detectable, REF: L1818-04P01C-1
FDA Device Recall #Z-1330-2024 — Class II — February 21, 2024
Recall Summary
| Recall Number | Z-1330-2024 |
| Classification | Class II — Moderate risk |
| Date Initiated | February 21, 2024 |
| Status | Ongoing |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Covidien |
| Location | Lafayette, CO |
| Product Type | Devices |
| Quantity | 755,800 |
Product Description
Situate Laparotomy Sponge, RF and X-ray Detectable, REF: L1818-04P01C-1
Reason for Recall
Packs including 5 laparotomy sponges may contain incorrect product with a blue loop, X-ray detectable feature, and ethylene oxide treatment; however, this incorrect product wasn't steam pre-treated prior to sterilization, and has no RF tag, which if non-sterile, may lead to infection, tissue trauma, and sepsis, and if in the scannable area, would lead a "CLEAR" display.
Distribution Pattern
US Nationwide distribution including in the states of AK, AL, AZ, CA, CO, FL, GA, IA, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, PA, RI, TN, TX, UT, VA, WA, WI, WV.
Lot / Code Information
UDI-DI: 20884521700953, 10884521700956. Lots: 230504KF, 220903KF, 221102KF
Other Recalls from Covidien
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-2151-2025 | Class II | Covidien Nellcor Bedside SpO2 Patient Monitorin... | Jun 23, 2025 |
| Z-1915-2025 | Class I | Newport HT70 Ventilator, REF: HT70M-JP-NA, HT70... | May 21, 2025 |
| Z-1390-2025 | Class I | Shiley Adult Flexible Tracheostomy Tube with Ta... | Feb 26, 2025 |
| Z-2717-2024 | Class II | Covideien McGRATH MAC Video Laryngoscope, REF: ... | Jul 9, 2024 |
| Z-2636-2024 | Class I | McGrath Mac Video Laryngoscope, REF: 300-000-00... | Jul 9, 2024 |
Frequently Asked Questions
Sterility recalls for medical devices vary in severity. If you have already had a procedure using a potentially non-sterile device, contact your healthcare provider immediately — you may need monitoring for signs of infection. Symptoms to watch for include fever, redness or swelling at the surgical site, unusual pain, or discharge. For devices that have not yet been used, they should be quarantined and returned to the manufacturer per the recall notice. Non-sterile implants can cause serious infections; early detection and treatment are critical.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.