Covidien Nellcor Bedside SpO2 Patient Monitoring System: REF: PM100N, 10005941, DLPM100N, DL10005...

FDA Device Recall #Z-2151-2025 — Class II — June 23, 2025

Recall Summary

Recall Number Z-2151-2025
Classification Class II — Moderate risk
Date Initiated June 23, 2025
Status Ongoing
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Covidien
Location Boulder, CO
Product Type Devices
Quantity 264,790

Product Description

Covidien Nellcor Bedside SpO2 Patient Monitoring System: REF: PM100N, 10005941, DLPM100N, DL10005941, DSPM100N, 10005941-SG, DS10005941, 10005941J, 10005941JP, 10005941JPN

Reason for Recall

Bedside SpO2 Patient Monitoring System alarms not heard/recognized, resulting in: treatment delay, lack of low oxygen saturation response, respiratory failure, arrhythmia. Highlighting operator's manual: 1)For patient Safety don't silence/decrease alarm volume, 2)Sensor off alarm (no perfusion detected) is low priority alarm, 3)Set monitors to Homecare Mode outside of hospital/professional setting

Distribution Pattern

Worldwide - US Nationwide distribution including in the states of MS, TX, FL, SC, VA, AZ, PA, MN, OH, CA, MI, CO, KS, IN, IL, TN, MO, IA, OK, WI, AK, NY, MD, AL, WV, GA, NC, AR, WA, SD, KY, DE, NJ, UT, NV, MA, MT, GU, NE, LA, HI, OR, RI, NH, ND, NM, ME, CT, DC, WY and the countries of Algeria, Andorra, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belgium, Bermuda, Bhutan, Bolivia, Bosnia And Herzegovina, Botswana, Brazil, Brunei Darussalam, Bulgaria, Cambodia, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, French Guiana, Germany, Ghana, Greece, Guadeloupe, Guatemala, Hong Kong, Hungary, India, Indonesia, Iran, Islamic Republic Of, Ireland, Isle of Man, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Maldives, Malta, Mauritius, Mexico, Montenegro, Morocco, Myanmar, Namibia, Netherlands, New Zealand, North Macedonia, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Vietnam, Zimbabwe.

Lot / Code Information

REF/UI-DI/Bundle Code/IFU Part Number: PM100N/10884521196728/NELL-HOMECARE, BNPULSOKSYMETR, PM100N, PM100NKIT, PM100NKITCHDE, PM100NKITCHFR, PM100NKITDA, PM100NKITDE-2, PM100NKITES, PM100NKITFI, PM100NKITFR, PM100NKIT-FR, PM100NKITIT, PM100NKIT-NL, PM100NKITNO, PM100NKITPT, PM100NKITSV, PM100NKITUK, PM100N-10, PM100N-2XMAXN, PM100N-2XMAXN-CC, PM100N-3DYS-CC, PM100N-DYS, PM100N-HC, PM100N-MAXN, PM100N-MAXN-CC, PM100N-OXIAN-CC/PT00156589 - AR - Arabic, PT00156609 - AR - Arabic, PT00156588 - BG - Bulgarian, PT00156610 - BG - Bulgarian, PT00156586 - CS - Czech, PT00156612 - CS - Czech, PT00156585 - DA - Danish, PT00156613 - DA - Danish, PT00156578 - DE - German, PT00156625 - DE - German, PT00156577 - EL - Greek, PT00156626 - EL - Greek, PT00156509 - EN - English, PT00156328 - EN - English, PT00156563 - ES - Spanish, PT00156639 - ES - Spanish, PT00156583 - ET - Estonian, PT00156615 - ET - Estonian, PT00156582 - FI - Finnish, PT00156616 - FI - Finnish, PT00156581 - FR - French, PT00156617 - FR - French, PT00156587 - HR - Croatian , PT00156611 - HR - Croatian , PT00156576 - HU - Hungarian, PT00156627 - HU - Hungarian, PT00156575 - IT - Italian, PT00156628 - IT - Italian, PT00156580 - KA - Georgian, PT00156618 - KA - Georgian, PT00156591 - KO - Korean, PT00156642 - KO - Korean, PT00156573 - LT - Lithuanian, PT00156630 - LT - Lithuanian, PT00156574 - LV - Latvian, PT00156629 - LV - Latvian, PT00156584 - NL - Dutch, PT00156614 - NL - Dutch, PT00156572 - NO - Norwegian, PT00156631 - NO - Norwegian, PT00156571 - PL - Polish, PT00156632 - PL - Polish, PT00156570 - PT-BR - Brazilian Portuguese, PT00156633 - PT-BR - Brazilian Portuguese, PT00156569 - RO - Romanian, PT00156634 - RO - Romanian, PT00156567 - RU - Russian, PT00156635 - RU - Russian, PT00156565 - SK - Slovak, PT00156637 - SK - Slovak, PT00156564 - SL - Slovenian, PT00156638 - SL - Slovenian, PT00156566 - SR - Serbian, PT00156636 - SR - Serbian, PT00156562 - SV Swedish, PT00156640 -SV - Swedish, PT00156561 - TR - Turkish, PT00156641 - TR - Turkish, PT00156590 - ZH - SimplifIed Chinese, PT00156643 - ZH - SimplifIed Chinese; 10005941/10884521163454/PM100N-10, PM100N-2XMAXN, PM100N-2XMAXN-CC, PM100N-3DYS-CC, PM100N-DYS, PM100N-HC, PM100N-MAXN, PM100N-MAXN-CC, PM100N-OXIAN-CC, BEDSIDESP2KIT2, NELLSPO2, PM100N-HC, NELL-HOMECARE, ROEBEDSIDEKIT, BEDSIDESP2PT2, BEDSIDESP2NO2, BEDSIDESP2PL2, BEDSIDESP2NL2, BEDSIDESP2KITNM, BEDSIDESP2K2UK, BEDSIDESP2K2IT, BEDSIDESP2K2FR, BEDSIDESP2K2-2, BEDSIDEKITISR/PT00156645 - AR - Arabic, PT00156647 - BG - Bulgarian, PT00156649 - CS - Czech, PT00156650 - DA - Danish, PT00156656 - DE - German, PT00156657 - EL - Greek, PT00156324 - EN - English, PT00156673 - ES - Spanish, PT00156652 - ET - Estonian, PT00156653 - FI - Finnish, PT00156654 - FR - French, PT00156648 - HR - Croatian, PT00156658 - HU - Hungarian, PT00156659 - IT - Italian, PT00156655 - KA - Georgian, PT00156677 - KO - Korean, PT00156661 - LT - Lithuanian, PT00156660 - LV - Latvian, PT00156651 - NL - Dutch, PT00156662 - NO - Norwegian, PT00156664 - PL - Polish, PT00156666 PT-BR - Brazilian Portuguese, PT00156667 - RO - Romanian, PT00156668 - RU - Russian, PT00156671 - SK - Slovak, PT00156672 - SL - Slovenian, PT00156669 - SR - Serbian, PT00156674 - SV - Swedish, PT00156675 - TR - Turkish, PT00156676 - ZH - Simplified Chinese, NELL-SPO2-AD3+ - Canadian, NELL-SPO2-AD3 - Canadian; 10005941/A8845211634501; DLPM100N/10884521527607; DL10005941/10884521173293; DSPM100N/DSPM100N; 10005941-SG/10884521171534/BNINHLLCODE, BNINNEL100A, BNINNEL100AS, BNINNEL100AVJ, BNINNEL100AYS, BNINNELDYS, BNHKLIBRANEOSET, BNHKLIBRASET; DS10005941/DS10005941; 10005941J/10005941JP; 10005941JP/884521512019; 10005941JPN/884521188689. All serial numbers

Other Recalls from Covidien

Recall # Classification Product Date
Z-1915-2025 Class I Newport HT70 Ventilator, REF: HT70M-JP-NA, HT70... May 21, 2025
Z-1390-2025 Class I Shiley Adult Flexible Tracheostomy Tube with Ta... Feb 26, 2025
Z-2717-2024 Class II Covideien McGRATH MAC Video Laryngoscope, REF: ... Jul 9, 2024
Z-2636-2024 Class I McGrath Mac Video Laryngoscope, REF: 300-000-00... Jul 9, 2024
Z-2637-2024 Class I McGRATH" MAC EMS Video Laryngoscope, REF: 300-2... Jul 9, 2024

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.