Atellica CH Iron3 - IVD use in the quantitative measurement of iron in human serum and plasma (li...

FDA Recall #Z-1549-2024 — Class II — March 13, 2024

Recall #Z-1549-2024 Date: March 13, 2024 Classification: Class II Status: Ongoing

Product Description

Atellica CH Iron3 - IVD use in the quantitative measurement of iron in human serum and plasma (lithium heparin and sodium heparin) using the Atellica¿ CI Analyzer and Atellica¿ CH Analyzer Siemens Material Number (SMN): 11537211

Reason for Recall

Potential for falsely elevated Chol_2, LDLC, and Trig_2 results on the Atellica CH and Atellica CI analyzers when the previous result in the cuvette was Iron3. Results in a positive bias ranging from 2-16% -impacts calibrator, quality control (QC), and patient results.

Recalling Firm

Siemens Healthcare Diagnostics, Inc. — Tarrytown, NY

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

8721 units

Distribution

Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, Colombia, Croatia, Curacao, St¿Eus, Czech Republic, Denmark, Estonia, Finland, France, Germany, ¿ Greece, Hungary, India, Ireland, Israel, Italy, Latvia, Lithuania, Mexico, Netherlands, ¿¿¿¿ New Zealand, Norway, Paraguay, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, U.A.E., United¿Kingdom.¿

Code Information

UDI-DI: 00630414610849 All lots

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated