Browse Device Recalls
682 FDA medical device safety recalls.
FDA Medical Device Recall Enforcement Database
Browse 682 FDA medical device recall enforcement actions. Each entry includes the device description, reason for recall, classification (Class I through III based on severity), recalling firm, and distribution details. Use the filters below to search by year, classification, state, or keyword.
Medical Device Recalls
Browse 682 FDA device recalls.
| Date | Product | Reason | Class | Firm |
|---|---|---|---|---|
| Jul 10, 2015 | WOUND CLOSURE TRAY II CUSTOMED 50/CS Used by medical staff to perform surg... | Recall initiated due to inadequate validation/qualification of sterilization processes, and packa... | Class II | Customed, Inc |
| Jul 10, 2015 | LAPAROTOMY PACK 4/CS Used by medical staff to perform surgical procedures a... | Recall initiated due to inadequate validation/qualification of sterilization processes, and packa... | Class II | Customed, Inc |
| Jul 10, 2015 | ORTHO-IMPLANT SURGICAL PACK, Used by medical staff to perform surgical ... | Recall initiated due to inadequate validation/qualification of sterilization processes, and packa... | Class II | Customed, Inc |
| Jul 10, 2015 | PERY GYN PACK CUSTOMED 10/CS Used by medical staff to perform surgical pro... | Recall initiated due to inadequate validation/qualification of sterilization processes, and packa... | Class II | Customed, Inc |
| Jul 10, 2015 | PERI GYN PACK 5/CS Used by medical staff to perform surgical procedures ... | Recall initiated due to inadequate validation/qualification of sterilization processes, and packa... | Class II | Customed, Inc |
| Jul 10, 2015 | CESAREAN PACK 3/CS Used by medical staff to perform surgical procedures an... | Recall initiated due to inadequate validation/qualification of sterilization processes, and packa... | Class II | Customed, Inc |
| Jul 10, 2015 | Gown, surgical 30/CS Used by medical staff to perform surgical procedures... | Recall initiated due to inadequate validation/qualification of sterilization processes, and packa... | Class II | Customed, Inc |
| Jul 10, 2015 | LABOR & DELIVERY 4/CS Used by medical staff to perform surgical procedures a... | Recall initiated due to inadequate validation/qualification of sterilization processes, and packa... | Class II | Customed, Inc |
| Jul 10, 2015 | LAPAROSCOPY PACK 6/CS Used by medical staff to perform surgical procedures a... | Recall initiated due to inadequate validation/qualification of sterilization processes, and packa... | Class II | Customed, Inc |
| Jul 10, 2015 | GAUZE SPONGE 2"X2" 8PLY COTTON 1500PK/CS Used by medical staff to perform s... | Recall initiated due to inadequate validation/qualification of sterilization processes, and packa... | Class II | Customed, Inc |
| Jul 10, 2015 | WOUND CLOSURE TRAY I CUSTOMED 50/CS Used by medical staff to perform surgi... | Recall initiated due to inadequate validation/qualification of sterilization processes, and packa... | Class II | Customed, Inc |
| Jul 10, 2015 | EXTREMITY ARM PACK 4/CS HOSP. DR. CENTER Used by medical staff to perform s... | Recall initiated due to inadequate validation/qualification of sterilization processes, and packa... | Class II | Customed, Inc |
| Jul 10, 2015 | WOUND MANAGEMENT CUSTOMED 20/CS Used by medical staff to perform surgical p... | Recall initiated due to inadequate validation/qualification of sterilization processes, and packa... | Class II | Customed, Inc |
| Jul 10, 2015 | SURG GOWN POLYREINF LGE CUSTOM 28/CS Used by medical staff to perform su... | Recall initiated due to inadequate validation/qualification of sterilization processes, and packa... | Class II | Customed, Inc |
| Jul 10, 2015 | UROLOGY PACK 4/CS Used by medical staff to perform surgical procedures and... | Recall initiated due to inadequate validation/qualification of sterilization processes, and packa... | Class II | Customed, Inc |
| Jul 10, 2015 | MINOR LACERATION TRAY CUSTOMED 50/CS Used by medical staff to perform su... | Recall initiated due to inadequate validation/qualification of sterilization processes, and packa... | Class II | Customed, Inc |
| Jul 10, 2015 | PREMIUM ARTHROSCOPY PK CUSTOMED 5/CS Used by medical staff to perform sur... | Recall initiated due to inadequate validation/qualification of sterilization processes, and packa... | Class II | Customed, Inc |
| Jul 10, 2015 | GENERAL MINOR PK 5/CS Used by medical staff to perform surgical procedures a... | Recall initiated due to inadequate validation/qualification of sterilization processes, and packa... | Class II | Customed, Inc |
| Jul 10, 2015 | MINOR PACK 4/CS Used by medical staff to perform surgical procedures and ... | Recall initiated due to inadequate validation/qualification of sterilization processes, and packa... | Class II | Customed, Inc |
| Jul 10, 2015 | LABOR & DELIVERY 4/CS Used by medical staff to perform surgical procedure... | Recall initiated due to inadequate validation/qualification of sterilization processes, and packa... | Class II | Customed, Inc |
| Oct 8, 2014 | Wound Management Tray Ill, Catalog number 900-540. Used by medical staff t... | Customed products are being recalled because of compromised sterility, not limited to potentially... | Class I | Customed, Inc |
| Oct 8, 2014 | Cataract Pack, Catalog # 900-944 Used by medical staff to perform patient ... | Significant compromised sterility, not limited to potentially damaged packaging. Serious deficie... | Class I | Customed, Inc |
| Oct 8, 2014 | Liposuction Pack, catalog #900-2656 Used by medical staff to perform patie... | Significant compromised sterility, not limited to potentially damaged packaging. Serious deficie... | Class I | Customed, Inc |
| Oct 8, 2014 | C-Section Surgical Pack, Catalog # 900-921A Used by medical staff to perfo... | Significant compromised sterility, not limited to potentially damaged packaging. Serious deficie... | Class I | Customed, Inc |
| Oct 8, 2014 | Wound Management Tray Ryder Memorial, Catalog number 900-1456. Used by med... | Customed products are being recalled because of compromised sterility, not limited to potentially... | Class I | Customed, Inc |
| Oct 8, 2014 | Extremity PACK, catalog #900-2615 Used by medical staff to perform patient... | Significant compromised sterility, not limited to potentially damaged packaging. Serious deficie... | Class I | Customed, Inc |
| Oct 8, 2014 | Laparotomy Surgical Pack, catalog #900-281A Used by medical staff to perfo... | Significant compromised sterility, not limited to potentially damaged packaging. Serious deficie... | Class I | Customed, Inc |
| Oct 8, 2014 | Laparoscopy Chole Pack, Catalog number 900-1688. Used by medical staff to ... | Customed products are being recalled because of compromised sterility, not limited to potentially... | Class I | Customed, Inc |
| Oct 8, 2014 | Surgeon Reinforced Gown XL, Catalog number 900-521. Used by medical staff ... | Customed products are being recalled because of compromised sterility, not limited to potentially... | Class I | Customed, Inc |
| Oct 8, 2014 | Major Laparotomy Pack, Catalog number 900-036. Used by medical staff to pe... | Customed products are being recalled because of compromised sterility, not limited to potentially... | Class I | Customed, Inc |
| Oct 8, 2014 | Transforaminal Lumbar Interbody Fusion (TLIF) pack, Catalog # 900-2918 Use... | Significant compromised sterility, not limited to potentially damaged packaging. Serious deficie... | Class I | Customed, Inc |
| Oct 8, 2014 | Irrigation Tray With Bulb Syringe, Catalog number 900-020. Used by medical... | Customed products are being recalled because of compromised sterility, not limited to potentially... | Class I | Customed, Inc |
| Oct 8, 2014 | Uretheral Catheter Tray, Catalog number 900-278. Used by medical staff to ... | Customed products are being recalled because of compromised sterility, not limited to potentially... | Class I | Customed, Inc |
| Oct 8, 2014 | Cesarean Pack, Catalog number 900-1689. Used by medical staff to perform p... | Customed products are being recalled because of compromised sterility, not limited to potentially... | Class I | Customed, Inc |
| Oct 8, 2014 | Umbilical Vessel Catheter Kit, Catalog number 900-1408. Used by medical st... | Customed products are being recalled because of compromised sterility, not limited to potentially... | Class I | Customed, Inc |
| Oct 8, 2014 | Arthroscopy Surgical Pack, Catalog # 900-740A Used by medical staff to per... | Significant compromised sterility, not limited to potentially damaged packaging. Serious deficie... | Class I | Customed, Inc |
| Oct 8, 2014 | Premium OB Pack, Catalog number 900-1297. Used by medical staff to perform... | Customed products are being recalled because of compromised sterility, not limited to potentially... | Class I | Customed, Inc |
| Oct 8, 2014 | Open Heart Tray, Catalog number 900-1196C Used by medical staff to perfor... | Customed products are being recalled because of compromised sterility, not limited to potentially... | Class I | Customed, Inc |
| Oct 8, 2014 | Laparoscopy Pack, Catalog number 900-1967. Used by medical staff to perfor... | Customed products are being recalled because of compromised sterility, not limited to potentially... | Class I | Customed, Inc |
| Oct 8, 2014 | Femoral Pack, Catalog number 900-164. Used by medical staff to perform pat... | Customed products are being recalled because of compromised sterility, not limited to potentially... | Class I | Customed, Inc |
| Oct 8, 2014 | Gown Surgical Non Reinforced XXL, Catalog number 900-2566. Used by medical... | Customed products are being recalled because of compromised sterility, not limited to potentially... | Class I | Customed, Inc |
| Oct 8, 2014 | Total Hysterectomy, catalog #900-2415 Used by medical staff to perform pat... | Significant compromised sterility, not limited to potentially damaged packaging. Serious deficie... | Class I | Customed, Inc |
| Oct 8, 2014 | Cystoscopy Pack III, catalog #900-2482 Used by medical staff to perform pa... | Significant compromised sterility, not limited to potentially damaged packaging. Serious deficie... | Class I | Customed, Inc |
| Oct 8, 2014 | Fanfold Drape, Catalog number 900-633. Used by medical staff to perform pa... | Significant compromised sterility, not limited to potentially damaged packaging. Serious deficie... | Class I | Customed, Inc |
| Oct 8, 2014 | Femoral Angiography Pack, Catalog number 900-796. Used by medical staff to... | Significant compromised sterility, not limited to potentially damaged packaging. Serious deficie... | Class I | Customed, Inc |
| Oct 8, 2014 | Laparotomy Basic Pack, Catalog number 900-547 Used by medical staff to pe... | Significant compromised sterility, not limited to potentially damaged packaging. Serious deficie... | Class I | Customed, Inc |
| Oct 8, 2014 | Laparotomy Pack, Catalog number 900-149. Used by medical staff to perform ... | Customed products are being recalled because of compromised sterility, not limited to potentially... | Class I | Customed, Inc |
| Oct 8, 2014 | Laparoscopy Pack, Catalog number 900-484 Used by medical staff to perform... | Significant compromised sterility, not limited to potentially damaged packaging. Serious deficie... | Class I | Customed, Inc |
| Oct 8, 2014 | Laparoscopy Pack, Catalog number 900-486. Used by medical staff to perform... | Significant compromised sterility, not limited to potentially damaged packaging. Serious deficie... | Class I | Customed, Inc |
| Oct 8, 2014 | Labor & Delivery Pack, Catalog number 900-168. Used by medical staff to pe... | Customed products are being recalled because of compromised sterility, not limited to potentially... | Class I | Customed, Inc |