Oxford Partial Knee System Left Medial Tibial Trial Tray Size E Product Usage: Instruments for ...

FDA Device Recall #Z-1161-2017 — Class II — January 5, 2016

Recall Summary

Recall Number Z-1161-2017
Classification Class II — Moderate risk
Date Initiated January 5, 2016
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Biomet U.K., Ltd.
Location Bridgend South Wales, N/A
Product Type Devices
Quantity 72 units in total

Product Description

Oxford Partial Knee System Left Medial Tibial Trial Tray Size E Product Usage: Instruments for use with the Oxford Uni Partial Knee System

Reason for Recall

Zimmer Biomet is conducting a medical device field action for various lot-specific Oxford Knee instruments due to potential alumina inclusions in the raw material batch used to produce the affected products. Inclusions contained within the finished product could lead to the cracking and separation of the instrument.

Distribution Pattern

Worldwide Distribution in the states of LA, NC, AR, MT, MO, CA, NY, KS, TX, MA, KY, OK, SC, NJ and the countries of Foreign: THAILAND, JAPAN, NETHERLANDS

Lot / Code Information

Item: 32-420736, Lots: ZB160501, ZB160601

Other Recalls from Biomet U.K., Ltd.

Recall # Classification Product Date
Z-1169-2017 Class II Oxford Unicompartmental Knee Phase 3 Shim Size ... Jan 5, 2016
Z-1164-2017 Class II Oxford Partial Knee System Right Medial Tibial ... Jan 5, 2016
Z-1156-2017 Class II Oxford Partial Knee System Right Medial Tibial ... Jan 5, 2016
Z-1175-2017 Class II Oxford Partial Knee System Phase 3 Tibial Templ... Jan 5, 2016
Z-1160-2017 Class II Oxford Partial Knee System Right Medial Tibial ... Jan 5, 2016

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.