uDR 596i Digital Medical X-Ray Imaging System - Product Usage: intended to use by a qualified/ t...

FDA Device Recall #Z-2521-2020 — Class II — February 27, 2020

Recall Summary

Recall Number Z-2521-2020
Classification Class II — Moderate risk
Date Initiated February 27, 2020
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Shanghai United Imaging Healthcare Co., Ltd.
Location Shanghai
Product Type Devices
Quantity 6 units

Product Description

uDR 596i Digital Medical X-Ray Imaging System - Product Usage: intended to use by a qualified/ trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, limbs and trunk.

Reason for Recall

Two issues were identified with the stationary x-ray system. 1. Some position illustrations in the patient exam interface show the incorrect anatomical position for the standard of care. When the image is acquired with the wrong reference position, clinical diagnostic requirements cannot be met.T2. Where image stitching is used, the left and right tag information may be lost. When these problems are encountered, an additional exposure is needed and the patient will receive more radiation.

Distribution Pattern

US Nationwide distribution including in the state of Texas.

Lot / Code Information

Model Number uDR 596i with software versions R002.19.0-1.8402-690106 and R002.19.0.8648.690112. Serial numbers: 270012, 270013, 270014, 270015, 270016 and 270017. UDI numbers: (01)06971576833023(240)88000093(21)270012; (01)06971576833023(240)88000093(21)270013; (01)06971576833023(240)88000093(21)270014; (01)06971576833023(240)88000093(21)270015; XXX; and XXX.

Other Recalls from Shanghai United Imaging Healthcare Co...

Recall # Classification Product Date
Z-1353-2023 Class II Shanghai United Imaging Healthcare Co., Ltd. Po... Mar 1, 2023
Z-0087-2023 Class II Positron Emission Tomography and Computed Tomog... May 30, 2022
Z-2594-2021 Class II uEXPLORER PET/CT System - Product Usage: intend... Aug 2, 2021
Z-2593-2021 Class II uMI 550 System - Product Usage: intended to be ... Aug 2, 2021
Z-1160-2021 Class II Digitial Medica X-ray Imagining System, Model: ... Jan 25, 2021

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.