Brand Name: Olympus Everest Bipolar Cutting Forceps w/Cord, 5mm 33 cm Product Name: Everest Bipo...
FDA Device Recall #Z-1639-2026 — Class II — February 25, 2026
Recall Summary
| Recall Number | Z-1639-2026 |
| Classification | Class II — Moderate risk |
| Date Initiated | February 25, 2026 |
| Status | Ongoing |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Olympus Corporation of the Americas |
| Location | Center Valley, PA |
| Product Type | Devices |
| Quantity | 164 units |
Product Description
Brand Name: Olympus Everest Bipolar Cutting Forceps w/Cord, 5mm 33 cm Product Name: Everest Bipolar Cutting Forceps Model/Catalog Number: 3006 Product Description: The OLYMPUS Everest Bipolar Cutting Forceps are intended to be passed through a 5 mm cannula. Coagulation is achieved using electrosurgical energy under visualization. The device is intended to be used with the bipolar outputs of compatible generators.
Reason for Recall
Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.
Distribution Pattern
Worldwide - US Nationwide and the countries of Canada, Mexico, EU, Australia, China, Singapore, Korea, Japan.
Lot / Code Information
Model: 3006; UDI: 00821925035898; Lot Numbers: All unexpired;
Other Recalls from Olympus Corporation of the Americas
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1641-2026 | Class II | Brand Name: Olympus HALO PKS Cutting Forceps, 5... | Feb 25, 2026 |
| Z-1640-2026 | Class II | Brand Name: Olympus PK Cutting Forceps, 5mm 33c... | Feb 25, 2026 |
| Z-1642-2026 | Class II | Brand Name: Olympus PKS Cutting Forceps, 5mm 33... | Feb 25, 2026 |
| Z-1644-2026 | Class II | Brand Name: Olympus PKS Cutting Forceps, 5mm 33... | Feb 25, 2026 |
| Z-1643-2026 | Class II | Brand Name: Olympus PKS Cutting Forceps w/Cord,... | Feb 25, 2026 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.