Sarns Modular Perfusion System 8000 base, 4 pump 100V. The Sarns Modular Perfusion System 8000...
FDA Device Recall #Z-2048-2012 — Class II — June 18, 2012
Recall Summary
| Recall Number | Z-2048-2012 |
| Classification | Class II — Moderate risk |
| Date Initiated | June 18, 2012 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Terumo Cardiovascular Systems Corporation |
| Location | Ann Arbor, MI |
| Product Type | Devices |
| Quantity | 26 units |
Product Description
Sarns Modular Perfusion System 8000 base, 4 pump 100V. The Sarns Modular Perfusion System 8000 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment.
Reason for Recall
Terumo Cardiovascular Systems has received several reports of malfunctions of the Air Bubble Detection System used with its heart-lung machines over the last three years. The malfunctions were expressed in various failure modes (over sensitivity, false alarms and were found to have various root causes. The malfunctions occurred in systems that were excluded from prior corrective actions or in syst
Distribution Pattern
Worldwide Distribution-USA (nationwide) and the countries of ARGENTINA, BANGLADESH, BELGIUM, Bolivia, BRAZIL, BRISBANE, CANADA, CHILE, China, COLOMBIA, COLOMBIA, Costa Rica, DENMARK, Dominican Republic, Ecuador, Egypt, FRANCE, GERMANY, Guatemala, HOLLAND, Honduras, Hong Kong, India, INDIA, Indonesia, Israel, ITALY, Jordan, KOREA, Kuwait, Lebanon, LEICESTERSHIRE, Malaysia, Mexico, NETHERLANDS, New Zealand, Nicaragua, NORWAY, Pakistan, Panama, PERU, Philippines, Russia, Saudi Arabia, SINGAPORE , SOUTH AFRICA, South Korea, SPAIN, Sri Lanka, SWEDEN, SWITZERLAND, Taiwan, THAILAND, Turkey, UNITED ARAB EMIRATES (UAE), Uruguay, Venezuela, Vietnam, WEST GERMANY and Yemen.
Lot / Code Information
Catalog Number 16409 and serial number 1001-1026
Other Recalls from Terumo Cardiovascular Systems Corpora...
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-2018-2025 | Class II | CDI OneView System BPM Probe: Cat. No. CDI75... | May 28, 2025 |
| Z-1492-2025 | Class II | Terumo CDI OneView Monitoring System Hematocrit... | Mar 3, 2025 |
| Z-1401-2024 | Class II | Cardiovascular Procedure Kits that include High... | Feb 20, 2024 |
| Z-1346-2023 | Class II | CAPIOX NX19 Hollow Fiber Oxygenator, Catalog Nu... | Mar 1, 2023 |
| Z-1347-2023 | Class II | CAPIOX FX25 Hollow Fiber Oxygenator, Catalog Nu... | Mar 1, 2023 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.