| Date | Product | Reason | Class | Firm |
|---|---|---|---|---|
| Sep 26, 2018 | Atellica IM 1300 Analyzer. In vitro diagnostic testing of clinical specimens.... | Multiple Issues Identified in Atellica Solution System Software in V 1.14.2 and lower in: Atellic... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Sep 26, 2018 | Atellica CH 930 Analyzer. In vitro diagnostic testing of clinical specimens ... | Multiple Issues Identified in Atellica Solution System Software in V 1.14.2 and lower in: Atellic... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Sep 7, 2018 | Dimension Lipase Flex Reagent Cartridge, Model Number 10460277 Product Usa... | There is lot to lot variability for QC and patient results when using the identified reagent cart... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Sep 7, 2018 | Dimension Vista Lipase Flex Reagent Cartridge, Model Number 10461745 Produ... | There is lot to lot variability for QC and patient results when using the identified reagent cart... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Aug 14, 2018 | APTIO CENTRIFUGE MODULE | The modules may have been shipped without a warning label for potential risk of interference to p... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Aug 14, 2018 | APTIO 9000 REFRIGERATED STORAGE MODULE | The modules may have been shipped without a warning label for potential risk of interference to p... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Aug 14, 2018 | APTIO RACK OUTPUT MODULE | The modules may have been shipped without a warning label for potential risk of interference to p... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Aug 14, 2018 | STREAMLAB - REFRIGERATED STORAGE MODULE 15000 | The modules may have been shipped without a warning label for potential risk of interference to p... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Aug 14, 2018 | STREAMLAB - REFRIGERATED STORAGE MODULE 9000 | The modules may have been shipped without a warning label for potential risk of interference to p... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Aug 14, 2018 | APTIO INPUT/OUTPUT MODULE | The modules may have been shipped without a warning label for potential risk of interference to p... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Aug 14, 2018 | STREAMLAB CORE UNIT / LYNX WITH IOM | The modules may have been shipped without a warning label for potential risk of interference to p... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Aug 14, 2018 | APTIO RACK INPUT MODULE | The modules may have been shipped without a warning label for potential risk of interference to p... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Aug 14, 2018 | APTIO 15000 REFRIGERATED STORAGE MODULE | The modules may have been shipped without a warning label for potential risk of interference to p... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Jul 16, 2018 | RAPIDPoint 405 Blood Gas Analyzer; 10282093, 10310464, 10314817, 10317193, 10... | Therapeutic levels (1mg/mL and 2mg/mL) of Hydroxocobalamin may interfere with tHb and some of the... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Jul 16, 2018 | RAPIDPoint 500 Blood Gas Analyzer; 10492730, 1069855, 10696857, 10697306 ... | Therapeutic levels (1mg/mL and 2mg/mL) of Hydroxocobalamin may interfere with tHb and some of the... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Jul 16, 2018 | RAPIDLab 1245 Blood Gas Analyzer; 10321844, 10337179, 10491393 Device is ... | Therapeutic levels (1mg/mL and 2mg/mL) of Hydroxocobalamin may interfere with tHb and some of the... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Jul 16, 2018 | RAPIDLab 1265 Blood Gas Analyzer; 110321852, 10470366, 10491395 Device is... | Therapeutic levels (1mg/mL and 2mg/mL) of Hydroxocobalamin may interfere with tHb and some of the... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Jun 29, 2018 | Atellica IM 1300 Analyzer (SMN 11066001). The Atellica" Solution is a mult... | Atellica Solution system software (SW) version 1.13 and lower, five (5) behavioral issues were id... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Jun 29, 2018 | Atellica CH 930 Analyzer (SMN 11067000) . The Atellica" Solution is a multi-... | Atellica Solution system software (SW) version 1.13 and lower, five (5) behavioral issues were id... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Jun 29, 2018 | Atellica Sample Handler Connect (SMN 11069018). The Atellica" Solution is a ... | Atellica Solution system software (SW) version 1.13 and lower, five (5) behavioral issues were id... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Jun 29, 2018 | Atellica Sample Handler Additional (SMN 11069004). The Atellica" Solution ... | Atellica Solution system software (SW) version 1.13 and lower, five (5) behavioral issues were id... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Jun 29, 2018 | Atellica IM 1600 Analyzer (SMN 11066000). The Atellica" Solution is a multi-... | Atellica Solution system software (SW) version 1.13 and lower, five (5) behavioral issues were id... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Jun 29, 2018 | Atellica Sample Handler Prime (SMN 11069001). The Atellica" Solution is a ... | Atellica Solution system software (SW) version 1.13 and lower, five (5) behavioral issues were id... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Apr 19, 2018 | Dimension Vista Gentamicin Flex reagent cartridge, DV GENT, gentamicin enzyme... | The affected lots may exhibit inaccuracy for Quality Control and patient samples at the low end o... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Apr 19, 2018 | Dimension Gentamicin Flex reagent cartridge, DM GENT, gentamicin enzyme immu... | The affected lots may exhibit inaccuracy for Quality Control and patient samples at the low end o... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Mar 27, 2018 | ADVIA Chemistry Urea Nitrogen Reagent (material number 10309051) for use on A... | Urea Nitrogen reagent kit demonstrated an increased incidence of Calibration Failures. Results ca... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Mar 21, 2018 | Dimension Vista¿ CTNI Sample Diluent, KD692, SMN# 10445205 | The diluent may have an incomplete slit on the septum in the cap of the vial. A diluent cap with... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Mar 21, 2018 | Dimension Vista¿ MULTI 1 SDIL, KD693, SMN# 10469971 | The diluent may have an incomplete slit on the septum in the cap of the vial. A diluent cap with... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Mar 21, 2018 | Dimension Vista¿ MULTI 2 SDIL, KD694, SMN# 10483586 | The diluent may have an incomplete slit on the septum in the cap of the vial. A diluent cap with... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Mar 12, 2018 | Syva¿ EMIT¿ II Plus 6-Acetylmorphine Catalog # for 1000 mL: 9R129UL /SMN# 10... | Incorrect calibrator level listed in qualitative calibration steps | Class III | Siemens Healthcare Diagnostics, Inc. |
| Mar 12, 2018 | Syva¿ EMIT¿ II Plus 6-Acetyl morphine Catalog # for 28 mL: 9R039UL /SMN# 104... | Incorrect calibrator level listed in qualitative calibration steps | Class III | Siemens Healthcare Diagnostics, Inc. |
| Feb 26, 2018 | Atellica Solution and Atellica Sample Handler Prime for in vitro diagnostic t... | When more than one immunoassay analyzer is connected to a single solution system, reporting and c... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Feb 26, 2018 | Atelia IM 1300 Analyzer for in vitro diagnostic testing of clinical specimens. | When more than one immunoassay analyzer is connected to a single solution system, reporting and c... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Feb 26, 2018 | Analyzer and Atellica IM 1600 Analyzer system for in vitro diagnostic testing... | When more than one immunoassay analyzer is connected to a single solution system, reporting and c... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Jan 30, 2018 | Dimension Vista¿ Calcium Flex¿ reagent cartridge, Dimension Vista¿ CA, K1023,... | May produce erroneously low results from specific well sets. | Class II | Siemens Healthcare Diagnostics, Inc. |
| Jan 2, 2018 | ADVIA Centaur¿ Folate (FOL) | Concentrations for the level of biotin that does not interfere is not currently listed in the ins... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Jan 2, 2018 | IMMULITE ¿ /IMMULITE ¿ 1000 TG | Concentrations for the level of biotin that does not interfere is not currently listed in the ins... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Jan 2, 2018 | Dimension Vista¿ Cardiac Troponin I (CTNI) Flex¿ reagent cartridge | Concentrations for the level of biotin that does not interfere is not currently listed in the ins... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Jan 2, 2018 | IMMULITE ¿ 2000/IMMULITE ¿ 2000 XPi TG | Concentrations for the level of biotin that does not interfere is not currently listed in the ins... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Jan 2, 2018 | IMMULITE 2000/IMMULITE ¿ 2000 XPi OM-MA (CA125) | Concentrations for the level of biotin that does not interfere is not currently listed in the ins... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Jan 2, 2018 | Dimension¿ Thyroid Stimulating Hormone (TSHL) Flex¿ reagent cartridge | The concentrations for biotin listed in the non-interfering substances section of the instruction... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Jan 2, 2018 | IMMULITE ¿ 2000/IMMULITE ¿ 2000 XPi 3g Allergy specific IgE Universal Kit | Concentrations for the level of biotin that does not interfere is not currently listed in the ins... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Jan 2, 2018 | IMMULITE ¿ /IMMULITE ¿ 1000 Vitamin B12 | Concentrations for the level of biotin that does not interfere is not currently listed in the ins... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Jan 2, 2018 | ADVIA Centaur¿ Cyclosporine | The concentrations for biotin listed in the non-interfering substances section of the instruction... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Jan 2, 2018 | IMMULITE ¿ 2000/IMMULITE ¿ 2000 XPi Folic Acid | Concentrations for the level of biotin that does not interfere is not currently listed in the ins... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Jan 2, 2018 | Dimension Vista¿ Thyroid Stimulating Hormone (TSH) Flex¿ reagent cartridge | The concentrations for biotin listed in the non-interfering substances section of the instruction... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Jan 2, 2018 | Dimension¿ Sirolimus (SIRO) Flex¿ reagent cartridge | Concentrations for the level of biotin that does not interfere is not currently listed in the ins... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Jan 2, 2018 | IMMULITE ¿ 2000/IMMULITE ¿ 2000 XPi CEA | Concentrations for the level of biotin that does not interfere is not currently listed in the ins... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Jan 2, 2018 | IMMULITE ¿ /IMMULITE ¿ 1000 CEA | Concentrations for the level of biotin that does not interfere is not currently listed in the ins... | Class II | Siemens Healthcare Diagnostics, Inc. |
| Jan 2, 2018 | ADVIA Centaur¿ HBc IgM | Concentrations for the level of biotin that does not interfere is not currently listed in the ins... | Class II | Siemens Healthcare Diagnostics, Inc. |