| Date | Product | Reason | Class | Firm |
|---|---|---|---|---|
| May 29, 2015 | GE Healthcare Innova IGS 540. Angiographic X-ray systems used in generatin... | GE Healthcare has recently become aware of a potential safety issue that may result in the loss o... | Class II | GE Healthcare |
| May 29, 2015 | GE Healthcare Innova 3100IQ. Angiographic X-ray systems used in generating... | GE Healthcare has recently become aware of a potential safety issue that may result in the loss o... | Class II | GE Healthcare |
| May 29, 2015 | GE Healthcare Discovery IGS 740 & GE Healthcare Innova IGS 740. Angiograph... | GE Healthcare has recently become aware of a potential safety issue that may result in the loss o... | Class II | GE Healthcare |
| May 29, 2015 | GE Healthcare Innova IGS 620. Angiographic X-ray systems used in generatin... | GE Healthcare has recently become aware of a potential safety issue that may result in the loss o... | Class II | GE Healthcare |
| May 29, 2015 | GE Healthcare Innova IGS 630. Angiographic X-ray systems used in generatin... | GE Healthcare has recently become aware of a potential safety issue that may result in the loss o... | Class II | GE Healthcare |
| May 29, 2015 | GE Healthcare Innova IGS 530. Angiographic X-ray systems used in generatin... | GE Healthcare has recently become aware of a potential safety issue that may result in the loss o... | Class II | GE Healthcare |
| May 29, 2015 | ELI 380 Electrocardiograph. The ELI 380 is intended to be a high-performance... | Mortara Instrument, Inc. has recently become aware of a potential safety hazard involving our EL... | Class II | Mortara Instrument, Inc |
| May 29, 2015 | GE Healthcare Innova 2100IQ. Angiographic X-ray systems used in generating... | GE Healthcare has recently become aware of a potential safety issue that may result in the loss o... | Class II | GE Healthcare |
| May 29, 2015 | GE Healthcare Innova IGS 520. Angiographic X-ray systems used in generatin... | GE Healthcare has recently become aware of a potential safety issue that may result in the loss o... | Class II | GE Healthcare |
| May 29, 2015 | GE Healthcare Innova 2121IQ. Angiographic X-ray systems used in generating... | GE Healthcare has recently become aware of a potential safety issue that may result in the loss o... | Class II | GE Healthcare |
| May 29, 2015 | GE Healthcare Optima CL 323i. Angiographic X-ray systems used in generatin... | GE Healthcare has recently become aware of a potential safety issue that may result in the loss o... | Class II | GE Healthcare |
| May 29, 2015 | GE Healthcare Discovery IGS 730. Angiographic X-ray systems used in genera... | GE Healthcare has recently become aware of a potential safety issue that may result in the loss o... | Class II | GE Healthcare |
| May 29, 2015 | GE Healthcare Optima IGS 320. Angiographic X-ray systems used in generatin... | GE Healthcare has recently become aware of a potential safety issue that may result in the loss o... | Class II | GE Healthcare |
| May 4, 2015 | GE Healthcare, CARESCAPE Monitor B850, B650 and B450 (Bx50). | The heart rate could be calculated from pacer pulses without indication that pacemaker detection ... | Class II | GE Healthcare |
| Apr 24, 2015 | D-fend, (p/n 876446-HEL), and D-fend+ (p/n 881319-HEL) water trap. Used with... | A potential safety issue due to loss of gas monitoring associated with occlusion of specific D-Fe... | Class II | GE Healthcare |
| Apr 22, 2015 | CT LightSpeed 16 and Discovery ST Pet/CT DST. Catalogue # 2339985. Designe... | GE Healthcare has become aware that 3 bolts that secure a power supply to the rotating side of th... | Class II | GE Healthcare |
| Apr 9, 2015 | Merge RadSuite software. Radiological image processing system. | When RadSuite is used with IPID (Issuer of Patient ID) as a part of the "Patient Identifier," it ... | Class II | Merge Healthcare, Inc. |
| Mar 30, 2015 | Schiller AG ISA CO2 Sensor (AKA: PhaseIn Unit), Merge Part Number HW-HEMO-000... | The units may display the alarm "sampling line clogged" after the flow zeroing procedure that is ... | Class II | Merge Healthcare, Inc. |
| Mar 12, 2015 | GE Healthcare, Treadmill T2100. For Exercise Testing, facilitating accurate ... | GE Healthcare has recently become aware of a potential safety issue due to uncontrolled motion of... | Class II | GE Healthcare |
| Mar 9, 2015 | GE Healthcare, SIGNA 1.5T TWINSPEED, SIGNA INFINITY MRISYSTEM, SIGNA 3.0T INF... | GE Healthcare has become aware of a potential safety issue involving MRI systems due to software ... | Class II | GE Healthcare |
| Mar 9, 2015 | Kenex Radiation Shield installed with GE Healthcare Interventional, Surgery, ... | GE Healthcare has recently become aware of an improper installation of a ceiling suspended penden... | Class II | GE Healthcare |
| Mar 5, 2015 | GE Healthcare Millennium VG, Discovery VH and VariCam Nuclear Medicine system... | GE Healthcare has recently become aware of a potential safety issue due to a collimator falling o... | Class II | GE Healthcare |
| Mar 5, 2015 | Merge LIS software. | There is a potential for duplicate container numbers to be created for patients. | Class II | Merge Healthcare, Inc. |
| Feb 27, 2015 | GE Healthcare, Aespire 7900, Aespire View Anesthesia Machines. Model numbers... | Specific Aespire 7900 Anesthesia Machines have an increased likelihood of failure of the CPU circ... | Class II | GE Healthcare |
| Feb 20, 2015 | Endo TCS Driver Straight Hexalobe #6 P/N 5210-1004, Non-Sterile, Manufactured... | The TCS Straight Hexlobular #6 Drivers (5210-1004) tip sheared off during use in procedures that ... | Class II | TITAN SPINE, LLC |
| Feb 16, 2015 | GE Healthcare, Treadmill T2100 and Treadmill T2000 with power cord part numbe... | Potential safety issue due to improper assembly associated with the power cords of T2100 and T200... | Class II | GE Healthcare |
| Feb 2, 2015 | Axial Torque Limiting Driver (1st generation) Product Usage: The Bradshaw... | A component failure could cause a situation where the applied torque was significantly higher tha... | Class II | Bradshaw Medical Inc. |
| Jan 30, 2015 | GE Healthcare Proteus XR/a with wall stand model number 2260354. Intended ... | Potential breakage of the steel cable which supports the bucky device in certain wall stands of P... | Class II | GE Healthcare |
| Jan 27, 2015 | GE Healthcare Engstrom Carestation, Engstrom Pro, Cart Only S/5 EC. GE model ... | Potential failure of the caster mounting hardware, which may result in the ventilator falling/tip... | Class II | GE Healthcare |
| Jan 21, 2015 | Merge Hemo, 9.10, 9.20.0, 9.20.1,9.20.2, 9.30, 9.40.0, 9.40.1, 9.40.2 with Ma... | A system freeze-up of the Merge Hemo system, that included the PHASEIN End Tidal CO2 unit, result... | Class II | Merge Healthcare, Inc. |
| Jan 14, 2015 | Ikaria, INOmax DSIR (Delivery System), Model 10007. Nitric oxide delivery s... | Potential delivery failure alarm condition. INOmax DSIR with software version 2.0.4 and a certa... | Class II | INO Therapeutics (dba Ikaria) |
| Jan 7, 2015 | This correction applies to all GE Healthcare MRI systems with superconducting... | At certain sites, the MRU may not be connected to the magnet. In emergency situations, a discon... | Class I | GE Healthcare |
| Dec 19, 2014 | GE Healthcare Revolution CT., multi-slice (256 detector row) CT scanner. | Potential safety issue due to a required torque check on three bolted joints of the Revolution CT... | Class II | GE Healthcare |
| Dec 19, 2014 | GE Healthcare Automatic Mobile X-Ray (AMX) Series: Optima XR200 amx | An issue was identified with the mAs accuracy check performed on the system that could result in ... | Class II | GE Healthcare |
| Dec 19, 2014 | GE Healthcare Automatic Mobile X-Ray (AMX) Series: Brivo XR285amx | An issue was identified with the mAs accuracy check performed on the system that could result in ... | Class II | GE Healthcare |
| Dec 19, 2014 | GE Healthcare Automatic Mobile X-Ray (AMX) Series: Optima XR220 amx | An issue was identified with the mAs accuracy check performed on the system that could result in ... | Class II | GE Healthcare |
| Dec 15, 2014 | PF4 Concentrated Wash (10x), REF 403624, packaged in the following assay kits... | Some Lots of PF4 Concentrated Wash (10X) are a slightly more dilute concentration. This may lead ... | Class II | Genetic Testing Institute,inc |
| Dec 10, 2014 | GE Healthcare, Discovery IGS 730 & Discovery IGS 740. Discovery IGS 730 : ... | Potential unintentional system motion or rotation due to a positioning user interface issue with ... | Class II | GE Healthcare |
| Dec 8, 2014 | GE Healthcare, Proteus XR/a with wall stand model number 600-0301. | GE Healthcare has recently become aware of a potential safety issue involving the breakage of the... | Class II | GE Healthcare |
| Nov 21, 2014 | TomoTherapy Treatment System with software versions 2.0.1 / 2.0.2 / 2.0.3/ (H... | Accuray has become aware of a potential safety issue related to the TomoTherapy Treatment System ... | Class II | TomoTherapy Incorporated |
| Nov 12, 2014 | The Revolution CT is a multi-slice (256 detector row) CT scanner consisting o... | A required quality control test was not performed during installation associated with the softwar... | Class II | GE Healthcare |
| Oct 24, 2014 | GE Discovery MR950 MRI system | The alignment lasers are missing the labels required by radiation safety regulations. | Class II | GE Healthcare |
| Oct 17, 2014 | INOMAX DSIR Nitric Oxide delivery system, Model 10007, software version 3.1.0... | An issue has been identified in the INOmax DSIR system that could result in monitored Nitric Oxid... | Class II | INO Therapeutics (dba Ikaria) |
| Oct 14, 2014 | GE Healthcare Varicam/VG and VG Hawkeye devices. Models H3000YA, H3000YB, H... | GE is issuing this recall due to a potential safety issue related to a portion of the system fall... | Class II | GE Healthcare, LLC |
| Oct 14, 2014 | GE Healthcare Infinia (including Infinia, Infinia Hawkeye, Infinia II, Infini... | GE is issuing this recall due to a potential safety issue related to a portion of the system fall... | Class II | GE Healthcare, LLC |
| Oct 14, 2014 | Brivo NM615, is an all-purpose, single detector integrated nuclear imaging sy... | GE is issuing this recall due to a potential safety issue related to a portion of the system fall... | Class II | GE Healthcare, LLC |
| Oct 14, 2014 | Optima NM/CT 640, Models H3100YA, H3100YC. Product Usage: The GE Optima... | GE is issuing this recall due to a potential safety issue related to a portion of the system fall... | Class II | GE Healthcare, LLC |
| Oct 14, 2014 | Discovery NM/CT 670 dual detector free-geometry integrated nuclear imaging sy... | GE is issuing this recall due to a potential safety issue related to a portion of the system fall... | Class II | GE Healthcare, LLC |
| Oct 14, 2014 | Discovery NM 630, dual detector free-geometry integrated nuclear imaging syst... | GE is updating the Preventative Maintenance procedure and schedule, and added an additional Preve... | Class II | GE Healthcare, LLC |
| Oct 10, 2014 | Transport Chair and Rollator in 1, Model HE21630011 Burgundy and HE21630011 B... | Silver Star Brands is recalling the Transport Chair and Rollator in One because they have receive... | Class II | Silver Star Brands |