Babylog VN500 Ventilator with Installed CO2 Measurement Option-intended for the ventilation of ne...

FDA Device Recall #Z-0109-2022 — Class II — September 10, 2021

Recall Summary

Recall Number Z-0109-2022
Classification Class II — Moderate risk
Date Initiated September 10, 2021
Status Ongoing
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Draeger Medical, Inc.
Location Telford, PA
Product Type Devices
Quantity 2,169 units

Product Description

Babylog VN500 Ventilator with Installed CO2 Measurement Option-intended for the ventilation of neonatal patients from 0.4 kg (0.88 lbs) up to 10 kg (22 lbs) and pediatric patients from 5 kg (11 lbs) up to 20 kg (44 lbs) bodyweight. Catalog Number: 8417400

Reason for Recall

Software 2.51 and Lower with Installed CO2 Measurement Option, may result in Restart of the Ventilator, brief cessation of ventilation can occur and loss of PEEP for approx. 8 seconds

Distribution Pattern

US Nationwide

Lot / Code Information

Software 2.51 and Lower

Other Recalls from Draeger Medical, Inc.

Recall # Classification Product Date
Z-0668-2025 Class I Brand Name: Atlan Product Name: Atlan A300XL ... Oct 15, 2024
Z-0304-2025 Class I Brand Name: Atlan Product Name: Atlan A350 Mo... Oct 15, 2024
Z-0305-2025 Class I Brand Name: Atlan Product Name: Atlan A350XL ... Oct 15, 2024
Z-0667-2025 Class I Brand Name: Atlan Product Name: Atlan A300 ... Oct 15, 2024
Z-1512-2024 Class I Dr¿ger Perseus A500 Anesthesia Workstation Mar 20, 2024

Frequently Asked Questions

A software recall means the device's embedded software or firmware has a defect that could affect its performance or safety. Many software recalls are corrected through firmware updates that can be applied without physically replacing the device. For implantable devices, the update may be delivered wirelessly during a routine clinic visit. For external devices, the manufacturer may provide updated software files or replacement units. Contact your healthcare provider to determine whether your specific device and software version are affected and what action is recommended.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.