Heater Cooler Unit Usage: The Heater-Cooler Unit (HCU 30) supplies temperature-controlled wat...

FDA Device Recall #Z-0348-2017 — Class II — September 27, 2016

Recall Summary

Recall Number Z-0348-2017
Classification Class II — Moderate risk
Date Initiated September 27, 2016
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Maquet Cardiovascular Us Sales, Llc
Location Wayne, NJ
Product Type Devices
Quantity 305 units

Product Description

Heater Cooler Unit Usage: The Heater-Cooler Unit (HCU 30) supplies temperature-controlled water for cardioplegia heat exchangers and for blood heat exchangers in extra corporeal circulation and for blankets with which patients can be warmed or cooled. The HCU 30 is intended to circulate water through heat exchange circuits to warm or cool a patient during short duration cardiopulmonary bypass procedures lasting 6 hours or less.

Reason for Recall

There is the potential for contaminated water to enter parts of the Heater-Cooler Unit circuits. There is a potential risk of bacterial contamination including Mycobacteria species.

Distribution Pattern

Worldwide Distribution - US Nationwide

Lot / Code Information

HCU 30 Base Unit 100-120 V, Model Number 704629 I 70103.4371 HCU 30 Base Unit Canadian/US, Model Number 705501 I 70103.4653 HCU 30 Base Unit 200-240 V, Model Number 939001 I 70102.8718 All serial numbers are affected.

Other Recalls from Maquet Cardiovascular Us Sales, Llc

Recall # Classification Product Date
Z-0632-2021 Class II HLS Set Advanced Model Name: BEQ-HLS 5050 USA; ... Nov 20, 2020
Z-0633-2021 Class II HLS Set Advanced, Model: BEQ-HLS 7050 USA; HLS ... Nov 20, 2020
Z-0443-2021 Class II Blood Monitoring Unit (BMU 40), model no. 70104... Sep 30, 2020
Z-0133-2021 Class II Yuno II Mobile Operating Table - Product Usage:... Sep 24, 2020
Z-0132-2021 Class II Yuno Mobile Operating Table - Product Usage: In... Sep 24, 2020

Frequently Asked Questions

Sterility recalls for medical devices vary in severity. If you have already had a procedure using a potentially non-sterile device, contact your healthcare provider immediately — you may need monitoring for signs of infection. Symptoms to watch for include fever, redness or swelling at the surgical site, unusual pain, or discharge. For devices that have not yet been used, they should be quarantined and returned to the manufacturer per the recall notice. Non-sterile implants can cause serious infections; early detection and treatment are critical.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.