Brand Name: UroPass Ureteral Access Sheaths Product Name: Ureteral Access Sheaths Model/Catalog...

FDA Device Recall #Z-0364-2025 — Class II — September 11, 2024

Recall Summary

Recall Number Z-0364-2025
Classification Class II — Moderate risk
Date Initiated September 11, 2024
Status Ongoing
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Olympus Corporation of the Americas
Location Center Valley, PA
Product Type Devices
Quantity 14,093 units

Product Description

Brand Name: UroPass Ureteral Access Sheaths Product Name: Ureteral Access Sheaths Model/Catalog Number: 61024BX, 61038BX, 61046BX, 61054BX, 61124BX, 61138BX, 61146BX, 61154BX, 61224BX, 61238BX, 61246BX, 61254BX, 61324BX, 61338BX, 61346BX, 61354BX Software Version: N/A Product Description: The Olympus UroPass Ureteral Access Sheath Set consists of a hydrophilic coated outer sheath and an inner tapered dilator intended to establish a conduit for the passage of endoscopes and retrieval devices into the ureter. The hydrophilic coating on the UroPass Ureteral Access Sheath eases passage and placement. Both the outer sheath and inner dilator are radio-opaque for ease of viewing radiographically. Intended for single use only. Component: N/A

Reason for Recall

Exposure of the UroPass product to Ultraviolet ( UV ) Radiation can cause brittleness of the device dilator tip, which may lead to broken dilator tips in the package or in patients during surgical procedures.

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Australia, Canada, China, Germany, Hong Kong, Japan, Republic of South Korea, Singapore.

Lot / Code Information

Model/UDI-DI: 61024BX/00821925035317, 61038BX/00821925035324, 61046BX/00821925035331, 61054BX/00821925035348, 61124BX/00821925035355, 61138BX/00821925035362, 61146BX/00821925035379, 61154BX/00821925035386, 61224BX/00821925035393, 61238BX/00821925035409, 61246BX/00821925035416, 61254BX/00821925035423, 61324BX/00821925035430, 61338BX/00821925035447, 61346BX/00821925035454, 61354BX/00821925035461.

Other Recalls from Olympus Corporation of the Americas

Recall # Classification Product Date
Z-1641-2026 Class II Brand Name: Olympus HALO PKS Cutting Forceps, 5... Feb 25, 2026
Z-1640-2026 Class II Brand Name: Olympus PK Cutting Forceps, 5mm 33c... Feb 25, 2026
Z-1642-2026 Class II Brand Name: Olympus PKS Cutting Forceps, 5mm 33... Feb 25, 2026
Z-1639-2026 Class II Brand Name: Olympus Everest Bipolar Cutting For... Feb 25, 2026
Z-1643-2026 Class II Brand Name: Olympus PKS Cutting Forceps w/Cord,... Feb 25, 2026

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.