| Date | Product | Reason | Class | Firm |
|---|---|---|---|---|
| Sep 29, 2023 | Alinity c Iron Reagent, Reference Number 08P3920 | A high recovery of Iron assay has been observed at some customer sites. Internal studies confirm... | Class II | Sentinel CH SpA |
| Sep 29, 2023 | IRON assay, Reference Numbers 6K95-41 and 6K95-30 | A high recovery of Iron assay has been observed at some customer sites. Internal studies confirm... | Class II | Sentinel CH SpA |
| Sep 25, 2023 | Uretero-reno fiberscope-For endoscopic diagnosis and treatment within the ure... | Channel air drying process was not validated, and that a small percentage of scopes returned to c... | Class II | Aizu Olympus Co., Ltd. |
| Sep 25, 2023 | Hysteroscope-For endoscopic diagnosis within the uterus Model Number: HYF-1T. | Channel air drying process was not validated, and that a small percentage of scopes returned to c... | Class II | Aizu Olympus Co., Ltd. |
| Sep 25, 2023 | Mobile airway scope-diagnosis and observation to access airway anatomy, endot... | Channel air drying process was not validated, and that a small percentage of scopes returned to c... | Class II | Aizu Olympus Co., Ltd. |
| Sep 25, 2023 | Ultrasonic gastrovideoscope-Real time ultrasound imaging,ultrasound guided ne... | Channel air drying process was not validated, and that a small percentage of scopes returned to c... | Class II | Aizu Olympus Co., Ltd. |
| Sep 25, 2023 | Colonoscope: for endoscopy and endoscopic surgery within the lower digestive ... | Channel air drying process was not validated, and that a small percentage of scopes returned to c... | Class II | Aizu Olympus Co., Ltd. |
| Sep 25, 2023 | Small intestinal videoscope-For endoscopy and endoscopic surgery within the u... | Channel air drying process was not validated, and that a small percentage of scopes returned to c... | Class II | Aizu Olympus Co., Ltd. |
| Sep 25, 2023 | Hysterofiberscope-For endoscopic diagnosis within the uterus. Model Number:... | Channel air drying process was not validated, and that a small percentage of scopes returned to c... | Class II | Aizu Olympus Co., Ltd. |
| Sep 25, 2023 | Duodenovideoscope-EndoTherapy accessories (such as a biopsy forceps), and oth... | Channel air drying process was not validated, and that a small percentage of scopes returned to c... | Class II | Aizu Olympus Co., Ltd. |
| Sep 25, 2023 | Gastrointestinal videoscope-indicated for use within the upper digestive trac... | Channel air drying process was not validated, and that a small percentage of scopes returned to c... | Class II | Aizu Olympus Co., Ltd. |
| Sep 25, 2023 | Hysterovideoscope-For endoscopic diagnosis within the uterus Model Number: H... | Channel air drying process was not validated, and that a small percentage of scopes returned to c... | Class II | Aizu Olympus Co., Ltd. |
| Sep 25, 2023 | Tracheal videoscope- For airway management, which includes observation to acc... | Channel air drying process was not validated, and that a small percentage of scopes returned to c... | Class II | Aizu Olympus Co., Ltd. |
| Sep 25, 2023 | Cystonephroscope-endoscopic diagnosis and treatment within the bladder, ureth... | Channel air drying process was not validated, and that a small percentage of scopes returned to c... | Class II | Aizu Olympus Co., Ltd. |
| Sep 25, 2023 | Transnasal esophagovideoscope-For endoscopy and endoscopic surgery within the... | Channel air drying process was not validated, and that a small percentage of scopes returned to c... | Class II | Aizu Olympus Co., Ltd. |
| Sep 25, 2023 | Ureteralscope- for endoscopic diagnosis and treatment within the ureter, kidn... | Channel air drying process was not validated, and that a small percentage of scopes returned to c... | Class II | Aizu Olympus Co., Ltd. |
| Sep 25, 2023 | Bronchoscope - Ultrasonic: endoscopic real-time ultrasonic imaging, ultrasoun... | Channel air drying process was not validated, and that a small percentage of scopes returned to c... | Class II | Aizu Olympus Co., Ltd. |
| Sep 25, 2023 | Colonoscope-indicated for use within the lower digestive tract (including the... | Channel air drying process was not validated, and that a small percentage of scopes returned to c... | Class II | Aizu Olympus Co., Ltd. |
| Sep 25, 2023 | Rhinolaryngoscope-for endoscopic diagnosis and treatment within the nasal lum... | Channel air drying process was not validated, and that a small percentage of scopes returned to c... | Class II | Aizu Olympus Co., Ltd. |
| Sep 25, 2023 | Bronchoscope-for endoscopy and endoscopic surgery within the airways and trac... | Channel air drying process was not validated, and that a small percentage of scopes returned to c... | Class II | Aizu Olympus Co., Ltd. |
| Sep 25, 2023 | Choleodochoscope. Model Numbers: CHF-BP30, CHF-CB30L, CHF-P60. | Channel air drying process was not validated, and that a small percentage of scopes returned to c... | Class III | Aizu Olympus Co., Ltd. |
| Sep 25, 2023 | Gastroscope-endoscopic ultrasound (EUS) guided fine needle aspiration (FNA) a... | Channel air drying process was not validated, and that a small percentage of scopes returned to c... | Class II | Aizu Olympus Co., Ltd. |
| Sep 25, 2023 | Tracheal fiberscope-airway management, which includes observation to access a... | Channel air drying process was not validated, and that a small percentage of scopes returned to c... | Class II | Aizu Olympus Co., Ltd. |
| Sep 25, 2023 | Signmoidovideoscope-For endoscopy and endoscopic surgery within the lower dig... | Channel air drying process was not validated, and that a small percentage of scopes returned to c... | Class II | Aizu Olympus Co., Ltd. |
| Sep 18, 2023 | IBA Proton Therapy System - PROTEUS 235. PTS-11 versions, PTS-12 versions wit... | The patient positioning system can perform motions in unexpected directions, creating the potenti... | Class II | Ion Beam Applications S.A. |
| Sep 13, 2023 | Astral 100 and Astral 150 ventilators | If ventilator is on internal battery, not intended to serve as a primary power source, low/critic... | Class I | ResMed Ltd. |
| Sep 12, 2023 | RayCare 5B and 6A including service packs- An oncology information system use... | An issue was found in RayCare 5B, 6A, including service packs, where values will not be visible f... | Class II | RAYSEARCH LABORATORIES AB |
| Sep 11, 2023 | Sapphire Infusion Pumps, Models: Multi-Therapy (REF 15031-000-0028), Epidural... | Infusion Pumps with affected software revision may fail to detect air in line, which may lead to ... | Class I | EITAN MEDICAL LTD |
| Sep 8, 2023 | F2-01 Frame, physiological patient monitor (with arrhythmia detection or alarms) | There is a potential interruption of data communication between E-modules inserted in the F2-01 F... | Class II | GE Medical Systems China Co., Ltd. |
| Sep 6, 2023 | RX Daytona Plus (without ISE)- IVD Clinical Chemistry Analyser (1) RX4041, (... | RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not mee... | Class II | Randox Laboratories Ltd. |
| Sep 6, 2023 | AU/DxC AU Chemistry Transferrin REF: OSR6152 Immune complexes formed in solu... | The firm has identified that the Transferrin reagent does not meet the Instructions for Use (IFU)... | Class II | Beckman Coulter Biomedical Division |
| Sep 6, 2023 | RX Daytona Plus (with ISE)-IVD Clinical Chemistry Analyser Catalog Number: R... | RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not mee... | Class II | Randox Laboratories Ltd. |
| Sep 6, 2023 | RX Imola-IVD Clinical Chemistry Analyzer Catalog Number: RX4900. RX4900R (re... | RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not mee... | Class II | Randox Laboratories Ltd. |
| Sep 6, 2023 | RX Monaco-IVD Clinical Chemistry Analyser Catalog Number: RX5000 | RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not mee... | Class II | Randox Laboratories Ltd. |
| Sep 6, 2023 | RX Modena-IVD Clinical Chemistry Analyser Catalog Number: RX9000 | RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not mee... | Class II | Randox Laboratories Ltd. |
| Sep 6, 2023 | Venue Go Standard Carts Ref: (H45181VC and H45103VCW), used with Venue Go R2,... | Some Venue Go Standard Carts can develop an internal failure of the vertical/horizontal tilt adju... | Class II | GE MEDICAL SYSTEMS, ISRAEL LTD. |
| Sep 5, 2023 | MAZOR X robotic guidance system, Model TPL0059, with software versions 4.2.2 ... | Software update | Class II | Mazor Robotics Ltd |
| Aug 29, 2023 | PRIMA TT Glenoid Monoblock/Modular reverse TT baseplates: Description/Product... | Potential for baseplate peripheral holes to be out of specification due to a manufacturing issue,... | Class II | Limacorporate S.p.A |
| Aug 25, 2023 | Multichem IA Plus, Part Codes: a) 08P86-10, b) 08P86-19 | Technopath Manufacturing Ltd. has identified an issue with the glass vial commodity used to fill ... | Class II | TECHNO-PATH MANUFACTURING LTD. |
| Aug 25, 2023 | Multichem S Plus, Part Codes: a) 08P88-10, b) 08P88-11, c) 08P88-12 | Technopath Manufacturing Ltd. has identified an issue with the glass vial commodity used to fill ... | Class II | TECHNO-PATH MANUFACTURING LTD. |
| Aug 25, 2023 | Multichem P, Part Code 08P90-10 | Technopath Manufacturing Ltd. has identified an issue with the glass vial commodity used to fill ... | Class II | TECHNO-PATH MANUFACTURING LTD. |
| Aug 23, 2023 | LFS Flat screen arm system with surgical lamp for Models Aurora series, Auror... | The stopper block on the LFS bracket could be broken or missing. Without this stopper block, the ... | Class II | DKK Dai-Ichi Shomei Co., Ltd. |
| Aug 23, 2023 | LFS Flat screen support arm systems without surgical lamp for models Aurora s... | The stopper block on the LFS bracket could be broken or missing. Without this stopper block, the ... | Class II | DKK Dai-Ichi Shomei Co., Ltd. |
| Aug 22, 2023 | therascreen ¿KRAS RGQ PCR Kit (24)-IVD qualitative PCR assay used on the Roto... | Handbook Revision (Rev13) a dilution step is indicated as an Action when having the Flag SAMPLE_... | Class II | Qiagen GmbH |
| Aug 18, 2023 | OPMI LUMERA 300, REF 6137 | Surgical microscope for use in ophthalmology may be missing a screw from the suspension arm, whic... | Class II | Carl Zeiss Suzhou Co., Ltd. |
| Aug 17, 2023 | SOZO Bilateral Arm L-Dex Software | Bilateral L-Dex assessment software does not have the same level of sensitivity to help detect ea... | Class II | Impedimed Limited |
| Aug 17, 2023 | exoplan version 3.1 Rijeka Software -A medical software, intended to support ... | A software library filtering error has been discovered which occurs when users are using the Step... | Class II | Exocad GmbH |
| Aug 10, 2023 | Functional MR V1.0 | When exporting regions of interest in DICOM SEGMENTATION format, when simultaneously exporting mu... | Class II | OLEA MEDICAL |
| Aug 8, 2023 | Oculus Pentacam AXL Wave, Ref 70020, CE 0123 | Their is a potential that optical devices with insufficient anti-reflective coating may lead to... | Class II | Oculus Optikgeraete GMBH |
| Aug 8, 2023 | Oculus Myopia Master , Ref 68100, CE 0123 | Their is a potential that optical devices with insufficient anti-reflective coating may lead to... | Class II | Oculus Optikgeraete GMBH |