Alaris Pump Module Model 8100

FDA Device Recall #Z-2736-2020 — Class I — June 30, 2020

Recall Summary

Recall Number Z-2736-2020
Classification Class I — Serious risk
Date Initiated June 30, 2020
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm CareFusion 303, Inc.
Location San Diego, CA
Product Type Devices
Quantity 1093588

Product Description

Alaris Pump Module Model 8100

Reason for Recall

Broken elements on pump module platen such as broken upper hinge post, lower hinge, and membrane frame can occur when the door is forcefully closed on an object, the device is dropped, or if improper cleaners are used on the membrane frame assembly. These may prevent devices from delivering accurate amounts of fluid, which may result in an over infusion, free-flow conditions, or under infusion.

Distribution Pattern

US: CA, FL, IN, TX, NC, LA, SC, VA, PA, OK, AZ, IL, KY, UT, AL, GA, TN, MA, MI, OH, AK, OR, MS, CO, NY, WI, KS, MO, MD, NV, AR, AS, ID, VT, NE, NJ, DC, HI, CT, MT, WV, NH, WA, DE, ME, NM, MN, WY, IA, SD, ND, MP, PR, GU. OUS: NL, MX, AU, MY, NZ, ZA, GB, FR, CH, CN, SG, TW, CA, IL, KW, QA, DE, BH, IN, SA, GI, AE, PK, PH, BE

Lot / Code Information

All serial numbers

Other Recalls from CareFusion 303, Inc.

Recall # Classification Product Date
Z-1645-2026 Class II BD Alaris" System with Guardrails" Suite MX wit... Feb 11, 2026
Z-1028-2026 Class II BD Pyxis ES Enterprise Server Catalog Numbers 1... Dec 3, 2025
Z-0962-2026 Class II BD Pyxis MedStation ES, REF: 323, BD Pyxis M... Nov 20, 2025
Z-1217-2026 Class II BD Alaris" Pump Module model 8100 UDI-DI code:... Nov 6, 2025
Z-0430-2026 Class I BD Alaris infusion Pump Module 8100, REF: 8100;... Oct 17, 2025

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.